A. Kandori, K. Tsukada, H. Horigome, M. Asaka, S. Shigemitsu, M. Takahashi, Y. Terada, T. Kubo, Akira Matsui, T. Mitsui
{"title":"Multichannel SQUID system for detecting fetal magnetocardiograms","authors":"A. Kandori, K. Tsukada, H. Horigome, M. Asaka, S. Shigemitsu, M. Takahashi, Y. Terada, T. Kubo, Akira Matsui, T. Mitsui","doi":"10.1109/ICBEM.1998.666429","DOIUrl":null,"url":null,"abstract":"A nine-channel SQUID magnetometer system is developed to detect weak magnetic fields generated by fetal cardiac activity. The nine 20-mm-diameter axial first-order gradiometers are mounted at intervals of 30 mm. A baseline length of 60 mm is applied to the pickup coil to measure the weak magnetic field from a fetal heart in a deep position. The SQUID is operated with a flux-locked loop (FLL) circuit and the FLL output is recorded at 1 kHz by a computer. Fetal magneto-cardiograms (FMCG) were measured for 109 fetuses of 20 to 39 weeks' gestation. In most cases, maternal MCG interference was negligible. R-wave peaks were detected in 81% of all subjects. The results of FMCG measurements and Fourier analysis of fetal the R-R interval are shown in this paper.","PeriodicalId":213764,"journal":{"name":"Proceedings of the 2nd International Conference on Bioelectromagnetism (Cat. No.98TH8269)","volume":"40 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1998-02-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"8","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Proceedings of the 2nd International Conference on Bioelectromagnetism (Cat. No.98TH8269)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/ICBEM.1998.666429","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 8
Abstract
A nine-channel SQUID magnetometer system is developed to detect weak magnetic fields generated by fetal cardiac activity. The nine 20-mm-diameter axial first-order gradiometers are mounted at intervals of 30 mm. A baseline length of 60 mm is applied to the pickup coil to measure the weak magnetic field from a fetal heart in a deep position. The SQUID is operated with a flux-locked loop (FLL) circuit and the FLL output is recorded at 1 kHz by a computer. Fetal magneto-cardiograms (FMCG) were measured for 109 fetuses of 20 to 39 weeks' gestation. In most cases, maternal MCG interference was negligible. R-wave peaks were detected in 81% of all subjects. The results of FMCG measurements and Fourier analysis of fetal the R-R interval are shown in this paper.