Investigation of the factors influencing spinal manipulative therapy force transmission through the thorax: a cadaveric study.

IF 2 4区 医学 Q2 REHABILITATION
Jérémie Mikhail, Martha Funabashi, Stéphane Sobczak, Martin Descarreaux, Isabelle Pagé
{"title":"Investigation of the factors influencing spinal manipulative therapy force transmission through the thorax: a cadaveric study.","authors":"Jérémie Mikhail, Martha Funabashi, Stéphane Sobczak, Martin Descarreaux, Isabelle Pagé","doi":"10.1186/s12998-023-00493-1","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Spinal manipulative therapy (SMT) clinical effects are believed to be linked to its force-time profile characteristics. Previous studies have revealed that the force measured at the patient-table interface is most commonly greater than the one applied at the clinician-patient interface. The factors explaining this force amplification remains unclear.</p><p><strong>Objective: </strong>To determine the difference between the force applied to a cadaveric specimen's thoracic spine and the resulting force measured by a force-sensing table, as well as to evaluate the relationship between this difference and both the SMT force-time characteristics and the specimens' characteristics.</p><p><strong>Methods: </strong>Twenty-five SMTs with different force-time profiles were delivered by an apparatus at the T7 vertebra of nine human cadaveric specimens lying prone on a treatment table equipped with a force plate. The difference between the force applied by the apparatus and the resulting force measured by the force plate was calculated in absolute force (F<sub>diff</sub>) and as the percentage of the applied force (F<sub>diff%</sub>). Kinematics markers were inserted into T6 to T8 spinous and transverse processes to evaluate vertebral displacements during the SMT thrusts. Mixed-effects linear models were run to evaluate the variance in F<sub>diff</sub> and F<sub>diff%</sub> explained by SMT characteristics (peak force, thrust duration and force application rate), T6 to T8 relative and total displacements, and specimens' characteristics (BMI, height, weight, kyphosis angle, thoracic thickness).</p><p><strong>Results: </strong>Sixty percent of the trials showed lower force measured at the force plate than the one applied at T7. F<sub>diff</sub>¸ was significantly predicted (R<sup>2</sup><sub>marginal</sub> = 0.54) by peak force, thrust duration, thoracic thickness and T6-T7 relative displacement in the z-axis (postero-anterior). F<sub>diff%</sub> was significantly predicted (R<sup>2</sup><sub>marginal</sub> = 0.56) by force application rate, thoracic thickness and total T6 displacements. For both dependant variables, thoracic thickness showed the highest R<sup>2</sup><sub>marginal</sub> out of all predictors.</p><p><strong>Conclusion: </strong>Difference in force between the clinician-patient and the patient-table interfaces is influenced by SMT force-time characteristics and by thoracic thickness. How these differences in force are associated with vertebral displacements remains unclear. Although further studies are needed, clinicians should consider thorax thickness as a possible modulator of forces being transmitted through it during prone SMT procedures.</p>","PeriodicalId":48572,"journal":{"name":"Chiropractic & Manual Therapies","volume":null,"pages":null},"PeriodicalIF":2.0000,"publicationDate":"2023-08-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10405484/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Chiropractic & Manual Therapies","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1186/s12998-023-00493-1","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"REHABILITATION","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Spinal manipulative therapy (SMT) clinical effects are believed to be linked to its force-time profile characteristics. Previous studies have revealed that the force measured at the patient-table interface is most commonly greater than the one applied at the clinician-patient interface. The factors explaining this force amplification remains unclear.

Objective: To determine the difference between the force applied to a cadaveric specimen's thoracic spine and the resulting force measured by a force-sensing table, as well as to evaluate the relationship between this difference and both the SMT force-time characteristics and the specimens' characteristics.

Methods: Twenty-five SMTs with different force-time profiles were delivered by an apparatus at the T7 vertebra of nine human cadaveric specimens lying prone on a treatment table equipped with a force plate. The difference between the force applied by the apparatus and the resulting force measured by the force plate was calculated in absolute force (Fdiff) and as the percentage of the applied force (Fdiff%). Kinematics markers were inserted into T6 to T8 spinous and transverse processes to evaluate vertebral displacements during the SMT thrusts. Mixed-effects linear models were run to evaluate the variance in Fdiff and Fdiff% explained by SMT characteristics (peak force, thrust duration and force application rate), T6 to T8 relative and total displacements, and specimens' characteristics (BMI, height, weight, kyphosis angle, thoracic thickness).

