Comparison of 1-year Follow-up Echocardiographic Outcomes of Sapien 3 Versus Evolut R Bioprosthetic Transcatheter Aortic Valves: A Single-center Retrospective Iranian Cohort Study.

Q3 Medicine
Mohammad Sahebjam, Arezou Zoroufian, Alimohammad Hajizeynali, Mojtaba Salarifar, Arash Jalali, Aryan Ayati, Mahkameh Farmanesh
{"title":"Comparison of 1-year Follow-up Echocardiographic Outcomes of Sapien 3 Versus Evolut R Bioprosthetic Transcatheter Aortic Valves: A Single-center Retrospective Iranian Cohort Study.","authors":"Mohammad Sahebjam,&nbsp;Arezou Zoroufian,&nbsp;Alimohammad Hajizeynali,&nbsp;Mojtaba Salarifar,&nbsp;Arash Jalali,&nbsp;Aryan Ayati,&nbsp;Mahkameh Farmanesh","doi":"10.1097/HPC.0000000000000321","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>The current study aimed to compare 1-year echocardiographic outcomes of the new generations of self-expanding (Evolut R) versus balloon-expandable (Sapien 3) bioprosthetic transcatheter aortic valves.</p><p><strong>Methods: </strong>In this study, gradients and flow velocities obtained from transthoracic Doppler-echocardiography were retrospectively collected from patients who underwent 2 new generations of transcatheter aortic valve implantation interventions with Sapien 3 and Evolut R valves. Patients underwent echocardiography before the procedure and at discharge, 6 months, and 1-year follow-up.</p><p><strong>Results: </strong>Of the 66 patients, 28 received Sapien 3 and 38 received Evolut R valves. Evolut R valve presented a lower mean gradient at all follow-up time points compared with Sapien 3 valves (14.4 mm Hg, 14.9 mm Hg, 15.5 mm Hg compared with 10.1 mm Hg, 11.6 mm Hg, 11.8 mm Hg, respectively; all P -values <0.001). Small valve sizes of Evolut R, including 23 and 26, had higher echocardiographic mean gradient or peak gradient at the time of discharge compared with larger valves, including sizes 29 and 34 (11.1 mm Hg and 11.2 mm Hg vs. 10.2 mm Hg, 9.1 mm Hg) and 1-year follow-up (11.0 mm Hg, 11.0 mm Hg vs. 9.9 mm Hg, 8.4 mm Hg; all P -values = 0.001). Although Sapien 3 valves demonstrated a higher peak gradient in smaller sizes at discharge (18.44 mm Hg in size 23 vs. 17.9 mm Hg, 16.5 mm Hg in size 26 and 29, respectively; P = 0.001), the peak gradients did not show a statistically significant difference in the 1-year follow-up.</p><p><strong>Conclusions: </strong>The current study detected significantly lower mean and peak gradients in Evolut R compared with Sapien 3 at all follow-up time points. Furthermore, smaller valve sizes were associated with significantly higher gradients at all follow-ups, regardless of the valve type.</p>","PeriodicalId":35914,"journal":{"name":"Critical Pathways in Cardiology","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2023-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Critical Pathways in Cardiology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1097/HPC.0000000000000321","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: The current study aimed to compare 1-year echocardiographic outcomes of the new generations of self-expanding (Evolut R) versus balloon-expandable (Sapien 3) bioprosthetic transcatheter aortic valves.

Methods: In this study, gradients and flow velocities obtained from transthoracic Doppler-echocardiography were retrospectively collected from patients who underwent 2 new generations of transcatheter aortic valve implantation interventions with Sapien 3 and Evolut R valves. Patients underwent echocardiography before the procedure and at discharge, 6 months, and 1-year follow-up.

Results: Of the 66 patients, 28 received Sapien 3 and 38 received Evolut R valves. Evolut R valve presented a lower mean gradient at all follow-up time points compared with Sapien 3 valves (14.4 mm Hg, 14.9 mm Hg, 15.5 mm Hg compared with 10.1 mm Hg, 11.6 mm Hg, 11.8 mm Hg, respectively; all P -values <0.001). Small valve sizes of Evolut R, including 23 and 26, had higher echocardiographic mean gradient or peak gradient at the time of discharge compared with larger valves, including sizes 29 and 34 (11.1 mm Hg and 11.2 mm Hg vs. 10.2 mm Hg, 9.1 mm Hg) and 1-year follow-up (11.0 mm Hg, 11.0 mm Hg vs. 9.9 mm Hg, 8.4 mm Hg; all P -values = 0.001). Although Sapien 3 valves demonstrated a higher peak gradient in smaller sizes at discharge (18.44 mm Hg in size 23 vs. 17.9 mm Hg, 16.5 mm Hg in size 26 and 29, respectively; P = 0.001), the peak gradients did not show a statistically significant difference in the 1-year follow-up.

Conclusions: The current study detected significantly lower mean and peak gradients in Evolut R compared with Sapien 3 at all follow-up time points. Furthermore, smaller valve sizes were associated with significantly higher gradients at all follow-ups, regardless of the valve type.

Sapien 3与Evolut R生物人工经导管主动脉瓣1年随访超声心动图结果的比较:一项单中心回顾性伊朗队列研究
目的:本研究旨在比较新一代自扩式(Evolut R)与球囊可扩式(Sapien 3)经导管生物假体主动脉瓣1年的超声心动图结果。方法:本研究回顾性收集经胸多普勒超声心动图获得的梯度和血流速度,这些患者接受了2代新一代经导管主动脉瓣植入术,采用Sapien 3和Evolut R瓣膜。患者在手术前、出院时、6个月和1年随访时接受超声心动图检查。结果:66例患者中,Sapien 3瓣膜28例,Evolut R瓣膜38例。与Sapien 3瓣膜相比,Evolut R瓣膜在所有随访时间点的平均梯度较低(分别为14.4 mm Hg、14.9 mm Hg、15.5 mm Hg比10.1 mm Hg、11.6 mm Hg、11.8 mm Hg;结论:本研究发现,在所有随访时间点,Evolut R的平均梯度和峰值梯度明显低于Sapien 3。此外,在所有随访中,无论瓣膜类型如何,较小的瓣膜通径与显著较高的梯度相关。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Critical Pathways in Cardiology
Critical Pathways in Cardiology Medicine-Medicine (all)
CiteScore
1.90
自引率
0.00%
发文量
52
期刊介绍: Critical Pathways in Cardiology provides a single source for the diagnostic and therapeutic protocols in use at hospitals worldwide for patients with cardiac disorders. The Journal presents critical pathways for specific diagnoses—complete with evidence-based rationales—and also publishes studies of these protocols" effectiveness.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信