{"title":"当使用肝素血浆管时,在Li联XL分析仪上对25-OH维生素D免疫测定的未暴露干扰。","authors":"Julie Jacobs,&nbsp;Frank Vanneste,&nbsp;Julie Hotton,&nbsp;Nathalie Miller,&nbsp;Mathieu Cauchie,&nbsp;Laetitia Brassinne,&nbsp;Antoine Mairesse","doi":"10.1080/00365513.2023.2241356","DOIUrl":null,"url":null,"abstract":"<p><p>This study investigates the impact of sample type on the measurement of 25-OH-vitamin D using the Liaison XL (Diasorin) and Cobas e801 (Roche). This investigation was motivated by the need to optimize sample volume usage, which led us to adopt the use of heparin plasma, an alternative proposed by Diasorin in their specification. Discordant and unexplainable results were observed, prompting us to evaluate the effect of sample type on the accuracy of the 25-OH-vitamin D measurements. We collected 34 different paired samples from a randomly selected patients who had two types of tubes taken simultaneously: serum-gel and lithium-heparin plasma tubes. The 25-OH-vitamin D levels were measured using Cobas e801 and Liaison. Statistical analysis was performed using the Mann-Whitney test to calculate the p-value. Biases were also calculated. When comparing the heparin matrix with the serum matrix on the Liaison XL analyzer, a higher proportion (<i>p</i> < .0001; 79% versus 64%) of patients were classified in the 'normal group', while fewer were classified in the 'insufficiency' or 'deficiency group'. The heparin tubes on the Liaison XL analyzer showed a mean bias of 57.5%) (<i>p</i>-value < .001; 95%CI: 37.6-77.4) compared to the serum tubes. On the other hand, the heparin tubes on the Cobas e801 analyzer showed a mean bias of -0.2% (95%CI: -4.8 to 4.5) compared to the serum tubes. It is imperative for laboratory professionals to be aware of this interference for an accurate measurement of 25-OH-vitamin D levels on the Liaison XL. Further research is needed to understand the mechanism of this interference.</p>","PeriodicalId":21474,"journal":{"name":"Scandinavian Journal of Clinical & Laboratory Investigation","volume":" ","pages":"390-393"},"PeriodicalIF":1.3000,"publicationDate":"2023-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Undisclosed interference in 25-OH-Vitamin D immunoassay on Liaison XL analyzer when using heparin plasma tubes.\",\"authors\":\"Julie Jacobs,&nbsp;Frank Vanneste,&nbsp;Julie Hotton,&nbsp;Nathalie Miller,&nbsp;Mathieu Cauchie,&nbsp;Laetitia Brassinne,&nbsp;Antoine Mairesse\",\"doi\":\"10.1080/00365513.2023.2241356\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>This study investigates the impact of sample type on the measurement of 25-OH-vitamin D using the Liaison XL (Diasorin) and Cobas e801 (Roche). This investigation was motivated by the need to optimize sample volume usage, which led us to adopt the use of heparin plasma, an alternative proposed by Diasorin in their specification. Discordant and unexplainable results were observed, prompting us to evaluate the effect of sample type on the accuracy of the 25-OH-vitamin D measurements. We collected 34 different paired samples from a randomly selected patients who had two types of tubes taken simultaneously: serum-gel and lithium-heparin plasma tubes. The 25-OH-vitamin D levels were measured using Cobas e801 and Liaison. Statistical analysis was performed using the Mann-Whitney test to calculate the p-value. Biases were also calculated. When comparing the heparin matrix with the serum matrix on the Liaison XL analyzer, a higher proportion (<i>p</i> < .0001; 79% versus 64%) of patients were classified in the 'normal group', while fewer were classified in the 'insufficiency' or 'deficiency group'. The heparin tubes on the Liaison XL analyzer showed a mean bias of 57.5%) (<i>p</i>-value < .001; 95%CI: 37.6-77.4) compared to the serum tubes. On the other hand, the heparin tubes on the Cobas e801 analyzer showed a mean bias of -0.2% (95%CI: -4.8 to 4.5) compared to the serum tubes. It is imperative for laboratory professionals to be aware of this interference for an accurate measurement of 25-OH-vitamin D levels on the Liaison XL. Further research is needed to understand the mechanism of this interference.</p>\",\"PeriodicalId\":21474,\"journal\":{\"name\":\"Scandinavian Journal of Clinical & Laboratory Investigation\",\"volume\":\" \",\"pages\":\"390-393\"},\"PeriodicalIF\":1.3000,\"publicationDate\":\"2023-10-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Scandinavian Journal of Clinical & Laboratory Investigation\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1080/00365513.2023.2241356\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2023/7/28 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q4\",\"JCRName\":\"MEDICINE, RESEARCH & EXPERIMENTAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Scandinavian Journal of Clinical & Laboratory Investigation","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/00365513.2023.2241356","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/7/28 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"MEDICINE, RESEARCH & EXPERIMENTAL","Score":null,"Total":0}
引用次数: 0

