Evaluation of the Effect of Intubation Box use on Tracheal Intubation Difficulty with King Vision® and Truview Videolaryngoscope in Manikin in a Tertiary Care Hospital.
{"title":"Evaluation of the Effect of Intubation Box use on Tracheal Intubation Difficulty with King Vision<sup>®</sup> and Truview Videolaryngoscope in Manikin in a Tertiary Care Hospital.","authors":"Nazia Nazir","doi":"10.2478/rjaic-2021-0004","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The procedures of introducing an airway by intubation are associated with increased risk of aerosolisation of SARS-CoV-2 virus, posing a high risk to the personnel involved. Newer and novel methods such as the intubation box have been developed to increase the safety of healthcare workers during intubation.</p><p><strong>Methods design: </strong>In this study, 33 anaesthesiologist and critical care specialists intubated the trachea of the airway manikin (US Laerdal Medical AS™) 4 times using a King Vision<sup>®</sup> videolaryngoscope and TRUVIEW PCD™ videolaryngoscope (with and without an intubation box as described by Lai). Intubation time was primary outcome. Secondary outcomes were first-pass intubation success rate, percentage of glottic opening (POGO) score and peak force to maxillary incisors.</p><p><strong>Results: </strong>Intubation time and the number of times a click was heard during tracheal intubation were considerably higher in both groups when an intubation box was used (Table 1). When comparing the two laryngoscopes, the King Vision<sup>®</sup> videolaryngoscope enabled much less time to intubate than did the TRUVIEW laryngoscope, both with and without the intubation box. (P<0.001) In both laryngoscope groups, first-pass successful intubation was higher without the intubation box, although the difference was statistically insignificant. POGO score was not affected by intubation box but a higher score was observed with King Vision<sup>®</sup> laryngoscope (Tables 1,2).</p><p><strong>Conclusion: </strong>This study indicates that use of an intubation box makes intubation difficult and increases the time needed to perform it. King Vision<sup>®</sup> videolaryngoscope results in lesser intubation time and better glottic view as compared to TRUVIEW laryngoscope.</p>","PeriodicalId":21279,"journal":{"name":"Romanian journal of anaesthesia and intensive care","volume":"28 1","pages":"25-28"},"PeriodicalIF":0.0000,"publicationDate":"2021-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/2d/3e/rjaic-28-025.PMC9949008.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Romanian journal of anaesthesia and intensive care","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2478/rjaic-2021-0004","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0
Abstract
Background: The procedures of introducing an airway by intubation are associated with increased risk of aerosolisation of SARS-CoV-2 virus, posing a high risk to the personnel involved. Newer and novel methods such as the intubation box have been developed to increase the safety of healthcare workers during intubation.
Methods design: In this study, 33 anaesthesiologist and critical care specialists intubated the trachea of the airway manikin (US Laerdal Medical AS™) 4 times using a King Vision® videolaryngoscope and TRUVIEW PCD™ videolaryngoscope (with and without an intubation box as described by Lai). Intubation time was primary outcome. Secondary outcomes were first-pass intubation success rate, percentage of glottic opening (POGO) score and peak force to maxillary incisors.
Results: Intubation time and the number of times a click was heard during tracheal intubation were considerably higher in both groups when an intubation box was used (Table 1). When comparing the two laryngoscopes, the King Vision® videolaryngoscope enabled much less time to intubate than did the TRUVIEW laryngoscope, both with and without the intubation box. (P<0.001) In both laryngoscope groups, first-pass successful intubation was higher without the intubation box, although the difference was statistically insignificant. POGO score was not affected by intubation box but a higher score was observed with King Vision® laryngoscope (Tables 1,2).
Conclusion: This study indicates that use of an intubation box makes intubation difficult and increases the time needed to perform it. King Vision® videolaryngoscope results in lesser intubation time and better glottic view as compared to TRUVIEW laryngoscope.
期刊介绍:
The Romanian Journal of Anaesthesia and Intensive Care is the official journal of the Romanian Society of Anaesthesia and Intensive Care and has been published continuously since 1994. It is intended mainly for anaesthesia and intensive care providers, but it is also aimed at specialists in emergency medical care and in pain research and management. The Journal is indexed in Scopus, Embase, PubMed Central as well as the databases of the Romanian Ministry of Education and Research (CNCSIS) B+ category. The Journal publishes two issues per year, the first one in April and the second one in October, and contains original articles, reviews, case reports, letters to the editor, book reviews and commentaries. The Journal is distributed free of charge to the members of the Romanian Society of Anaesthesia and Intensive Care.