{"title":"A novel limit alarmer for reliable intensive care monitoring","authors":"A. Makivirta, A. Kari, E. Koski, T. Sukuvaara","doi":"10.1109/CIC.1989.130530","DOIUrl":null,"url":null,"abstract":"A median filter is combined with narrow adjustable monitoring limits and wide physiologically determined alarm limits. This ensures both quick alarming in the case of a catastrophic failure and reliable therapy monitoring with alarms. Results of the clinical assessment of this dual-limit alarmer are provided. Results of the clinical evaluation indicate a significant reduction in the frequency of false alarm (63% per patient hour).<<ETX>>","PeriodicalId":161494,"journal":{"name":"[1989] Proceedings. Computers in Cardiology","volume":"246 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1989-09-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"2","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"[1989] Proceedings. Computers in Cardiology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/CIC.1989.130530","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 2
Abstract
A median filter is combined with narrow adjustable monitoring limits and wide physiologically determined alarm limits. This ensures both quick alarming in the case of a catastrophic failure and reliable therapy monitoring with alarms. Results of the clinical assessment of this dual-limit alarmer are provided. Results of the clinical evaluation indicate a significant reduction in the frequency of false alarm (63% per patient hour).<>