{"title":"使用ArcScan insight 100高频超声测量前晶状体带长度的可重复性和再现性。","authors":"Zehui Zhu,&nbsp;Han Zou,&nbsp;Hongzhe Li,&nbsp;Xueer Wu,&nbsp;Yiyi Wang,&nbsp;Zhangliang Li,&nbsp;Yune Zhao","doi":"10.1080/17434440.2023.2223967","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>This study aimed to evaluate the intra-examiner repeatability and inter-examiner reproducibility in lens zonular length measurements using very high-frequency digital ultrasound (Insight 100).</p><p><strong>Methods: </strong>Two examiners performed ultrasound imaging independently in each subject. The length of temporal and nasal zonules were then measured with a built-in software. Coefficient of variations (CVs) of the three repeated measurements were used to determine intra-examiner variances. Inter-examiner reproducibility was evaluated using intraclass correlation coefficients (ICCs) and the Bland-Altman method.</p><p><strong>Results: </strong>40 eyes of 40 subjects (14male and 26female; mean age 23.9 ± 2.4 years) were included in the study. The CVs for intra-examiner measurement were 2.74% temporally and 4.32% nasally for Examiner 1, and were 1.96% temporally and 1.75% nasally for Examiner 2. For inter-examiner reproducibility, all ICCs were above 0.9. However, there were significant differences between the two examiners in temporal zonular length measurements (<i>p</i> = 0.001), and the differences mainly came from measuring the zonular length manually (<i>p</i> = 0.001) rather than recording images (<i>p</i> = 0.480). No significant differences were found between two measurements by the same examiner after one month (all <i>p</i> > 0.05, all ICCs > 0.8).</p><p><strong>Conclusion: </strong>The Insight 100 device can be used to measure the length of anterior lens zonule with relatively good repeatability and reproducibility.</p><p><strong>Clinical trial registration: </strong>www.clinicaltrials.gov identifier is NCT05657951.</p>","PeriodicalId":12330,"journal":{"name":"Expert Review of Medical Devices","volume":"20 8","pages":"703-710"},"PeriodicalIF":2.9000,"publicationDate":"2023-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Repeatability and reproducibility of anterior lens zonule length measurement using ArcScan insight 100 very high-frequency ultrasound.\",\"authors\":\"Zehui Zhu,&nbsp;Han Zou,&nbsp;Hongzhe Li,&nbsp;Xueer Wu,&nbsp;Yiyi Wang,&nbsp;Zhangliang Li,&nbsp;Yune Zhao\",\"doi\":\"10.1080/17434440.2023.2223967\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objectives: </strong>This study aimed to evaluate the intra-examiner repeatability and inter-examiner reproducibility in lens zonular length measurements using very high-frequency digital ultrasound (Insight 100).</p><p><strong>Methods: </strong>Two examiners performed ultrasound imaging independently in each subject. The length of temporal and nasal zonules were then measured with a built-in software. Coefficient of variations (CVs) of the three repeated measurements were used to determine intra-examiner variances. Inter-examiner reproducibility was evaluated using intraclass correlation coefficients (ICCs) and the Bland-Altman method.</p><p><strong>Results: </strong>40 eyes of 40 subjects (14male and 26female; mean age 23.9 ± 2.4 years) were included in the study. The CVs for intra-examiner measurement were 2.74% temporally and 4.32% nasally for Examiner 1, and were 1.96% temporally and 1.75% nasally for Examiner 2. For inter-examiner reproducibility, all ICCs were above 0.9. However, there were significant differences between the two examiners in temporal zonular length measurements (<i>p</i> = 0.001), and the differences mainly came from measuring the zonular length manually (<i>p</i> = 0.001) rather than recording images (<i>p</i> = 0.480). No significant differences were found between two measurements by the same examiner after one month (all <i>p</i> > 0.05, all ICCs > 0.8).</p><p><strong>Conclusion: </strong>The Insight 100 device can be used to measure the length of anterior lens zonule with relatively good repeatability and reproducibility.</p><p><strong>Clinical trial registration: </strong>www.clinicaltrials.gov identifier is NCT05657951.</p>\",\"PeriodicalId\":12330,\"journal\":{\"name\":\"Expert Review of Medical Devices\",\"volume\":\"20 8\",\"pages\":\"703-710\"},\"PeriodicalIF\":2.9000,\"publicationDate\":\"2023-07-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Expert Review of Medical Devices\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.1080/17434440.2023.2223967\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"ENGINEERING, BIOMEDICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Expert Review of Medical Devices","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1080/17434440.2023.2223967","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

