Andrew Tabner, Philip Bryson, Nicholas Tilbury, Benjamin McGregor, Alistair Wesson, Gareth D Hughes, Gareth R Hughes, Graham Johnson
{"title":"An evaluation of the NUI Compact Chest Compression Device (NCCD), a mechanical CPR device suitable for use in the saturation diving environment.","authors":"Andrew Tabner, Philip Bryson, Nicholas Tilbury, Benjamin McGregor, Alistair Wesson, Gareth D Hughes, Gareth R Hughes, Graham Johnson","doi":"10.28920/dhm53.3.181-188","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Provision of manual chest compressions in a diving bell using a conventional technique is often impossible, and alternative techniques are poorly evidenced in terms of efficacy and sustainability. The first mechanical cardiopulmonary resuscitation (CPR) device suitable for use in this environment, the NUI Compact Chest Compression Device (NCCD), has recently been designed and manufactured. This study assessed both the efficacy of the device in delivering chest compressions to both prone and seated manikins, and the ability of novice users to apply and operate it.</p><p><strong>Methods: </strong>Compression efficacy was assessed using a Resusi Anne QCPR intelligent manikin, and the primary outcome was the proportion of compressions delivered to target depth (50-60 mm). The gold standard was that achieved by expert CPR providers delivering manual CPR; the LUCAS 3 mCPR device was a further comparator.</p><p><strong>Results: </strong>The NCCD delivered 100% of compressions to target depth compared to 98% for the gold standard (interquartile range 1.5%) and 98% for the LUCAS 3 when applied to both supine and seated manikins. The NCCD sometimes became dislodged and had to be reapplied when used with a seated manikin.</p><p><strong>Conclusions: </strong>The NCCD can deliver chest compressions at target rate and depth to both supine and seated manikins with efficacy equivalent to manual CPR and the LUCAS 3. It can become dislodged when applied to a seated manikin; its design has now been altered to prevent this. New users can be trained in use of the NCCD quickly, but practise is required to ensure effective use.</p>","PeriodicalId":11296,"journal":{"name":"Diving and hyperbaric medicine","volume":"53 3","pages":"181-188"},"PeriodicalIF":0.8000,"publicationDate":"2023-09-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10597600/pdf/DHM-53-181.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Diving and hyperbaric medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.28920/dhm53.3.181-188","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"PUBLIC, ENVIRONMENTAL & OCCUPATIONAL HEALTH","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction: Provision of manual chest compressions in a diving bell using a conventional technique is often impossible, and alternative techniques are poorly evidenced in terms of efficacy and sustainability. The first mechanical cardiopulmonary resuscitation (CPR) device suitable for use in this environment, the NUI Compact Chest Compression Device (NCCD), has recently been designed and manufactured. This study assessed both the efficacy of the device in delivering chest compressions to both prone and seated manikins, and the ability of novice users to apply and operate it.
Methods: Compression efficacy was assessed using a Resusi Anne QCPR intelligent manikin, and the primary outcome was the proportion of compressions delivered to target depth (50-60 mm). The gold standard was that achieved by expert CPR providers delivering manual CPR; the LUCAS 3 mCPR device was a further comparator.
Results: The NCCD delivered 100% of compressions to target depth compared to 98% for the gold standard (interquartile range 1.5%) and 98% for the LUCAS 3 when applied to both supine and seated manikins. The NCCD sometimes became dislodged and had to be reapplied when used with a seated manikin.
Conclusions: The NCCD can deliver chest compressions at target rate and depth to both supine and seated manikins with efficacy equivalent to manual CPR and the LUCAS 3. It can become dislodged when applied to a seated manikin; its design has now been altered to prevent this. New users can be trained in use of the NCCD quickly, but practise is required to ensure effective use.
期刊介绍:
Diving and Hyperbaric Medicine (DHM) is the combined journal of the South Pacific Underwater Medicine Society (SPUMS) and the European Underwater and Baromedical Society (EUBS). It seeks to publish papers of high quality on all aspects of diving and hyperbaric medicine of interest to diving medical professionals, physicians of all specialties, scientists, members of the diving and hyperbaric industries, and divers. Manuscripts must be offered exclusively to Diving and Hyperbaric Medicine, unless clearly authenticated copyright exemption accompaniesthe manuscript. All manuscripts will be subject to peer review. Accepted contributions will also be subject to editing.