High frame rate synthetic aperture vector flow imaging for transthoracic echocardiography

C. A. Villagómez-Hoyos, M. Stuart, Thor Bechsgaard, M. Nielsen, J. Jensen
{"title":"High frame rate synthetic aperture vector flow imaging for transthoracic echocardiography","authors":"C. A. Villagómez-Hoyos, M. Stuart, Thor Bechsgaard, M. Nielsen, J. Jensen","doi":"10.1117/12.2216707","DOIUrl":null,"url":null,"abstract":"This work presents the first in vivo results of 2-D high frame rate vector velocity imaging for transthoracic cardiac imaging. Measurements are made on a healthy volunteer using the SARUS experimental ultrasound scanner connected to an intercostal phased-array probe. Two parasternal long-axis view (PLAX) are obtained, one centred at the aortic valve and another centred at the left ventricle. The acquisition sequence was composed of 3 diverging waves for high frame rate synthetic aperture flow imaging. For verification a phantom measurement is performed on a transverse straight 5 mm diameter vessel at a depth of 100 mm in a tissue-mimicking phantom. A flow pump produced a 2 ml/s constant flow with a peak velocity of 0.2 m/s. The average estimated flow angle in the ROI was 86.22° ± 6.66° with a true flow angle of 90°. A relative velocity bias of −39% with a standard deviation of 13% was found. In-vivo acquisitions show complex flow patterns in the heart. In the aortic valve view, blood is seen exiting the left ventricle cavity through the aortic valve into the aorta during the systolic phase of the cardiac cycle. In the left ventricle view, blood flow is seen entering the left ventricle cavity through the mitral valve and splitting in two ways when approximating the left ventricle wall. The work presents 2-D velocity estimates on the heart from a non-invasive transthoracic scan. The ability of the method detecting flow regardless of the beam angle could potentially reveal a more complete view of the flow patterns presented on the heart.","PeriodicalId":228011,"journal":{"name":"SPIE Medical Imaging","volume":"45 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2016-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"7","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"SPIE Medical Imaging","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1117/12.2216707","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 7

Abstract

This work presents the first in vivo results of 2-D high frame rate vector velocity imaging for transthoracic cardiac imaging. Measurements are made on a healthy volunteer using the SARUS experimental ultrasound scanner connected to an intercostal phased-array probe. Two parasternal long-axis view (PLAX) are obtained, one centred at the aortic valve and another centred at the left ventricle. The acquisition sequence was composed of 3 diverging waves for high frame rate synthetic aperture flow imaging. For verification a phantom measurement is performed on a transverse straight 5 mm diameter vessel at a depth of 100 mm in a tissue-mimicking phantom. A flow pump produced a 2 ml/s constant flow with a peak velocity of 0.2 m/s. The average estimated flow angle in the ROI was 86.22° ± 6.66° with a true flow angle of 90°. A relative velocity bias of −39% with a standard deviation of 13% was found. In-vivo acquisitions show complex flow patterns in the heart. In the aortic valve view, blood is seen exiting the left ventricle cavity through the aortic valve into the aorta during the systolic phase of the cardiac cycle. In the left ventricle view, blood flow is seen entering the left ventricle cavity through the mitral valve and splitting in two ways when approximating the left ventricle wall. The work presents 2-D velocity estimates on the heart from a non-invasive transthoracic scan. The ability of the method detecting flow regardless of the beam angle could potentially reveal a more complete view of the flow patterns presented on the heart.
用于经胸超声心动图的高帧率合成孔径矢量流成像
这项工作提出了第一个体内结果的二维高帧率矢量速度成像经胸心脏成像。使用SARUS实验超声扫描仪连接肋间相控阵探针对健康志愿者进行测量。获得两个胸骨旁长轴视图(PLAX),一个以主动脉瓣为中心,另一个以左心室为中心。采集序列由3个发散波组成,用于高帧率合成孔径流成像。为了验证,在组织模拟模型中,在深度为100mm的横向直5mm容器上进行了模拟测量。流量泵产生2ml /s的恒定流量,峰值速度为0.2 m/s。ROI内流动角的估计平均值为86.22°±6.66°,真实流动角为90°。相对速度偏差为- 39%,标准差为13%。活体获取显示心脏内复杂的血流模式。在心脏周期的收缩期,主动脉瓣图显示血液从左心室腔通过主动脉瓣进入主动脉。左心室图,血流通过二尖瓣进入左心室腔,并在接近左心室壁时分成两路。这项工作提出了通过非侵入性经胸扫描对心脏的二维速度估计。无论光束角度如何,该方法检测血流的能力可能潜在地揭示心脏上血流模式的更完整视图。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信