Reciprocity Calibration of Hydrophones in High Intensity Focused Ultrasound Field

Longyang Jia, W. Shou, Bing Hu
{"title":"Reciprocity Calibration of Hydrophones in High Intensity Focused Ultrasound Field","authors":"Longyang Jia, W. Shou, Bing Hu","doi":"10.11648/j.ajpa.20200802.11","DOIUrl":null,"url":null,"abstract":"The primary problem of safety and efficiency for the high intensity therapeutic ultrasound (HITU) is the acoustic measure and dose control. The key technique is the pressure and intensity in the acoustic field especial in the focal region using the small calibrated hydrophone. The calibration accuracy of the used hydrophone is very important for HITU. Although the small hydrophone calibration has realized but there was no report of the hydrophone calibration in high pressure field. In this paper, our objective is to develop an absolute calibration method for the measurement of free field voltage sensitivity of hydrophone for high intensity focused ultrasound. First the acoustic pressure at the focal point by the self-reciprocity method of spherically curved auxiliary transducer is calibrated, then the free field voltage sensitivity of hydrophone at the geometric focal point of the calibrated pressure is obtained. The spatial average effect of acoustic pressure on hydrophone surface at the focal point is theoretically modified, and the expression and value table of correction coefficient of spatial average effect of hydrophone are given. The maximum acoustic pressure measured at the focal point was up to 5.58MPa (1.02kW/cm2) and used to calibrate a hydrophone from 0.95 MHz to 1.10 MHz with maximum local distortion parameter 0.72. The results show the rationality and feasibility of the measurement principle and method.","PeriodicalId":329149,"journal":{"name":"American Journal of Physics and Applications","volume":"66 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2020-05-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"American Journal of Physics and Applications","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.11648/j.ajpa.20200802.11","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

The primary problem of safety and efficiency for the high intensity therapeutic ultrasound (HITU) is the acoustic measure and dose control. The key technique is the pressure and intensity in the acoustic field especial in the focal region using the small calibrated hydrophone. The calibration accuracy of the used hydrophone is very important for HITU. Although the small hydrophone calibration has realized but there was no report of the hydrophone calibration in high pressure field. In this paper, our objective is to develop an absolute calibration method for the measurement of free field voltage sensitivity of hydrophone for high intensity focused ultrasound. First the acoustic pressure at the focal point by the self-reciprocity method of spherically curved auxiliary transducer is calibrated, then the free field voltage sensitivity of hydrophone at the geometric focal point of the calibrated pressure is obtained. The spatial average effect of acoustic pressure on hydrophone surface at the focal point is theoretically modified, and the expression and value table of correction coefficient of spatial average effect of hydrophone are given. The maximum acoustic pressure measured at the focal point was up to 5.58MPa (1.02kW/cm2) and used to calibrate a hydrophone from 0.95 MHz to 1.10 MHz with maximum local distortion parameter 0.72. The results show the rationality and feasibility of the measurement principle and method.
高强度聚焦超声场中水听器的互易校准
高强度治疗超声(HITU)安全性和有效性的首要问题是声学测量和剂量控制。利用小型标定水听器测量声场,特别是焦点区域的声场压力和声场强度是关键技术。使用的水听器的标定精度对HITU至关重要。虽然实现了小型水听器的标定,但目前尚无高压场水听器标定的报道。本文的目的是建立一种测量高强度聚焦超声水听器自由场电压灵敏度的绝对校准方法。首先利用球弯曲辅助换能器的自互易法对焦点处的声压进行标定,然后得到标定压力几何焦点处水听器的自由场电压灵敏度。从理论上修正了焦点处水听器表面声压的空间平均效应,给出了水听器空间平均效应修正系数的表达式和取值表。在焦点处测得的最大声压可达5.58MPa (1.02kW/cm2),用于校准0.95 ~ 1.10 MHz范围内的水听器,最大局部失真参数为0.72。结果表明了测量原理和方法的合理性和可行性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信