Novel method of automatic auscultation for blood pressure measurement using pulses in cuff pressure and korotkoff sound

D. Park, H. Oh, J.H. Kang, I.Y. Kim, Y. Chee, J.S. Lee
{"title":"Novel method of automatic auscultation for blood pressure measurement using pulses in cuff pressure and korotkoff sound","authors":"D. Park, H. Oh, J.H. Kang, I.Y. Kim, Y. Chee, J.S. Lee","doi":"10.1109/CIC.2008.4749007","DOIUrl":null,"url":null,"abstract":"In the NIBP (Non-Invasive Blood Pressure) measurement, we need well trained observer and there are differences in reading values between observers. We suggest new automated auscultation algorithms to select the proper epoch of K-sound for each heart beat. The sound for each epoch is converted into frequency domain and total power was calculated. We decided SBP(Systolic Blood Pressure) point when there was abrupt jump from the baseline. DBP (Diastolic blood pressure) was decided when there was the biggest drop in total power of successive beats. For the validation of our suggested method, human observer assessed SBP/DBP according to EHS (European Hypertension Society) recommendation. The mean differences between the automatic method and traditional auscultation method were 2.0 mmHg for SBP and 3.1 mmHg for DBP. We also took the paired t-test with the two results, and there was no statistically significant difference between these two methods.","PeriodicalId":194782,"journal":{"name":"2008 Computers in Cardiology","volume":"53 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2008-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"11","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2008 Computers in Cardiology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/CIC.2008.4749007","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 11

Abstract

In the NIBP (Non-Invasive Blood Pressure) measurement, we need well trained observer and there are differences in reading values between observers. We suggest new automated auscultation algorithms to select the proper epoch of K-sound for each heart beat. The sound for each epoch is converted into frequency domain and total power was calculated. We decided SBP(Systolic Blood Pressure) point when there was abrupt jump from the baseline. DBP (Diastolic blood pressure) was decided when there was the biggest drop in total power of successive beats. For the validation of our suggested method, human observer assessed SBP/DBP according to EHS (European Hypertension Society) recommendation. The mean differences between the automatic method and traditional auscultation method were 2.0 mmHg for SBP and 3.1 mmHg for DBP. We also took the paired t-test with the two results, and there was no statistically significant difference between these two methods.
利用袖带压力脉冲和克罗特科夫声自动听诊血压测量的新方法
在NIBP(无创血压)测量中,我们需要训练有素的观察者,观察者之间的读数值存在差异。我们提出了新的自动听诊算法,为每次心跳选择合适的k音纪元。将每个历元的声音转换为频域,计算总功率。当从基线突然跳跃时,我们决定收缩压点。当连续搏动的总功率下降幅度最大时,测定舒张压。为了验证我们建议的方法,人类观察员根据EHS(欧洲高血压协会)的建议评估收缩压/舒张压。自动方法与传统听诊方法的平均差值分别为收缩压2.0 mmHg和舒张压3.1 mmHg。我们还对两种结果进行了配对t检验,两种方法的差异无统计学意义。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信