M. G. Naón, S. Marelli, A. P. Moorhead, B. Saggin, G. Moschioni, M. Tarabini
{"title":"Development of a device to impose medio-lateral whole-body vibration while walking","authors":"M. G. Naón, S. Marelli, A. P. Moorhead, B. Saggin, G. Moschioni, M. Tarabini","doi":"10.1109/MeMeA52024.2021.9478701","DOIUrl":null,"url":null,"abstract":"A compact and light-weight set-up to impose medio-lateral vibration while walking has been designed and manufactured. A vibrating plate was actuated by a motor and a linear guide. After the design and finite-element analysis, the set-up has been manufactured and tested in the frequency band 0.5-4 Hz with amplitude below ± 20 mm, that is compatible for human testing. Total harmonic distortion has been measured below −30 dB. Crosstalk values along Z and Y measured at the corners of the moving platform were lower than 0.1 at 4 Hz. Given the previous analyses, the set-up can be used to test in a movement laboratory the effect of vibration during walking, mounting a walking pad on the platform.","PeriodicalId":429222,"journal":{"name":"2021 IEEE International Symposium on Medical Measurements and Applications (MeMeA)","volume":"44 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2021-06-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2021 IEEE International Symposium on Medical Measurements and Applications (MeMeA)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/MeMeA52024.2021.9478701","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
A compact and light-weight set-up to impose medio-lateral vibration while walking has been designed and manufactured. A vibrating plate was actuated by a motor and a linear guide. After the design and finite-element analysis, the set-up has been manufactured and tested in the frequency band 0.5-4 Hz with amplitude below ± 20 mm, that is compatible for human testing. Total harmonic distortion has been measured below −30 dB. Crosstalk values along Z and Y measured at the corners of the moving platform were lower than 0.1 at 4 Hz. Given the previous analyses, the set-up can be used to test in a movement laboratory the effect of vibration during walking, mounting a walking pad on the platform.