Prospective External Validation of an Algorithm Predicting Hourly Basal Insulin Infusion Rates from Characteristics of Patients with Type 1 Diabetes Treated with Insulin Pumps.
Jana S Schmelzer, Melanie Kahle-Stephan, Juris J Meier, Michael A Nauck
{"title":"Prospective External Validation of an Algorithm Predicting Hourly Basal Insulin Infusion Rates from Characteristics of Patients with Type 1 Diabetes Treated with Insulin Pumps.","authors":"Jana S Schmelzer, Melanie Kahle-Stephan, Juris J Meier, Michael A Nauck","doi":"10.1055/a-2118-2011","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>We previously published an algorithm predicting 24 h basal insulin infusion profiles in insulin pump-treated subjects with type 1 diabetes profiles from six subject characteristics. This algorithm was to be externally validated in an independent environment and patient population.</p><p><strong>Methods: </strong>Thirty-two patients with pump-treated type diabetes were switched to their individually algorithm-derived basal insulin infusion profile, and the appropriateness of fasting glycemic control was scrutinized by means of a supervised 24 h fast. Primary endpoint was appropriate fasting glycemic control according to pre-defined criteria in at least 80% of the cohort.</p><p><strong>Results: </strong>In 24 out of 32 patients switching to the algorithm-derived basal insulin infusion rate and undergoing a 24-h fasting period, appropriate glycemic control was achieved (=75%, lower than the pre-defined threshold of 80%), two patients discontinued the fast due to hyperglycemia, and six finished the fasting period, however, with inappropriate fasting glycemic control (entirely due to hyperglycemic episodes). There were no obvious differences in baseline characteristics between those with appropriate vs. inappropriate fasting glycemic control on the basal insulin infusion rate provided by the algorithm.</p><p><strong>Conclusion: </strong>In conclusion, when testing fasting glycemic control with an algorithm-derived individual basal insulin infusion profile during a 24 h fasting period in a cohort unrelated in terms of the hospital environment and catchment area, the success rate was lower than a pre-defined threshold for concluding utility of this algorithm. Therefore, applying this algorithm in order to initiate or optimize basal insulin infusion profiles in type 1 diabetes cannot be generally recommended.</p>","PeriodicalId":12241,"journal":{"name":"Experimental and Clinical Endocrinology & Diabetes","volume":" ","pages":"539-547"},"PeriodicalIF":1.6000,"publicationDate":"2023-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Experimental and Clinical Endocrinology & Diabetes","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1055/a-2118-2011","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/7/20 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"ENDOCRINOLOGY & METABOLISM","Score":null,"Total":0}
引用次数: 0
Abstract
Background: We previously published an algorithm predicting 24 h basal insulin infusion profiles in insulin pump-treated subjects with type 1 diabetes profiles from six subject characteristics. This algorithm was to be externally validated in an independent environment and patient population.
Methods: Thirty-two patients with pump-treated type diabetes were switched to their individually algorithm-derived basal insulin infusion profile, and the appropriateness of fasting glycemic control was scrutinized by means of a supervised 24 h fast. Primary endpoint was appropriate fasting glycemic control according to pre-defined criteria in at least 80% of the cohort.
Results: In 24 out of 32 patients switching to the algorithm-derived basal insulin infusion rate and undergoing a 24-h fasting period, appropriate glycemic control was achieved (=75%, lower than the pre-defined threshold of 80%), two patients discontinued the fast due to hyperglycemia, and six finished the fasting period, however, with inappropriate fasting glycemic control (entirely due to hyperglycemic episodes). There were no obvious differences in baseline characteristics between those with appropriate vs. inappropriate fasting glycemic control on the basal insulin infusion rate provided by the algorithm.
Conclusion: In conclusion, when testing fasting glycemic control with an algorithm-derived individual basal insulin infusion profile during a 24 h fasting period in a cohort unrelated in terms of the hospital environment and catchment area, the success rate was lower than a pre-defined threshold for concluding utility of this algorithm. Therefore, applying this algorithm in order to initiate or optimize basal insulin infusion profiles in type 1 diabetes cannot be generally recommended.
期刊介绍:
Publishing outstanding articles from all fields of endocrinology and diabetology, from molecular biology to clinical research, this journal is a brilliant resource. Since being published in English in 1983, the popularity of this journal has grown steadily, reflecting the importance of this publication within its field.
Original contributions and short communications appear in each issue along with reviews addressing current topics. In addition, supplementary issues are published each year presenting abstracts or proceedings of national and international scientific meetings.
The journal was initially published in German and is still the oldest endocrinological periodical in the German-language market!