Uji Kesesuaian Berkas Cahaya Kolimasi pada Pesawat Sinar-X di Politeknik Al-Islam Bandung

Oktarina Damayanti, Indra Gunawan
{"title":"Uji Kesesuaian Berkas Cahaya Kolimasi pada Pesawat Sinar-X di Politeknik Al-Islam Bandung","authors":"Oktarina Damayanti, Indra Gunawan","doi":"10.31983/JIMED.V7I2.6596","DOIUrl":null,"url":null,"abstract":"Background : Suitability test for light collimation are tests to ensure that x-ray aircraft meet radiation safety requirements and provide precise and accurate diagnostic or diagnostic information. Suitability test for light collimation is carried out on x-ray aircraft at Al-Islam Bandung Politeknik.Methods: The research method used in quantitative experimental method. The suitability test was carried out with a collimator test tool and beam alignment test tool placed on the tape then performed an experiment using an exposure factor of 75 Kv, 200 mA and 8 mAs with FFD 100 cm.Results: Based on the tests that have been carried out, it is known that there is a discrepancy in the width of the collimation beam. The large discrepancy between the light field area and the radiation field area, that is, the edge X₁ 0.3 cm, the edge X₂ 0.5 cm, the edge Y₁ 0.5 cm, the edge Y₂ 0.1 cm. So the value of ∆X is 0.8% FFD, the value of ∆Y is 0.6% FFD, and the value of ∆X + ∆Y is 1.4% FFD. While the deviation of the central point is less or right than 3⁰. These results indicate that the discrepancy in the width of the light beam collimation and center point deviations is still within the tolerance limits.Conclusions: There is a discrepancy between the X-ray light field and the collimated light field, but it is still within the limit of the test value to the passing value of the predefined test. ","PeriodicalId":205622,"journal":{"name":"Jurnal Imejing Diagnostik (JImeD)","volume":"238 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2021-07-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Jurnal Imejing Diagnostik (JImeD)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.31983/JIMED.V7I2.6596","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Background : Suitability test for light collimation are tests to ensure that x-ray aircraft meet radiation safety requirements and provide precise and accurate diagnostic or diagnostic information. Suitability test for light collimation is carried out on x-ray aircraft at Al-Islam Bandung Politeknik.Methods: The research method used in quantitative experimental method. The suitability test was carried out with a collimator test tool and beam alignment test tool placed on the tape then performed an experiment using an exposure factor of 75 Kv, 200 mA and 8 mAs with FFD 100 cm.Results: Based on the tests that have been carried out, it is known that there is a discrepancy in the width of the collimation beam. The large discrepancy between the light field area and the radiation field area, that is, the edge X₁ 0.3 cm, the edge X₂ 0.5 cm, the edge Y₁ 0.5 cm, the edge Y₂ 0.1 cm. So the value of ∆X is 0.8% FFD, the value of ∆Y is 0.6% FFD, and the value of ∆X + ∆Y is 1.4% FFD. While the deviation of the central point is less or right than 3⁰. These results indicate that the discrepancy in the width of the light beam collimation and center point deviations is still within the tolerance limits.Conclusions: There is a discrepancy between the X-ray light field and the collimated light field, but it is still within the limit of the test value to the passing value of the predefined test. 
背景:光准直适用性测试是确保x射线飞机满足辐射安全要求并提供精确和准确的诊断或诊断信息的测试。在万隆机场的x射线机上进行了光准直的适用性测试。方法:研究方法采用定量实验方法。在胶带上放置准直仪测试工具和光束对准测试工具进行适用性测试,然后在75 Kv, 200 mA和8 mA的暴露系数下进行实验,FFD为100 cm。结果:根据已进行的测试,已知准直光束的宽度存在差异。光场面积与辐射场面积差异大,即边X₁0.3 cm,边X₂0.5 cm,边Y₁0.5 cm,边Y₂0.1 cm。因此,∆X的值为0.8% FFD,∆Y的值为0.6% FFD,∆X +∆Y的值为1.4% FFD。而中心点的偏差小于或小于3⁰。这些结果表明,光束准直宽度和中心点偏差的差异仍然在公差范围内。结论:x射线光场与准直光场存在差异,但仍在测试值与预先设定的测试通过值的范围内。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信