{"title":"Calibration Standard for Impulse Energy of Defibrillator and Defibrillator Analyzer","authors":"Wei Zhao, H. Shao, Huanghui Zhang","doi":"10.1109/MeMeA49120.2020.9137277","DOIUrl":null,"url":null,"abstract":"Impulse energy is one of the most important parameters that need to be calibrated for defibrillator and defibrillator analyzer. The measurement accuracy of commercial defibrillator analyzer could reach ± 1 % of reading plus 0.1 Joule nowadays. So, it placed a greater demand on the measurement standard used for the calibration. In the National Institute of Metrology (NIM) China, we had developed a set of national measurement standard. The calibration uncertainty on the defibrillator analyzer is evaluated to be better than 0.06% of reading plus 0.06 Joule. In this manuscript, the components of the system and the uncertainty evaluation of the measurement standard is introduced.","PeriodicalId":152478,"journal":{"name":"2020 IEEE International Symposium on Medical Measurements and Applications (MeMeA)","volume":"31 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2020-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2020 IEEE International Symposium on Medical Measurements and Applications (MeMeA)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/MeMeA49120.2020.9137277","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1
Abstract
Impulse energy is one of the most important parameters that need to be calibrated for defibrillator and defibrillator analyzer. The measurement accuracy of commercial defibrillator analyzer could reach ± 1 % of reading plus 0.1 Joule nowadays. So, it placed a greater demand on the measurement standard used for the calibration. In the National Institute of Metrology (NIM) China, we had developed a set of national measurement standard. The calibration uncertainty on the defibrillator analyzer is evaluated to be better than 0.06% of reading plus 0.06 Joule. In this manuscript, the components of the system and the uncertainty evaluation of the measurement standard is introduced.