Simplified Solution to Eliminating Electrical Noise During Cardiac Surgery

Nicklett Johnston, M. Jessen, M. Zeidler
{"title":"Simplified Solution to Eliminating Electrical Noise During Cardiac Surgery","authors":"Nicklett Johnston, M. Jessen, M. Zeidler","doi":"10.1051/ject/1996284205","DOIUrl":null,"url":null,"abstract":"Electrical noise will be present in the operating room even under ideal circumstances. The addition of the cardiopulmonary bypass machine to the other machinery in the operating room introduces one more source of electrical noise. Identifying and managing this interference may aid in reducing artifactual signals on the electrocardiogram (EKG) monitor.\nIf electrical interference occurs during the onset of cardiopulmonary bypass, an uninterpretable rhythm pattern on the electrocardiogram monitor may be present.\nAdding an extra ground from the main pump head to the heater/cooler helps to reduce noise and, therefore, effectively diminishes pump-generated artifact on the EKG.","PeriodicalId":309024,"journal":{"name":"The Journal of ExtraCorporeal Technology","volume":"141 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1996-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"The Journal of ExtraCorporeal Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1051/ject/1996284205","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

Abstract

Electrical noise will be present in the operating room even under ideal circumstances. The addition of the cardiopulmonary bypass machine to the other machinery in the operating room introduces one more source of electrical noise. Identifying and managing this interference may aid in reducing artifactual signals on the electrocardiogram (EKG) monitor. If electrical interference occurs during the onset of cardiopulmonary bypass, an uninterpretable rhythm pattern on the electrocardiogram monitor may be present. Adding an extra ground from the main pump head to the heater/cooler helps to reduce noise and, therefore, effectively diminishes pump-generated artifact on the EKG.
心脏手术中消除电噪声的简化解决方案
即使在理想的情况下,手术室也会有电气噪声。在手术室的其他机器中增加体外循环机器,又增加了一个电噪声源。识别和处理这种干扰可能有助于减少在心电图(EKG)监视器上的人为信号。如果在体外循环开始时发生电干扰,心电图监护仪上可能出现无法解释的节律模式。在主泵头和加热器/冷却器之间增加一个额外的接地,有助于降低噪音,从而有效地减少泵产生的心电图伪影。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信