新型自定位球囊可扩张经导管生物主动脉瓣系统在绵羊主动脉绑扎模型中的长期生物功能表现。

IF 2.5 3区 医学 Q2 CARDIAC & CARDIOVASCULAR SYSTEMS
Cardiology journal Pub Date : 2024-01-01 Epub Date: 2023-07-31 DOI:10.5603/CJ.a2023.0048
Piotr P Buszman, Mateusz Kachel, Krzysztof P Milewski, Magdalena Michalak, Wojciech Domaradzki, Maciej Pruski, Michał Sobota, Carlos Fernandez, Karolina Łukasik, Jerzy Nożyński, Piotr Hirnle, Paweł Kaźmierczak, Andrzej Bochenek, Paweł E Buszman
{"title":"新型自定位球囊可扩张经导管生物主动脉瓣系统在绵羊主动脉绑扎模型中的长期生物功能表现。","authors":"Piotr P Buszman, Mateusz Kachel, Krzysztof P Milewski, Magdalena Michalak, Wojciech Domaradzki, Maciej Pruski, Michał Sobota, Carlos Fernandez, Karolina Łukasik, Jerzy Nożyński, Piotr Hirnle, Paweł Kaźmierczak, Andrzej Bochenek, Paweł E Buszman","doi":"10.5603/CJ.a2023.0048","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The aim of the study was to evaluate bio-functionality of a novel, proprietary balloon-expandable biological transcatheter aortic valve implantation (TAVI) system (InFlow, CardValve Consortium, Poland) in an ovine model of aortic banding.</p><p><strong>Methods: </strong>Surgical ascending aorta banding was created in 21 sheep. Two weeks later, 18 biological valves were implanted within the model using 15-16 F InFlow TAVI systems and carotid cut-down approach. Follow-up transthoracic echocardiography was performed at 30, 90, and 180-day. At designated time, animals were euthanized and valves harvested for analysis.</p><p><strong>Results: </strong>All sheep survived the banding procedure. There were 4 (22%) procedure related deaths within a 7-day period. During the observation an additional 2 sheep died. In one, the valve dislocated after the procedure - the animal was excluded. Two animals completed 30-day follow up, five 90-day follow-up and four terminal follow-up of 180 days. Valves examined via transesophageal echocardiography showed proper hemodynamic parameters without evidence of structural valve deterioration. The maximum and average flow gradients at 180 days were 31.4 (23.3-37.7) and 17.5 (13.1-20.2) mmHg, respectively. There was one case of moderate insufficiency and no case of perivalvular leaks. By histopathology, there were no inflammation, thrombosis, nor calcifications in any tested valves at long-term follow-up. Neointimal coverage of stent struts increased with time from basal part in \"early\" groups to nearly 3/4 of stent length in the 180-day group. The pannus tissue showed maturation that increased with time with no stenotic \"collar\" visible in orthotopically implanted valves.</p><p><strong>Conclusions: </strong>The study showed good hemodynamic performance, durability and biocompatibility of the novel biological THV.</p>","PeriodicalId":9492,"journal":{"name":"Cardiology journal","volume":" ","pages":"124-132"},"PeriodicalIF":2.5000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10919571/pdf/","citationCount":"0","resultStr":"{\"title\":\"Long-term bio-functional performance of a novel, self-positioning balloon expandable transcatheter biological aortic valve system in the ovine aortic banding model.\",\"authors\":\"Piotr P Buszman, Mateusz Kachel, Krzysztof P Milewski, Magdalena Michalak, Wojciech Domaradzki, Maciej Pruski, Michał Sobota, Carlos Fernandez, Karolina Łukasik, Jerzy Nożyński, Piotr Hirnle, Paweł Kaźmierczak, Andrzej Bochenek, Paweł E Buszman\",\"doi\":\"10.5603/CJ.a2023.0048\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>The aim of the study was to evaluate bio-functionality of a novel, proprietary balloon-expandable biological transcatheter aortic valve implantation (TAVI) system (InFlow, CardValve Consortium, Poland) in an ovine model of aortic banding.</p><p><strong>Methods: </strong>Surgical ascending aorta banding was created in 21 sheep. Two weeks later, 18 biological valves were implanted within the model using 15-16 F InFlow TAVI systems and carotid cut-down approach. Follow-up transthoracic echocardiography was performed at 30, 90, and 180-day. At designated time, animals were euthanized and valves harvested for analysis.</p><p><strong>Results: </strong>All sheep survived the banding procedure. There were 4 (22%) procedure related deaths within a 7-day period. During the observation an additional 2 sheep died. In one, the valve dislocated after the procedure - the animal was excluded. Two animals completed 30-day follow up, five 90-day follow-up and four terminal follow-up of 180 days. Valves examined via transesophageal echocardiography showed proper hemodynamic parameters without evidence of structural valve deterioration. The maximum and average flow gradients at 180 days were 31.4 (23.3-37.7) and 17.5 (13.1-20.2) mmHg, respectively. There was one case of moderate insufficiency and no case of perivalvular leaks. By histopathology, there were no inflammation, thrombosis, nor calcifications in any tested valves at long-term follow-up. Neointimal coverage of stent struts increased with time from basal part in \\\"early\\\" groups to nearly 3/4 of stent length in the 180-day group. The pannus tissue showed maturation that increased with time with no stenotic \\\"collar\\\" visible in orthotopically implanted valves.</p><p><strong>Conclusions: </strong>The study showed good hemodynamic performance, durability and biocompatibility of the novel biological THV.</p>\",\"PeriodicalId\":9492,\"journal\":{\"name\":\"Cardiology journal\",\"volume\":\" \",\"pages\":\"124-132\"},\"PeriodicalIF\":2.5000,\"publicationDate\":\"2024-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10919571/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Cardiology journal\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.5603/CJ.a2023.0048\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2023/7/31 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q2\",\"JCRName\":\"CARDIAC & CARDIOVASCULAR SYSTEMS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Cardiology journal","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.5603/CJ.a2023.0048","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/7/31 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