摘要

目的:本研究旨在评估使用甚高频数字超声(Insight 100)测量晶状体带状长度的检查员内部重复性和检查员之间重复性。方法:两名检查人员分别对每名受试者进行超声成像。然后用内置软件测量颞部和鼻腔小带的长度。使用三个重复测量的变异系数(cv)来确定考官内部的方差。使用类内相关系数(ICCs)和Bland-Altman方法评估检查者间的可重复性。结果:40例受试者40只眼,其中男14只,女26只;平均年龄23.9±2.4岁)纳入研究。主考官1颞部和鼻腔测量的cv值分别为2.74%和4.32%,主考官2颞部和鼻腔测量的cv值分别为1.96%和1.75%。对于审查员间的再现性,所有icc均在0.9以上。然而,两名检查者在颞带长度测量上存在显著差异(p = 0.001),差异主要来自人工测量(p = 0.001)而不是记录图像(p = 0.480)。1个月后同一检查者两次测量无显著差异(p均> 0.05,ICCs均> 0.8)。结论:Insight 100仪器可用于前晶状体小带长度的测量,具有较好的重复性和再现性。临床试验注册:www.clinicaltrials.gov标识符:NCT05657951。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Repeatability and reproducibility of anterior lens zonule length measurement using ArcScan insight 100 very high-frequency ultrasound.

Objectives: This study aimed to evaluate the intra-examiner repeatability and inter-examiner reproducibility in lens zonular length measurements using very high-frequency digital ultrasound (Insight 100).

Methods: Two examiners performed ultrasound imaging independently in each subject. The length of temporal and nasal zonules were then measured with a built-in software. Coefficient of variations (CVs) of the three repeated measurements were used to determine intra-examiner variances. Inter-examiner reproducibility was evaluated using intraclass correlation coefficients (ICCs) and the Bland-Altman method.

Results: 40 eyes of 40 subjects (14male and 26female; mean age 23.9 ± 2.4 years) were included in the study. The CVs for intra-examiner measurement were 2.74% temporally and 4.32% nasally for Examiner 1, and were 1.96% temporally and 1.75% nasally for Examiner 2. For inter-examiner reproducibility, all ICCs were above 0.9. However, there were significant differences between the two examiners in temporal zonular length measurements (p = 0.001), and the differences mainly came from measuring the zonular length manually (p = 0.001) rather than recording images (p = 0.480). No significant differences were found between two measurements by the same examiner after one month (all p > 0.05, all ICCs > 0.8).

Conclusion: The Insight 100 device can be used to measure the length of anterior lens zonule with relatively good repeatability and reproducibility.

Clinical trial registration: www.clinicaltrials.gov identifier is NCT05657951.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Expert Review of Medical Devices
Expert Review of Medical Devices 医学-工程:生物医学
CiteScore
5.90
自引率
3.20%
发文量
69
审稿时长
6-12 weeks
期刊介绍: The journal serves the device research community by providing a comprehensive body of high-quality information from leading experts, all subject to rigorous peer review. The Expert Review format is specially structured to optimize the value of the information to reader. Comprehensive coverage by each author in a key area of research or clinical practice is augmented by the following sections: Expert commentary - a personal view on the most effective or promising strategies Five-year view - a clear perspective of future prospects within a realistic timescale Key issues - an executive summary cutting to the author''s most critical points In addition to the Review program, each issue also features Medical Device Profiles - objective assessments of specific devices in development or clinical use to help inform clinical practice. There are also Perspectives - overviews highlighting areas of current debate and controversy, together with reports from the conference scene and invited Editorials.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信