{"title":"An implementation of motion artifacts elimination for PPG signal processing based on recursive least squares adaptive filter","authors":"Chih-Chin Wu, I-Wei Chen, W. Fang","doi":"10.1109/BIOCAS.2017.8325141","DOIUrl":null,"url":null,"abstract":"In Photoplethysmographic (PPG) signals analysis, the accuracy and stability are highly affected by Motion Artifacts (MAs) disturbances. In this paper, we adopt an adaptive and efficient approach based on the developed DC Remover method and Recursive Least Squares (RLS) adaptive filter for reducing MAs from PPG signals in real time. The experimental results of this work show a high correlation coefficient between Electrocardiography (ECG)-derived heart rate and PPG-derived heart rate, which is higher than 0.8504 of the R value, a high agreement by Bland-Altman analysis in the limits of agreement represent the 95% confidence interval and the standard deviation is 3.81 BPM (Beats Per Minutes). An overall PPG signal with higher signal quality is obtained. Further, the precision of heart rate calculated by PPG is improved.","PeriodicalId":361477,"journal":{"name":"2017 IEEE Biomedical Circuits and Systems Conference (BioCAS)","volume":"47 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2017-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"16","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2017 IEEE Biomedical Circuits and Systems Conference (BioCAS)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/BIOCAS.2017.8325141","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 16
Abstract
In Photoplethysmographic (PPG) signals analysis, the accuracy and stability are highly affected by Motion Artifacts (MAs) disturbances. In this paper, we adopt an adaptive and efficient approach based on the developed DC Remover method and Recursive Least Squares (RLS) adaptive filter for reducing MAs from PPG signals in real time. The experimental results of this work show a high correlation coefficient between Electrocardiography (ECG)-derived heart rate and PPG-derived heart rate, which is higher than 0.8504 of the R value, a high agreement by Bland-Altman analysis in the limits of agreement represent the 95% confidence interval and the standard deviation is 3.81 BPM (Beats Per Minutes). An overall PPG signal with higher signal quality is obtained. Further, the precision of heart rate calculated by PPG is improved.
在光容积脉搏波(PPG)信号分析中,运动伪影(MAs)干扰严重影响信号的准确性和稳定性。在本文中,我们采用一种基于所开发的DC去除方法和递归最小二乘(RLS)自适应滤波器的自适应高效方法来实时减少PPG信号中的MAs。本工作的实验结果表明,心电图(ECG)衍生心率与ppg衍生心率之间具有较高的相关系数,高于R值的0.8504,在一致性限中Bland-Altman分析具有较高的一致性,代表95%的置信区间,标准差为3.81 BPM (Beats Per Minutes)。得到了具有较高信号质量的整体PPG信号。进一步提高了PPG计算心率的精度。