Giuseppe Lippi, Laura Pighi, Marco Tosi, Marco Vettori, Giovanni Celegon, Emmanuel J Favaloro, Gian Luca Salvagno
{"title":"Effect of syringe underfilling on the quality of venous blood gas analysis.","authors":"Giuseppe Lippi, Laura Pighi, Marco Tosi, Marco Vettori, Giovanni Celegon, Emmanuel J Favaloro, Gian Luca Salvagno","doi":"10.1515/dx-2023-0096","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>There is limited information on the influence of collecting small amounts of blood on the quality of blood gas analysis. Therefore, the purpose of this study was to investigate the effects of different degrees of underfilling of syringes on test results of venous blood gas analysis.</p><p><strong>Methods: </strong>Venous blood was collected by venipuncture from 19 healthcare workers in three 1.0 mL syringes for blood gas analysis, by manually aspirating different volumes of blood (i.e., 1.0, 0.5 and 0.25 mL). Routine blood gas analysis was then immediately performed with GEM Premier 5,000. The results of the two underfilled syringes were compared with those of the reference syringe filled with appropriate blood volume.</p><p><strong>Results: </strong>The values of most assayed parameters did not differ significantly in the two underfilled syringes. Statistically significant variations were found for lactate, hematocrit and total hemoglobin, the values of which gradually increased as the fill volume diminished, as well as for sodium concentration, which decreased in both insufficiently filled blood gas syringes. The bias was clinically meaningful for lactate in syringe filled with 0.25 mL of blood, and for hematocrit, total hemoglobin and sodium in both syringes containing 0.5 and 0.25 mL of blood.</p><p><strong>Conclusions: </strong>Collection of smaller volumes of venous blood than the specified filling volume in blood gas syringes may have an effect on the quality of some test results, namely lactate, hematocrit, total hemoglobin and sodium. Specific indications must be given for standardizing the volume of blood to be collected within these syringes.</p>","PeriodicalId":11273,"journal":{"name":"Diagnosis","volume":" ","pages":"91-96"},"PeriodicalIF":2.2000,"publicationDate":"2023-09-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Diagnosis","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1515/dx-2023-0096","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/2/1 0:00:00","PubModel":"eCollection","JCR":"Q2","JCRName":"MEDICINE, GENERAL & INTERNAL","Score":null,"Total":0}
引用次数: 0
Abstract
Objectives: There is limited information on the influence of collecting small amounts of blood on the quality of blood gas analysis. Therefore, the purpose of this study was to investigate the effects of different degrees of underfilling of syringes on test results of venous blood gas analysis.
Methods: Venous blood was collected by venipuncture from 19 healthcare workers in three 1.0 mL syringes for blood gas analysis, by manually aspirating different volumes of blood (i.e., 1.0, 0.5 and 0.25 mL). Routine blood gas analysis was then immediately performed with GEM Premier 5,000. The results of the two underfilled syringes were compared with those of the reference syringe filled with appropriate blood volume.
Results: The values of most assayed parameters did not differ significantly in the two underfilled syringes. Statistically significant variations were found for lactate, hematocrit and total hemoglobin, the values of which gradually increased as the fill volume diminished, as well as for sodium concentration, which decreased in both insufficiently filled blood gas syringes. The bias was clinically meaningful for lactate in syringe filled with 0.25 mL of blood, and for hematocrit, total hemoglobin and sodium in both syringes containing 0.5 and 0.25 mL of blood.
Conclusions: Collection of smaller volumes of venous blood than the specified filling volume in blood gas syringes may have an effect on the quality of some test results, namely lactate, hematocrit, total hemoglobin and sodium. Specific indications must be given for standardizing the volume of blood to be collected within these syringes.
期刊介绍:
Diagnosis focuses on how diagnosis can be advanced, how it is taught, and how and why it can fail, leading to diagnostic errors. The journal welcomes both fundamental and applied works, improvement initiatives, opinions, and debates to encourage new thinking on improving this critical aspect of healthcare quality. Topics: -Factors that promote diagnostic quality and safety -Clinical reasoning -Diagnostic errors in medicine -The factors that contribute to diagnostic error: human factors, cognitive issues, and system-related breakdowns -Improving the value of diagnosis – eliminating waste and unnecessary testing -How culture and removing blame promote awareness of diagnostic errors -Training and education related to clinical reasoning and diagnostic skills -Advances in laboratory testing and imaging that improve diagnostic capability -Local, national and international initiatives to reduce diagnostic error