Using the BBraun BSmartTM Pressure Manometer to Prevent Unsafe Injection Pressures During Simulated Peripheral Nerve Blockade: A Pilot Study

Rebecca-Lea Smith, S. West, Jason Wilson
{"title":"Using the BBraun BSmartTM Pressure Manometer to Prevent Unsafe Injection Pressures During Simulated Peripheral Nerve Blockade: A Pilot Study","authors":"Rebecca-Lea Smith, S. West, Jason Wilson","doi":"10.2174/2589645802115010049","DOIUrl":null,"url":null,"abstract":"\n \n Peripheral nerve injury during regional anaesthesia may result from accidental intraneural placement of the needle, or forceful needle nerve contact. Intraneural injections are associated with increased resistance to injection, typically >15 psi. The BBraun BSmart™ is an inline mechanical manometer, offering a visual display of injection pressures.\n \n \n \n The primary objective of this study was to determine if using the BBraun BSmartTM manometer successfully prevents 90% of anaesthetists and anaesthetic assistants from injecting at pressures > 15 psi during simulated nerve block.\n \n \n \n This was a prospective observational study involving anaesthetists and anaesthetic assistants. Two 20 ml injections were performed by each participant, once when the BBraun BSmartTM manometer was obscured from view, and once with the manometer visible. A PendoTech PressureMATTMS recorded injection pressures.\n \n \n \n 39 participants completed the study, with a total of 78 injections recorded. During the study, 32 peak pressures during the 78 procedures were recorded above the recommended upper limit of 15 psi, 41% of the total injections. The peak pressure rose above 15 psi in 24/39 (62%) injections when the BBraun Bsmart™ manometer was obscured, but only in 8/39 (21%) injections when the manometer was visible.\n \n \n \n The BBraun Bsmart™ manometer did not successfully prevent 90% of anaesthetists or anaesthetic assistants from injecting at unsafe pressures. However, using the BBraun BSmart™ did reduce the number of unsafe injection pressures generated by participants. When utilised in conjunction with PNS and ultrasound guidance, this may offer additional safety during peripheral nerve blockade.\n","PeriodicalId":272376,"journal":{"name":"The Open Anesthesiology Journal","volume":"6 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2021-12-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"The Open Anesthesiology Journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2174/2589645802115010049","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Peripheral nerve injury during regional anaesthesia may result from accidental intraneural placement of the needle, or forceful needle nerve contact. Intraneural injections are associated with increased resistance to injection, typically >15 psi. The BBraun BSmart™ is an inline mechanical manometer, offering a visual display of injection pressures. The primary objective of this study was to determine if using the BBraun BSmartTM manometer successfully prevents 90% of anaesthetists and anaesthetic assistants from injecting at pressures > 15 psi during simulated nerve block. This was a prospective observational study involving anaesthetists and anaesthetic assistants. Two 20 ml injections were performed by each participant, once when the BBraun BSmartTM manometer was obscured from view, and once with the manometer visible. A PendoTech PressureMATTMS recorded injection pressures. 39 participants completed the study, with a total of 78 injections recorded. During the study, 32 peak pressures during the 78 procedures were recorded above the recommended upper limit of 15 psi, 41% of the total injections. The peak pressure rose above 15 psi in 24/39 (62%) injections when the BBraun Bsmart™ manometer was obscured, but only in 8/39 (21%) injections when the manometer was visible. The BBraun Bsmart™ manometer did not successfully prevent 90% of anaesthetists or anaesthetic assistants from injecting at unsafe pressures. However, using the BBraun BSmart™ did reduce the number of unsafe injection pressures generated by participants. When utilised in conjunction with PNS and ultrasound guidance, this may offer additional safety during peripheral nerve blockade.
使用BBraun BSmartTM压力计在模拟周围神经阻滞期间防止不安全注射压力:一项试点研究
区域麻醉时周围神经损伤可能是由于意外的神经内针头放置或强力的针神经接触造成的。神经内注射与注射阻力增加有关,通常> 15psi。BBraun BSmart™是一款内联机械压力计,提供喷射压力的视觉显示。本研究的主要目的是确定使用BBraun BSmartTM压力计是否能成功防止90%的麻醉师和麻醉助手在模拟神经阻滞期间注射压力> 15 psi。这是一项涉及麻醉师和麻醉师助理的前瞻性观察研究。每位参与者进行两次20毫升注射,一次是在BBraun BSmartTM压力计从视野中遮挡时,另一次是在压力计可见时。PendoTech PressureMATTMS记录了注入压力。39名参与者完成了这项研究,总共记录了78次注射。在研究过程中,78次操作中有32次的峰值压力超过了15psi的推荐上限,占总注入量的41%。当BBraun Bsmart™压力计被遮挡时,在24/39(62%)次注入中,峰值压力高于15 psi,而在8/39(21%)次注入中,当压力计可见时,峰值压力高于15 psi。BBraun Bsmart™血压计不能成功防止90%的麻醉师或麻醉助理在不安全的压力下进行注射。然而,使用BBraun BSmart™确实减少了参与者产生的不安全注射压力的数量。当与PNS和超声引导结合使用时,这可能会在周围神经阻滞期间提供额外的安全性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信