Brendon Hart, Nahush A Mokadam, Karina Anam, Samiya Saklayen, Galina Dimitrova, Alix Zuleta-Alarcon, Jordan Holloway, Hamdy Awad, David Convissar, Michael Essandoh
{"title":"Refractory Hypotension During Implantation of a 70 mL Total Artificial Heart in a Patient With Pectus Excavatum: A Case Report.","authors":"Brendon Hart, Nahush A Mokadam, Karina Anam, Samiya Saklayen, Galina Dimitrova, Alix Zuleta-Alarcon, Jordan Holloway, Hamdy Awad, David Convissar, Michael Essandoh","doi":"10.1177/10892532231157529","DOIUrl":null,"url":null,"abstract":"<p><p>The Syncardia total artificial heart system is the only commercially approved durable device for treating biventricular heart failure patients awaiting heart transplantation. Conventionally, the Syncardia total artificial heart system is implanted based on the distance from the anterior aspect of the 10th thoracic vertebra to the sternum and the patient's body surface area. However, this criterion does not account for chest wall musculoskeletal deformities. This case report describes a patient with a pectus excavatum who developed compression of the inferior vena cava after Syncardia total artificial heart implantation and how transesophageal echocardiography guided chest wall surgery to accommodate the total artificial heart system.</p>","PeriodicalId":46500,"journal":{"name":"Seminars in Cardiothoracic and Vascular Anesthesia","volume":null,"pages":null},"PeriodicalIF":1.1000,"publicationDate":"2023-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Seminars in Cardiothoracic and Vascular Anesthesia","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1177/10892532231157529","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/2/20 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"ANESTHESIOLOGY","Score":null,"Total":0}
引用次数: 1
Abstract
The Syncardia total artificial heart system is the only commercially approved durable device for treating biventricular heart failure patients awaiting heart transplantation. Conventionally, the Syncardia total artificial heart system is implanted based on the distance from the anterior aspect of the 10th thoracic vertebra to the sternum and the patient's body surface area. However, this criterion does not account for chest wall musculoskeletal deformities. This case report describes a patient with a pectus excavatum who developed compression of the inferior vena cava after Syncardia total artificial heart implantation and how transesophageal echocardiography guided chest wall surgery to accommodate the total artificial heart system.