{"title":"长轴与短轴视野免疫PET/CT:89Zr曲妥珠单抗的半定量评价。","authors":"Philipp Mohr, Joyce van Sluis, Laura Providência, Johannes H van Snick, Marjolijn N Lub-de Hooge, Antoon T Willemsen, Andor W J M Glaudemans, Ronald Boellaard, Adriaan A Lammertsma, Adrienne H Brouwers, Charalampos Tsoumpas","doi":"10.2967/jnumed.123.265621","DOIUrl":null,"url":null,"abstract":"<p><p>The purpose of this study was to quantify any differences between the SUVs of <sup>89</sup>Zr immuno-PET scans obtained using a PET/CT system with a long axial field of view (LAFOV; Biograph Vision Quadra) compared to a PET/CT system with a short axial field of view (SAFOV; Biograph Vision) and to evaluate how LAFOV PET scan duration affects image noise and SUV metrics. <b>Methods:</b> Five metastatic breast cancer patients were scanned consecutively on SAFOV and LAFOV PET/CT scanners. Four additional patients were scanned using only LAFOV PET/CT. Scans on both systems lasted approximately 30 min and were acquired 4 d after injection of 37 MBq of <sup>89</sup>Zr-trastuzumab. LAFOV list-mode data were reprocessed to obtain images acquired using shorter scan durations (15, 10, 7.5, 5, and 3 min). Volumes of interest were placed in healthy tissues, and tumors were segmented semiautomatically to compare coefficients of variation and to perform Bland-Altman analysis on SUV metrics (SUV<sub>max</sub>, SUV<sub>peak</sub>, and SUV<sub>mean</sub>). <b>Results:</b> Using 30-min images, 2 commonly used lesion SUV metrics were higher for SAFOV than for LAFOV PET (SUV<sub>max</sub>, 16.2% ± 13.4%, and SUV<sub>peak</sub>, 10.1% ± 7.2%), whereas the SUV<sub>mean</sub> of healthy tissues showed minimal differences (0.7% ± 5.8%). Coefficients of variation in the liver derived from 30-min SAFOV PET were between those of 3- and 5-min LAFOV PET. The smallest SUV<sub>max</sub> and SUV<sub>peak</sub> differences between SAFOV and LAFOV were found for 3-min LAFOV PET. <b>Conclusion:</b> LAFOV <sup>89</sup>Zr immuno-PET showed a lower SUV<sub>max</sub> and SUV<sub>peak</sub> than SAFOV because of lower image noise. LAFOV PET scan duration may be reduced at the expense of increasing image noise and bias in SUV metrics. Nevertheless, SUV<sub>peak</sub> showed only minimal bias when reducing scan duration from 30 to 10 min.</p>","PeriodicalId":16758,"journal":{"name":"Journal of Nuclear Medicine","volume":null,"pages":null},"PeriodicalIF":9.1000,"publicationDate":"2023-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"2","resultStr":"{\"title\":\"Long Versus Short Axial Field of View Immuno-PET/CT: Semiquantitative Evaluation for <sup>89</sup>Zr-Trastuzumab.\",\"authors\":\"Philipp Mohr, Joyce van Sluis, Laura Providência, Johannes H van Snick, Marjolijn N Lub-de Hooge, Antoon T Willemsen, Andor W J M Glaudemans, Ronald Boellaard, Adriaan A Lammertsma, Adrienne H Brouwers, Charalampos Tsoumpas\",\"doi\":\"10.2967/jnumed.123.265621\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The purpose of this study was to quantify any differences between the SUVs of <sup>89</sup>Zr immuno-PET scans obtained using a PET/CT system with a long axial field of view (LAFOV; Biograph Vision Quadra) compared to a PET/CT system with a short axial field of view (SAFOV; Biograph Vision) and to evaluate how LAFOV PET scan duration affects image noise and SUV metrics. <b>Methods:</b> Five metastatic breast cancer patients were scanned consecutively on SAFOV and LAFOV PET/CT scanners. Four additional patients were scanned using only LAFOV PET/CT. Scans on both systems lasted approximately 30 min and were acquired 4 d after injection of 37 MBq of <sup>89</sup>Zr-trastuzumab. LAFOV list-mode data were reprocessed to obtain images acquired using shorter scan durations (15, 10, 7.5, 5, and 3 min). Volumes of interest were placed in healthy tissues, and tumors were segmented semiautomatically to compare coefficients of variation and to perform Bland-Altman analysis on SUV metrics (SUV<sub>max</sub>, SUV<sub>peak</sub>, and