Refractory Hypotension During Implantation of a 70 mL Total Artificial Heart in a Patient With Pectus Excavatum: A Case Report.

IF 1.1 Q3 ANESTHESIOLOGY
Brendon Hart, Nahush A Mokadam, Karina Anam, Samiya Saklayen, Galina Dimitrova, Alix Zuleta-Alarcon, Jordan Holloway, Hamdy Awad, David Convissar, Michael Essandoh
{"title":"Refractory Hypotension During Implantation of a 70 mL Total Artificial Heart in a Patient With Pectus Excavatum: A Case Report.","authors":"Brendon Hart,&nbsp;Nahush A Mokadam,&nbsp;Karina Anam,&nbsp;Samiya Saklayen,&nbsp;Galina Dimitrova,&nbsp;Alix Zuleta-Alarcon,&nbsp;Jordan Holloway,&nbsp;Hamdy Awad,&nbsp;David Convissar,&nbsp;Michael Essandoh","doi":"10.1177/10892532231157529","DOIUrl":null,"url":null,"abstract":"<p><p>The Syncardia total artificial heart system is the only commercially approved durable device for treating biventricular heart failure patients awaiting heart transplantation. Conventionally, the Syncardia total artificial heart system is implanted based on the distance from the anterior aspect of the 10th thoracic vertebra to the sternum and the patient's body surface area. However, this criterion does not account for chest wall musculoskeletal deformities. This case report describes a patient with a pectus excavatum who developed compression of the inferior vena cava after Syncardia total artificial heart implantation and how transesophageal echocardiography guided chest wall surgery to accommodate the total artificial heart system.</p>","PeriodicalId":46500,"journal":{"name":"Seminars in Cardiothoracic and Vascular Anesthesia","volume":null,"pages":null},"PeriodicalIF":1.1000,"publicationDate":"2023-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Seminars in Cardiothoracic and Vascular Anesthesia","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1177/10892532231157529","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/2/20 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"ANESTHESIOLOGY","Score":null,"Total":0}
引用次数: 1

Abstract

The Syncardia total artificial heart system is the only commercially approved durable device for treating biventricular heart failure patients awaiting heart transplantation. Conventionally, the Syncardia total artificial heart system is implanted based on the distance from the anterior aspect of the 10th thoracic vertebra to the sternum and the patient's body surface area. However, this criterion does not account for chest wall musculoskeletal deformities. This case report describes a patient with a pectus excavatum who developed compression of the inferior vena cava after Syncardia total artificial heart implantation and how transesophageal echocardiography guided chest wall surgery to accommodate the total artificial heart system.

一例挖掘胸肌患者植入70mL全人工心脏期间的顽固性低血压:一例报告。
Syncardia全人工心脏系统是唯一一种商业批准的耐用设备,用于治疗等待心脏移植的双心室心力衰竭患者。传统上,根据从第10胸椎的前部到胸骨的距离和患者的体表面积来植入Syncardia全人工心脏系统。然而,这一标准并不适用于胸壁肌肉骨骼畸形。本病例报告描述了一名患有漏斗胸的患者,他在Syncardia全人工心脏植入术后出现下腔静脉压迫,以及经食道超声心动图如何指导胸壁手术以适应全人工心脏系统。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
3.60
自引率
14.30%
发文量
31
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信