摘要

本研究调查了样品类型对使用Li联XL(Diasorin)和Cobas e801(Roche)测量25-OH维生素D的影响。这项研究的动机是需要优化样本量的使用,这导致我们采用了肝素血浆的使用,Diasorin在其规范中提出了一种替代方案。观察到了不一致且无法解释的结果,促使我们评估样本类型对25-OH维生素D测量准确性的影响。我们从随机选择的患者中收集了34个不同的配对样本,这些患者同时使用了两种类型的试管:血清凝胶和肝素锂血浆试管。使用Cobas e801和Lions测量25 OH维生素D水平。使用Mann-Whitney检验进行统计分析以计算p值。还计算了偏差。当在Li联XL分析仪上比较肝素基质和血清基质时,更高的比例(p p值
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Undisclosed interference in 25-OH-Vitamin D immunoassay on Liaison XL analyzer when using heparin plasma tubes.

This study investigates the impact of sample type on the measurement of 25-OH-vitamin D using the Liaison XL (Diasorin) and Cobas e801 (Roche). This investigation was motivated by the need to optimize sample volume usage, which led us to adopt the use of heparin plasma, an alternative proposed by Diasorin in their specification. Discordant and unexplainable results were observed, prompting us to evaluate the effect of sample type on the accuracy of the 25-OH-vitamin D measurements. We collected 34 different paired samples from a randomly selected patients who had two types of tubes taken simultaneously: serum-gel and lithium-heparin plasma tubes. The 25-OH-vitamin D levels were measured using Cobas e801 and Liaison. Statistical analysis was performed using the Mann-Whitney test to calculate the p-value. Biases were also calculated. When comparing the heparin matrix with the serum matrix on the Liaison XL analyzer, a higher proportion (p < .0001; 79% versus 64%) of patients were classified in the 'normal group', while fewer were classified in the 'insufficiency' or 'deficiency group'. The heparin tubes on the Liaison XL analyzer showed a mean bias of 57.5%) (p-value < .001; 95%CI: 37.6-77.4) compared to the serum tubes. On the other hand, the heparin tubes on the Cobas e801 analyzer showed a mean bias of -0.2% (95%CI: -4.8 to 4.5) compared to the serum tubes. It is imperative for laboratory professionals to be aware of this interference for an accurate measurement of 25-OH-vitamin D levels on the Liaison XL. Further research is needed to understand the mechanism of this interference.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
3.50
自引率
4.80%
发文量
85
审稿时长
4-8 weeks
期刊介绍: The Scandinavian Journal of Clinical and Laboratory Investigation is an international scientific journal covering clinically oriented biochemical and physiological research. Since the launch of the journal in 1949, it has been a forum for international laboratory medicine, closely related to, and edited by, The Scandinavian Society for Clinical Chemistry. The journal contains peer-reviewed articles, editorials, invited reviews, and short technical notes, as well as several supplements each year. Supplements consist of monographs, and symposium and congress reports covering subjects within clinical chemistry and clinical physiology.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信