D Cathignol, J Y Chapelon, J Jossinet, B Lavandier, C Fourcade
{"title":"Detailed description of an implantable directional Doppler flowmeter.","authors":"D Cathignol, J Y Chapelon, J Jossinet, B Lavandier, C Fourcade","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>A Doppler flowmeter and the necessary modifications for implantation are described in detail. Since only part of the electronics was implanted a phase-locked loop had to be introduced in order to keep the flow measurement directional. The proper working of the apparatus is demonstrated in vitro and in vivo. As an example the result of flow studies in the aorta and the pulmonary artery after homotransplantation of the lung in dogs are given.</p>","PeriodicalId":75602,"journal":{"name":"Biotelemetry","volume":"3 2","pages":"117-28"},"PeriodicalIF":0.0000,"publicationDate":"1976-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biotelemetry","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
A Doppler flowmeter and the necessary modifications for implantation are described in detail. Since only part of the electronics was implanted a phase-locked loop had to be introduced in order to keep the flow measurement directional. The proper working of the apparatus is demonstrated in vitro and in vivo. As an example the result of flow studies in the aorta and the pulmonary artery after homotransplantation of the lung in dogs are given.