{"title":"Bioimpedance monitoring with improved accuracy using three-level stimulus","authors":"T. Paavle, P. Annus, A. Kuusik, R. Land, M. Min","doi":"10.1109/ECCTD.2007.4529620","DOIUrl":null,"url":null,"abstract":"Operation principles, modeling, and practical design of a novel system for bioimpedance (BI) measurement and monitoring are introduced in this paper. The system employs excitation and reference signals in a form of three-level rectangular waveforms with specially shortened duty cycles for eliminating some higher harmonics from the measured response. It enables to decrease systematic errors and ensures simplicity and power efficiency for wearable devices at the same time. The system is designed as a four-channel instrument, while both, the analog and digital approaches are discussed.","PeriodicalId":445822,"journal":{"name":"2007 18th European Conference on Circuit Theory and Design","volume":"37 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2007-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"4","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2007 18th European Conference on Circuit Theory and Design","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/ECCTD.2007.4529620","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 4

Abstract

Operation principles, modeling, and practical design of a novel system for bioimpedance (BI) measurement and monitoring are introduced in this paper. The system employs excitation and reference signals in a form of three-level rectangular waveforms with specially shortened duty cycles for eliminating some higher harmonics from the measured response. It enables to decrease systematic errors and ensures simplicity and power efficiency for wearable devices at the same time. The system is designed as a four-channel instrument, while both, the analog and digital approaches are discussed.
利用三电平刺激提高精度的生物阻抗监测
本文介绍了一种新型生物阻抗(BI)测量与监测系统的工作原理、建模和实际设计。该系统采用三电平矩形波形形式的激励和参考信号,特别缩短了占空比,以消除测量响应中的一些高次谐波。它可以减少系统误差,同时保证可穿戴设备的简单性和功耗效率。该系统设计为四通道仪器,并对模拟和数字两种方法进行了讨论。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信