{"title":"[Clinical experience with the cage electrode EMT 588 K. Threshold--current consumption--postoperative complications (author's transl)].","authors":"H W Präuer,&nbsp;M Lampadius","doi":"10.1055/s-0028-1096594","DOIUrl":null,"url":null,"abstract":"<p><p>The Lagergren cage electrode Elema EMT 588 K was implanted in 26 male and 21 female patients for permanent cardiac stimulation. 29 of the 47 cage electrodes were fitted with a platinum and 18 with an Elgiloy head. Acute threshold in the platinum group was 0,52 +/- 0,15 volts (x +/- s), the equivalent electrode resistance being 565 +/- 135 ohms. The threshold mean value for the Elgiloy cage electrodes was shown to be 0,36 +/- 0,11 volts, while the resistance values measured respectively 410 +/- 44 and 418 +/- 23 ohms depending on the method of measurement. In the postoperative period (up to 16 months) the following complications were observed: 3 dislocations (6,38%) and one case of rise in threshold (2,12%). The total correction rate (8,5%) corresponds almost exactly to that recorded by the author in a personal series of 400 electrode implantations of various designs. The cage electrode offers no advantages in comparison to other small surface electrodes, however, the relatively higher current consumption caused by low total electrode resistance must be regarded as disadvantages. Experience with the cage electrode has shown that the transitional resistance depends not only on the surface, but also essentially on the material and shape of the electrode tip.</p>","PeriodicalId":22981,"journal":{"name":"Thoraxchirurgie, vaskulare Chirurgie","volume":"26 1","pages":"39-43"},"PeriodicalIF":0.0000,"publicationDate":"1978-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1055/s-0028-1096594","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Thoraxchirurgie, vaskulare Chirurgie","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1055/s-0028-1096594","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

The Lagergren cage electrode Elema EMT 588 K was implanted in 26 male and 21 female patients for permanent cardiac stimulation. 29 of the 47 cage electrodes were fitted with a platinum and 18 with an Elgiloy head. Acute threshold in the platinum group was 0,52 +/- 0,15 volts (x +/- s), the equivalent electrode resistance being 565 +/- 135 ohms. The threshold mean value for the Elgiloy cage electrodes was shown to be 0,36 +/- 0,11 volts, while the resistance values measured respectively 410 +/- 44 and 418 +/- 23 ohms depending on the method of measurement. In the postoperative period (up to 16 months) the following complications were observed: 3 dislocations (6,38%) and one case of rise in threshold (2,12%). The total correction rate (8,5%) corresponds almost exactly to that recorded by the author in a personal series of 400 electrode implantations of various designs. The cage electrode offers no advantages in comparison to other small surface electrodes, however, the relatively higher current consumption caused by low total electrode resistance must be regarded as disadvantages. Experience with the cage electrode has shown that the transitional resistance depends not only on the surface, but also essentially on the material and shape of the electrode tip.

[临床经验]笼式电极EMT 588 K。阈值-电流消耗-术后并发症[作者简介]。
将Elema EMT 588 K拉格伦笼电极植入26例男性患者和21例女性患者进行永久性心脏刺激。47个笼状电极中有29个配有铂电极,18个配有Elgiloy电极头。铂组急性阈值为0.52 +/- 0.15伏(x +/- s),等效电极电阻为565 +/- 135欧姆。Elgiloy笼电极的阈值平均值显示为0,36 +/- 0,11伏,而电阻值分别测量为410 +/- 44和418 +/- 23欧姆,具体取决于测量方法。术后(16个月)出现以下并发症:脱位3例(6.38%),门限升高1例(2.12%)。总校正率(8.5%)几乎完全符合作者在各种设计的400个电极植入的个人系列中记录的。笼状电极与其他小表面电极相比没有优势,但由于电极总电阻低而导致的相对较高的电流消耗必须被视为缺点。笼状电极的经验表明,过渡电阻不仅取决于表面,而且本质上取决于电极尖端的材料和形状。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信