摘要

研究背景该研究旨在评估一种新型、专有的球囊扩张型生物经导管主动脉瓣植入(TAVI)系统(InFlow,CardValve Consortium,波兰)在绵羊主动脉束带模型中的生物功能:方法:对 21 只绵羊的升主动脉进行手术捆绑。两周后,使用 15-16 F InFlow TAVI 系统和颈动脉切开法在模型中植入 18 个生物瓣膜。分别在 30 天、90 天和 180 天进行经胸超声心动图随访。在指定时间对动物实施安乐死,并采集瓣膜进行分析:结果:所有绵羊均在捆绑过程中存活。7天内有4只羊(22%)因手术死亡。在观察期间,又有 2 只羊死亡。其中一只羊的瓣膜在手术后脱位--该动物被排除在外。两只绵羊完成了 30 天的随访,五只完成了 90 天的随访,四只完成了 180 天的终末随访。经食道超声心动图检查的瓣膜显示血流动力学参数正常,没有瓣膜结构退化的迹象。180天时的最大和平均血流阶差分别为31.4(23.3-37.7)毫米汞柱和17.5(13.1-20.2)毫米汞柱。有一例中度血流不全,没有一例瓣周漏。通过组织病理学检查,在长期随访中,所有接受检测的瓣膜均未出现炎症、血栓或钙化。支架支柱的新内膜覆盖率随着时间的推移而增加,从 "早期 "组的基底部分增加到 180 天组支架长度的近 3/4。随着时间的推移,瓣膜组织逐渐成熟,在正位植入的瓣膜上看不到狭窄的 "领圈":研究表明,新型生物 THV 具有良好的血液动力学性能、耐用性和生物相容性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Long-term bio-functional performance of a novel, self-positioning balloon expandable transcatheter biological aortic valve system in the ovine aortic banding model.

Background: The aim of the study was to evaluate bio-functionality of a novel, proprietary balloon-expandable biological transcatheter aortic valve implantation (TAVI) system (InFlow, CardValve Consortium, Poland) in an ovine model of aortic banding.

Methods: Surgical ascending aorta banding was created in 21 sheep. Two weeks later, 18 biological valves were implanted within the model using 15-16 F InFlow TAVI systems and carotid cut-down approach. Follow-up transthoracic echocardiography was performed at 30, 90, and 180-day. At designated time, animals were euthanized and valves harvested for analysis.

Results: All sheep survived the banding procedure. There were 4 (22%) procedure related deaths within a 7-day period. During the observation an additional 2 sheep died. In one, the valve dislocated after the procedure - the animal was excluded. Two animals completed 30-day follow up, five 90-day follow-up and four terminal follow-up of 180 days. Valves examined via transesophageal echocardiography showed proper hemodynamic parameters without evidence of structural valve deterioration. The maximum and average flow gradients at 180 days were 31.4 (23.3-37.7) and 17.5 (13.1-20.2) mmHg, respectively. There was one case of moderate insufficiency and no case of perivalvular leaks. By histopathology, there were no inflammation, thrombosis, nor calcifications in any tested valves at long-term follow-up. Neointimal coverage of stent struts increased with time from basal part in "early" groups to nearly 3/4 of stent length in the 180-day group. The pannus tissue showed maturation that increased with time with no stenotic "collar" visible in orthotopically implanted valves.

Conclusions: The study showed good hemodynamic performance, durability and biocompatibility of the novel biological THV.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Cardiology journal
Cardiology journal CARDIAC & CARDIOVASCULAR SYSTEMS-
CiteScore
5.10
自引率
10.30%
发文量
188
审稿时长
4-8 weeks
期刊介绍: Cardiology Journal is a scientific, peer-reviewed journal covering a broad spectrum of topics in cardiology. The journal has been published since 1994 and over the years it has become an internationally recognized journal of cardiological and medical community. Cardiology Journal is the journal for practicing cardiologists, researchers, and young trainees benefiting from broad spectrum of useful educational content.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信