A biomechanical study of 150/spl deg/ vs. 135/spl deg/ hip screws in femoral neck fractures

G. Lynn, D. P. Mukherjee, R. N. Kruse, K. Sadasivan*, J. Albright
{"title":"A biomechanical study of 150/spl deg/ vs. 135/spl deg/ hip screws in femoral neck fractures","authors":"G. Lynn, D. P. Mukherjee, R. N. Kruse, K. Sadasivan*, J. Albright","doi":"10.1109/SBEC.1995.514427","DOIUrl":null,"url":null,"abstract":"Femoral neck fractures have been difficult to treat with nonunion and avascular necrosis reported in 10% to 40% of cases. The hypothesis tested here was to determine whether a femoral neck fracture fixed with a 150/spl deg/ screw is more stable than that fixed with a 135/spl deg/ screw. The objectives were to measure axial stiffness, torsional stiffness, and failure strength of femoral neck fractures (Pauwels III). Also, CT derived bone mineral density (BMD in units of K/sub 2/KPO/sub 4/ mg/cc equivalent) was measured. The following results were obtained: 1) The 150/spl deg/ hip screw demonstrated greater axial stiffness, as well as a preferred mode of failure (impaction at the fracture site). This is less likely to lead to nonunion or avascular necrosis than is the shearing action at the fracture site with the 135/spl deg/ hip screw. 2) The 135/spl deg/ hip screw demonstrated greater resistance to rotation. 3) CT derived BMD did not correlate with either axial or torsional stiffness of intact or fixed bones.","PeriodicalId":332563,"journal":{"name":"Proceedings of the 1995 Fourteenth Southern Biomedical Engineering Conference","volume":"5 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1995-04-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Proceedings of the 1995 Fourteenth Southern Biomedical Engineering Conference","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/SBEC.1995.514427","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Femoral neck fractures have been difficult to treat with nonunion and avascular necrosis reported in 10% to 40% of cases. The hypothesis tested here was to determine whether a femoral neck fracture fixed with a 150/spl deg/ screw is more stable than that fixed with a 135/spl deg/ screw. The objectives were to measure axial stiffness, torsional stiffness, and failure strength of femoral neck fractures (Pauwels III). Also, CT derived bone mineral density (BMD in units of K/sub 2/KPO/sub 4/ mg/cc equivalent) was measured. The following results were obtained: 1) The 150/spl deg/ hip screw demonstrated greater axial stiffness, as well as a preferred mode of failure (impaction at the fracture site). This is less likely to lead to nonunion or avascular necrosis than is the shearing action at the fracture site with the 135/spl deg/ hip screw. 2) The 135/spl deg/ hip screw demonstrated greater resistance to rotation. 3) CT derived BMD did not correlate with either axial or torsional stiffness of intact or fixed bones.
股骨颈骨折150/spl°/与135/spl°/髋螺钉的生物力学研究
股骨颈骨折难以治疗,有10% - 40%的病例伴有骨不连和无血管坏死。本研究检验的假设是确定150/spl°螺钉固定股骨颈骨折是否比135/spl°螺钉固定股骨颈骨折更稳定。目的是测量股骨颈骨折的轴向刚度、扭转刚度和破坏强度(Pauwels III)。此外,还测量了CT衍生的骨矿物质密度(BMD,单位为K/sub 2/KPO/sub 4/ mg/cc当量)。得到以下结果:1)150/spl度/髋关节螺钉具有更大的轴向刚度,以及首选的失效模式(在骨折部位撞击)。与135/spl度/髋螺钉在骨折部位的剪切作用相比,这更不容易导致骨不连或缺血性坏死。2) 135/spl度/髋关节螺钉表现出更大的旋转阻力。3) CT得出的骨密度与完整或固定骨的轴向或扭转刚度无关。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信