Calibration procedures for a 10.6-μm laser calorimeter

M. Kennedy, D. Ristau, G. Dumitru, D. Sporea, Clementina A. Timus
{"title":"Calibration procedures for a 10.6-μm laser calorimeter","authors":"M. Kennedy, D. Ristau, G. Dumitru, D. Sporea, Clementina A. Timus","doi":"10.1117/12.312718","DOIUrl":null,"url":null,"abstract":"The paper presents the theoretical background of laser calorimetry as a method to investigate the absorptance of optical components, according to the recently adopted ISO Standard 11 551. An experimental set-up for optical absorption measurements, including a cw CO2 laser and a calorimetric chamber, is presented. The calibration of the laser calorimetric set-up is also described. The calibration involved several steps: laser power calibration, evaluation of chamber sensitivity, electrical calibration using electrically heated samples and optical calibration utilizing 100% absorptance samples. There were measured the conversion factors for the laser power and for the sample temperature; there were calculated over-all correction factors for the different substrates and they could be placed within +/- 4% around their mean value.","PeriodicalId":383583,"journal":{"name":"ROMOPTO International Conference on Micro- to Nano- Photonics III","volume":"114 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1998-07-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"ROMOPTO International Conference on Micro- to Nano- Photonics III","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1117/12.312718","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

Abstract

The paper presents the theoretical background of laser calorimetry as a method to investigate the absorptance of optical components, according to the recently adopted ISO Standard 11 551. An experimental set-up for optical absorption measurements, including a cw CO2 laser and a calorimetric chamber, is presented. The calibration of the laser calorimetric set-up is also described. The calibration involved several steps: laser power calibration, evaluation of chamber sensitivity, electrical calibration using electrically heated samples and optical calibration utilizing 100% absorptance samples. There were measured the conversion factors for the laser power and for the sample temperature; there were calculated over-all correction factors for the different substrates and they could be placed within +/- 4% around their mean value.
10.6 μm激光量热计校准程序
本文介绍了激光量热法作为一种研究光学元件吸光度的方法的理论背景,该方法是根据最近采用的ISO标准11551。介绍了一种光学吸收测量的实验装置,包括一个连续波CO2激光器和一个量热室。还介绍了激光量热装置的标定。校准包括几个步骤:激光功率校准,腔室灵敏度评估,使用电加热样品的电校准和使用100%吸收样品的光学校准。测量了激光功率和样品温度的转换系数;计算了不同底物的总体校正因子,它们可以在平均值周围的+/- 4%范围内。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信