SUV<sub>mean</sub>). <b>Results:</b> Using 30-min images, 2 commonly used lesion SUV metrics were higher for SAFOV than for LAFOV PET (SUV<sub>max</sub>, 16.2% ± 13.4%, and SUV<sub>peak</sub>, 10.1% ± 7.2%), whereas the SUV<sub>mean</sub> of healthy tissues showed minimal differences (0.7% ± 5.8%). Coefficients of variation in the liver derived from 30-min SAFOV PET were between those of 3- and 5-min LAFOV PET. The smallest SUV<sub>max</sub> and SUV<sub>peak</sub> differences between SAFOV and LAFOV were found for 3-min LAFOV PET. <b>Conclusion:</b> LAFOV <sup>89</sup>Zr immuno-PET showed a lower SUV<sub>max</sub> and SUV<sub>peak</sub> than SAFOV because of lower image noise. LAFOV PET scan duration may be reduced at the expense of increasing image noise and bias in SUV metrics. Nevertheless, SUV<sub>peak</sub> showed only minimal bias when reducing scan duration from 30 to 10 min.</p>\",\"PeriodicalId\":16758,\"journal\":{\"name\":\"Journal of Nuclear Medicine\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":9.1000,\"publicationDate\":\"2023-11-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"2\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Nuclear Medicine\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.2967/jnumed.123.265621\",\"RegionNum\":1,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2023/8/3 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q1\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Nuclear Medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.2967/jnumed.123.265621","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/8/3 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
Long Versus Short Axial Field of View Immuno-PET/CT: Semiquantitative Evaluation for 89Zr-Trastuzumab.
The purpose of this study was to quantify any differences between the SUVs of 89Zr immuno-PET scans obtained using a PET/CT system with a long axial field of view (LAFOV; Biograph Vision Quadra) compared to a PET/CT system with a short axial field of view (SAFOV; Biograph Vision) and to evaluate how LAFOV PET scan duration affects image noise and SUV metrics. Methods: Five metastatic breast cancer patients were scanned consecutively on SAFOV and LAFOV PET/CT scanners. Four additional patients were scanned using only LAFOV PET/CT. Scans on both systems lasted approximately 30 min and were acquired 4 d after injection of 37 MBq of 89Zr-trastuzumab. LAFOV list-mode data were reprocessed to obtain images acquired using shorter scan durations (15, 10, 7.5, 5, and 3 min). Volumes of interest were placed in healthy tissues, and tumors were segmented semiautomatically to compare coefficients of variation and to perform Bland-Altman analysis on SUV metrics (SUVmax, SUVpeak, and SUVmean). Results: Using 30-min images, 2 commonly used lesion SUV metrics were higher for SAFOV than for LAFOV PET (SUVmax, 16.2% ± 13.4%, and SUVpeak, 10.1% ± 7.2%), whereas the SUVmean of healthy tissues showed minimal differences (0.7% ± 5.8%). Coefficients of variation in the liver derived from 30-min SAFOV PET were between those of 3- and 5-min LAFOV PET. The smallest SUVmax and SUVpeak differences between SAFOV and LAFOV were found for 3-min LAFOV PET. Conclusion: LAFOV 89Zr immuno-PET showed a lower SUVmax and SUVpeak than SAFOV because of lower image noise. LAFOV PET scan duration may be reduced at the expense of increasing image noise and bias in SUV metrics. Nevertheless, SUVpeak showed only minimal bias when reducing scan duration from 30 to 10 min.
期刊介绍:
The Journal of Nuclear Medicine (JNM), self-published by the Society of Nuclear Medicine and Molecular Imaging (SNMMI), provides readers worldwide with clinical and basic science investigations, continuing education articles, reviews, employment opportunities, and updates on practice and research. In the 2022 Journal Citation Reports (released in June 2023), JNM ranked sixth in impact among 203 medical journals worldwide in the radiology, nuclear medicine, and medical imaging category.