Results: Sixty percent of the trials showed lower force measured at the force plate than the one applied at T7. Fdiff¸ was significantly predicted (R2marginal = 0.54) by peak force, thrust duration, thoracic thickness and T6-T7 relative displacement in the z-axis (postero-anterior). Fdiff% was significantly predicted (R2marginal = 0.56) by force application rate, thoracic thickness and total T6 displacements. For both dependant variables, thoracic thickness showed the highest R2marginal out of all predictors.

Conclusion: Difference in force between the clinician-patient and the patient-table interfaces is influenced by SMT force-time characteristics and by thoracic thickness. How these differences in force are associated with vertebral displacements remains unclear. Although further studies are needed, clinicians should consider thorax thickness as a possible modulator of forces being transmitted through it during prone SMT procedures.

Abstract Image

Abstract Image

Abstract Image

影响脊柱推拿疗法经胸腔力传递的因素探讨:一项尸体研究。
背景:脊柱推拿疗法(SMT)的临床效果被认为与其力-时间剖面特征有关。先前的研究表明,在病人-桌子界面测量的力通常大于在临床-病人界面施加的力。解释这种力量放大的因素尚不清楚。目的:确定施加在尸体标本胸椎上的力与力传感表测得的力之间的差异,并评估这种差异与SMT力-时间特征和标本特征之间的关系。方法:用器械对9具躺在装有测力板的治疗台上的人体T7椎体进行25次不同力-时间分布的smt。仪器施加的力与测力板测量的结果力之间的差值以绝对力(Fdiff)和施加力的百分比(Fdiff%)计算。运动学标记插入T6至T8棘突和横突,以评估SMT推力期间的椎体位移。使用混合效应线性模型来评估SMT特征(峰值力、推力持续时间和施力率)、T6至T8相对位移和总位移以及标本特征(BMI、身高、体重、后凸角、胸厚)解释的Fdiff和Fdiff%的方差。结果:60%的试验显示力板处测得的力低于T7处施加的力。峰值力、推力持续时间、胸椎厚度和z轴(后前)T6-T7相对位移显著预测Fdiff (R2marginal = 0.54)。Fdiff%可通过施力率、胸椎厚度和T6移位量显著预测(R2marginal = 0.56)。对于这两个因变量,胸廓厚度在所有预测因子中显示出最高的r2边际。结论:SMT力-时间特征和胸廓厚度影响了临床-患者和患者-手术台界面的力差异。这些力的差异与椎体位移的关系尚不清楚。尽管需要进一步的研究,临床医生应该考虑胸腔厚度作为俯卧SMT过程中通过它传递的力的可能调节剂。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Chiropractic & Manual Therapies
Chiropractic & Manual Therapies Medicine-Complementary and Alternative Medicine
CiteScore
3.20
自引率
15.80%
发文量
48
审稿时长
20 weeks
期刊介绍: Chiropractic & Manual Therapies publishes manuscripts on all aspects of evidence-based information that is clinically relevant to chiropractors, manual therapists and related health care professionals. Chiropractic & Manual Therapies is an open access journal that aims to provide chiropractors, manual therapists and related health professionals with clinically relevant, evidence-based information. Chiropractic and other manual therapies share a relatively broad diagnostic practice and treatment scope, emphasizing the structure and function of the body''s musculoskeletal framework (especially the spine). The practices of chiropractic and manual therapies are closely associated with treatments including manipulation, which is a key intervention. The range of services provided can also include massage, mobilisation, physical therapies, dry needling, lifestyle and dietary counselling, plus a variety of other associated therapeutic and rehabilitation approaches. Chiropractic & Manual Therapies continues to serve as a critical resource in this field, and as an open access publication, is more readily available to practitioners, researchers and clinicians worldwide.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信