{"title":"A CMOS detector system for fluorescent bio-sensing application","authors":"Nan Liu, Zhiliang Hong, Ran Liu","doi":"10.1109/ASSCC.2008.4708795","DOIUrl":null,"url":null,"abstract":"This system integrates a CMOS process compatible photodiode, a low noise capacitive trans-impedance amplifier (CTIA) and a 12-bit pipelined analog-to-digital converter (ADC). The chip fabricated in a 0.18-mum standard CMOS technology occupies 3 MMZ and consumes 37 mW. Experimental results show that the Nwell/Psub photodiode used in the fluorescent detecting experiment has a sensitivity of 0.1 A/W at 515 nm and a dark current of 3 nA/cm2. The maximum differential and integral nonlinearity of the designed ADC are +0.8 LSB and -3 LSB, respectively. With a solution volume of 0.5 mul, the detector system is able to detect the fluorescein solution concentration as low as 625 ng/ml. The minimum detectable fluorescent intensity and photocurrent are 18 nW/cm2 and 36 fA, respectively.","PeriodicalId":143173,"journal":{"name":"2008 IEEE Asian Solid-State Circuits Conference","volume":"70 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2008-12-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"7","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2008 IEEE Asian Solid-State Circuits Conference","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/ASSCC.2008.4708795","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 7
Abstract
This system integrates a CMOS process compatible photodiode, a low noise capacitive trans-impedance amplifier (CTIA) and a 12-bit pipelined analog-to-digital converter (ADC). The chip fabricated in a 0.18-mum standard CMOS technology occupies 3 MMZ and consumes 37 mW. Experimental results show that the Nwell/Psub photodiode used in the fluorescent detecting experiment has a sensitivity of 0.1 A/W at 515 nm and a dark current of 3 nA/cm2. The maximum differential and integral nonlinearity of the designed ADC are +0.8 LSB and -3 LSB, respectively. With a solution volume of 0.5 mul, the detector system is able to detect the fluorescein solution concentration as low as 625 ng/ml. The minimum detectable fluorescent intensity and photocurrent are 18 nW/cm2 and 36 fA, respectively.