Ambient light exposure of photostimulable phosphor plates: is there a safe limit for acceptable image quality?

IF 2.9 2区 医学 Q1 DENTISTRY, ORAL SURGERY & MEDICINE
Dento maxillo facial radiology Pub Date : 2023-10-01 Epub Date: 2023-09-04 DOI:10.1259/dmfr.20230174
Matheus Sampaio-Oliveira, Luiz Eduardo Marinho-Vieira, Francisco Haiter-Neto, Deborah Q Freitas, Matheus L Oliveira
{"title":"Ambient light exposure of photostimulable phosphor plates: is there a safe limit for acceptable image quality?","authors":"Matheus Sampaio-Oliveira, Luiz Eduardo Marinho-Vieira, Francisco Haiter-Neto, Deborah Q Freitas, Matheus L Oliveira","doi":"10.1259/dmfr.20230174","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>To subjectively assess radiographs obtained with photostimulable phosphor (PSP) plates exposed to clinical levels of ambient light prior to read-out to potentially set a safe limit for acceptable image quality.</p><p><strong>Methods and materials: </strong>Six dental regions of a dry human skull were X-rayed using PSP plates from VistaScan and Express under four exposure times: 0.1, 0.2, 0.32, and 0.4 s. Before read-out, the PSP plates were exposed to ambient light for 0, 5, 10, 30, 60, and 90 s. Six observers were asked to classify the 288 resulting radiographs as acceptable or unacceptable based on the identification of anatomical structures and global image quality. The number of answers classifying radiographs as unacceptable was used to calculate a rejection rate; a pairwise comparison for better image quality was further conducted among radiographs considered acceptable. Reproducibility was tested by having 25% of all experimental groups reassessed.</p><p><strong>Results: </strong>Intra- and interobserver agreement ranged from 0.87 to 1.00 and from 0.81 to 0.92, respectively. Exposure of PSP plates to ambient light increased rejection rates mostly as of 10 s. In the pairwise comparison, subtle differences were observed between radiographs obtained with PSP plates not exposed and those exposed to ambient light for 5 s.</p><p><strong>Conclusions: </strong>Ambient light exposure of PSP plates impairs the image quality of radiographs. A safe limit of ambient light exposure of 5 s for VistaScan and Express should be considered. Ambient light exposure of PSP plates within safe limits can avoid retakes and reduce unnecessary patient exposure to X-rays.</p>","PeriodicalId":11261,"journal":{"name":"Dento maxillo facial radiology","volume":null,"pages":null},"PeriodicalIF":2.9000,"publicationDate":"2023-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10552132/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Dento maxillo facial radiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1259/dmfr.20230174","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/9/4 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives: To subjectively assess radiographs obtained with photostimulable phosphor (PSP) plates exposed to clinical levels of ambient light prior to read-out to potentially set a safe limit for acceptable image quality.

Methods and materials: Six dental regions of a dry human skull were X-rayed using PSP plates from VistaScan and Express under four exposure times: 0.1, 0.2, 0.32, and 0.4 s. Before read-out, the PSP plates were exposed to ambient light for 0, 5, 10, 30, 60, and 90 s. Six observers were asked to classify the 288 resulting radiographs as acceptable or unacceptable based on the identification of anatomical structures and global image quality. The number of answers classifying radiographs as unacceptable was used to calculate a rejection rate; a pairwise comparison for better image quality was further conducted among radiographs considered acceptable. Reproducibility was tested by having 25% of all experimental groups reassessed.

Results: Intra- and interobserver agreement ranged from 0.87 to 1.00 and from 0.81 to 0.92, respectively. Exposure of PSP plates to ambient light increased rejection rates mostly as of 10 s. In the pairwise comparison, subtle differences were observed between radiographs obtained with PSP plates not exposed and those exposed to ambient light for 5 s.

Conclusions: Ambient light exposure of PSP plates impairs the image quality of radiographs. A safe limit of ambient light exposure of 5 s for VistaScan and Express should be considered. Ambient light exposure of PSP plates within safe limits can avoid retakes and reduce unnecessary patient exposure to X-rays.

光刺激荧光板的环境光暴露:可接受的图像质量有安全限制吗?
目的:在读出之前,主观评估暴露于临床环境光水平的光刺激性磷光体(PSP)板获得的射线照片,以潜在地设定可接受图像质量的安全极限。方法和材料:使用VistaScan和Express的PSP板,在0.1、0.2、0.32和0.4四个暴露时间下,对干燥的人类头骨的六个牙齿区域进行X射线照相 在读出之前,PSP板暴露于环境光0、5、10、30、60和90 s.六名观察者被要求根据解剖结构和全局图像质量的识别,将288张射线照片分为可接受或不可接受。将射线照片分类为不可接受的答案数量用于计算拒绝率;在被认为可接受的射线照片之间进一步进行了用于更好图像质量的成对比较。通过重新评估所有实验组的25%来测试再现性。结果:观察者内部和观察者之间的一致性分别为0.87至1.00和0.81至0.92。PSP板暴露在环境光下增加了排斥率,主要是从10 s.在成对比较中,在未暴露PSP板和暴露于环境光5年的PSP板获得的射线照片之间观察到细微差异 结论:PSP板的环境光照射会损害射线照片的图像质量。环境光暴露的安全限值为5 应考虑VistaScan和Express的s。PSP板在安全范围内的环境光照射可以避免重拍,并减少患者不必要的X射线照射。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
5.60
自引率
9.10%
发文量
65
审稿时长
4-8 weeks
期刊介绍: Dentomaxillofacial Radiology (DMFR) is the journal of the International Association of Dentomaxillofacial Radiology (IADMFR) and covers the closely related fields of oral radiology and head and neck imaging. Established in 1972, DMFR is a key resource keeping dentists, radiologists and clinicians and scientists with an interest in Head and Neck imaging abreast of important research and developments in oral and maxillofacial radiology. The DMFR editorial board features a panel of international experts including Editor-in-Chief Professor Ralf Schulze. Our editorial board provide their expertise and guidance in shaping the content and direction of the journal. Quick Facts: - 2015 Impact Factor - 1.919 - Receipt to first decision - average of 3 weeks - Acceptance to online publication - average of 3 weeks - Open access option - ISSN: 0250-832X - eISSN: 1476-542X
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信