{"title":"一种移动式心律远程监测系统的临床评价。","authors":"Hristo Mateev,&nbsp;Iana Simova,&nbsp;Tzvetana Katova,&nbsp;Nikolay Dimitrov","doi":"10.5402/2012/192670","DOIUrl":null,"url":null,"abstract":"<p><p>Purpose. To evaluate the clinical applicability of a telemonitoring system: telemetric system for collection and distant surveillance of medical information (TEMEO). Methods. We evaluated 60 patients, applying simultaneously standard Holter ECG and telemonitoring. Two different comparisons were performed: (1) TEMEO ECG with standard 12-lead ECG; (2) TEMEO Holter with standard ECG Holter. Results. We found a very high coincidence rate (99.3%) between TEMEO derived ECGs and standard ECGs. Intraclass correlation coefficient analysis revealed high and significant correlation coefficients regarding average, maximal, and minimal heart rate, % of time in tachycardia, single supraventricular ectopic beats (SVEB), and single and couplets of ventricular ectopic beats (VEB) between Holter ECG and TEMEO derived parameters. Couplets of SVEB were recorded as different by the two monitoring systems, however, with a borderline statistical significance. Conclusions. TEMEO derived ECGs have a very high coincidence rate with standard ECGs. TEMEO patient monitoring provides results that are similar to those derived from a standard Holter ECG.</p>","PeriodicalId":73519,"journal":{"name":"ISRN cardiology","volume":"2012 ","pages":"192670"},"PeriodicalIF":0.0000,"publicationDate":"2012-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.5402/2012/192670","citationCount":"11","resultStr":"{\"title\":\"Clinical evaluation of a mobile heart rhythm telemonitoring system.\",\"authors\":\"Hristo Mateev,&nbsp;Iana Simova,&nbsp;Tzvetana Katova,&nbsp;Nikolay Dimitrov\",\"doi\":\"10.5402/2012/192670\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Purpose. To evaluate the clinical applicability of a telemonitoring system: telemetric system for collection and distant surveillance of medical information (TEMEO). Methods. We evaluated 60 patients, applying simultaneously standard Holter ECG and telemonitoring. Two different comparisons were performed: (1) TEMEO ECG with standard 12-lead ECG; (2) TEMEO Holter with standard ECG Holter. Results. We found a very high coincidence rate (99.3%) between TEMEO derived ECGs and standard ECGs. Intraclass correlation coefficient analysis revealed high and significant correlation coefficients regarding average, maximal, and minimal heart rate, % of time in tachycardia, single supraventricular ectopic beats (SVEB), and single and couplets of ventricular ectopic beats (VEB) between Holter ECG and TEMEO derived parameters. Couplets of SVEB were recorded as different by the two monitoring systems, however, with a borderline statistical significance. Conclusions. TEMEO derived ECGs have a very high coincidence rate with standard ECGs. TEMEO patient monitoring provides results that are similar to those derived from a standard Holter ECG.</p>\",\"PeriodicalId\":73519,\"journal\":{\"name\":\"ISRN cardiology\",\"volume\":\"2012 \",\"pages\":\"192670\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2012-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.5402/2012/192670\",\"citationCount\":\"11\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"ISRN cardiology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.5402/2012/192670\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"ISRN cardiology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.5402/2012/192670","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 11

摘要

目的。评估远程监测系统的临床适用性:用于收集和远程监测医疗信息的遥测系统(TEMEO)。方法。我们评估了60例患者,同时应用标准动态心电图和远程监护。进行两种不同的比较:(1)TEMEO心电图与标准12导联心电图;(2) TEMEO动态心电图与标准心电图动态心电图。结果。我们发现TEMEO衍生的心电图与标准心电图的符合率非常高(99.3%)。类内相关系数分析显示,动态心电图和TEMEO衍生参数之间的平均、最大和最小心率、心动过速时间%、单次室上异位搏(SVEB)、单次和双次室上异位搏(VEB)具有高度和显著的相关系数。然而,两种监测系统记录的SVEB对联是不同的,具有临界统计学意义。结论。TEMEO导出的心电图与标准心电图有很高的符合率。TEMEO患者监测提供的结果与标准动态心电图相似。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

Clinical evaluation of a mobile heart rhythm telemonitoring system.

Clinical evaluation of a mobile heart rhythm telemonitoring system.

Clinical evaluation of a mobile heart rhythm telemonitoring system.

Clinical evaluation of a mobile heart rhythm telemonitoring system.

Purpose. To evaluate the clinical applicability of a telemonitoring system: telemetric system for collection and distant surveillance of medical information (TEMEO). Methods. We evaluated 60 patients, applying simultaneously standard Holter ECG and telemonitoring. Two different comparisons were performed: (1) TEMEO ECG with standard 12-lead ECG; (2) TEMEO Holter with standard ECG Holter. Results. We found a very high coincidence rate (99.3%) between TEMEO derived ECGs and standard ECGs. Intraclass correlation coefficient analysis revealed high and significant correlation coefficients regarding average, maximal, and minimal heart rate, % of time in tachycardia, single supraventricular ectopic beats (SVEB), and single and couplets of ventricular ectopic beats (VEB) between Holter ECG and TEMEO derived parameters. Couplets of SVEB were recorded as different by the two monitoring systems, however, with a borderline statistical significance. Conclusions. TEMEO derived ECGs have a very high coincidence rate with standard ECGs. TEMEO patient monitoring provides results that are similar to those derived from a standard Holter ECG.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信