{"title":"医疗设备和护理环境中金属表面处理的可清洁性。","authors":"Jesiska Tandy,&nbsp;Kathy Hanhquynh Le,&nbsp;Garrett Michael Deane,&nbsp;Steven Joseph Burns","doi":"10.2345/1943-5967-56.2.29","DOIUrl":null,"url":null,"abstract":"<p><p>The surface finish of reusable medical devices often is a consideration during design, with attention paid to utility, aesthetics, and cost. To study the cleanability of metal surfaces of varying roughness and finish, soil made of bovine blood and egg yolk was placed on nickel alloy surface of varying average roughness (Ra) values (nominal values of 2-500 μin) and finishes (lapped, ground, milled, profiled, and shape turned). A commercially available wipe consisting of quaternary ammonium compound and less than 20% alcohol was applied for a total of eight wipe cycles to remove the soil. The wipe was applied in both horizontal and vertical directions. Evaluation methods for cleanliness include visual inspection and adenosine triphosphate (ATP) measurement. Rougher surfaces above nominal Ra of 250 were found to have higher ATP readings when wiped in both horizontal and vertical directions. In addition, different surface finishes have different cleanabilities despite similar nominal Ra. To ensure optimal cleaning, surfaces should be cleaned in multiple directions. In the future, similar studies will be coupled with efficacy studies and surfaces made with other materials will be investigated.</p>","PeriodicalId":35656,"journal":{"name":"Biomedical Instrumentation and Technology","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2022-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9767432/pdf/i1943-5967-56-2-29.pdf","citationCount":"0","resultStr":"{\"title\":\"Cleanability of Metal Surface Finishes Found in Medical Devices and the Environment of Care.\",\"authors\":\"Jesiska Tandy,&nbsp;Kathy Hanhquynh Le,&nbsp;Garrett Michael Deane,&nbsp;Steven Joseph Burns\",\"doi\":\"10.2345/1943-5967-56.2.29\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The surface finish of reusable medical devices often is a consideration during design, with attention paid to utility, aesthetics, and cost. To study the cleanability of metal surfaces of varying roughness and finish, soil made of bovine blood and egg yolk was placed on nickel alloy surface of varying average roughness (Ra) values (nominal values of 2-500 μin) and finishes (lapped, ground, milled, profiled, and shape turned). A commercially available wipe consisting of quaternary ammonium compound and less than 20% alcohol was applied for a total of eight wipe cycles to remove the soil. The wipe was applied in both horizontal and vertical directions. Evaluation methods for cleanliness include visual inspection and adenosine triphosphate (ATP) measurement. Rougher surfaces above nominal Ra of 250 were found to have higher ATP readings when wiped in both horizontal and vertical directions. In addition, different surface finishes have different cleanabilities despite similar nominal Ra. To ensure optimal cleaning, surfaces should be cleaned in multiple directions. In the future, similar studies will be coupled with efficacy studies and surfaces made with other materials will be investigated.</p>\",\"PeriodicalId\":35656,\"journal\":{\"name\":\"Biomedical Instrumentation and Technology\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2022-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9767432/pdf/i1943-5967-56-2-29.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Biomedical Instrumentation and Technology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.2345/1943-5967-56.2.29\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biomedical Instrumentation and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2345/1943-5967-56.2.29","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

摘要

可重复使用医疗器械的表面光洁度通常是设计过程中的一个考虑因素,需要注意实用性、美观性和成本。为了研究不同粗糙度和光洁度的金属表面的可清洁性,将牛血液和蛋黄制成的土壤放置在不同平均粗糙度(Ra)值(标称值为2-500 μin)和光洁度(研磨、研磨、铣削、轮廓和车削)的镍合金表面。采用由季铵化合物和少于20%酒精组成的市售擦拭剂,共进行8次擦拭以去除土壤。在水平和垂直方向上都进行了擦拭。洁净度评价方法包括目视检查和三磷酸腺苷(ATP)测定。当在水平和垂直方向上擦拭时,发现标称Ra为250以上的粗糙表面具有更高的ATP读数。此外,尽管标称Ra相似,但不同的表面处理具有不同的清洁能力。为了确保最佳的清洁,表面应从多个方向清洁。在未来,类似的研究将与功效研究相结合,并将研究用其他材料制成的表面。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

Cleanability of Metal Surface Finishes Found in Medical Devices and the Environment of Care.

Cleanability of Metal Surface Finishes Found in Medical Devices and the Environment of Care.

The surface finish of reusable medical devices often is a consideration during design, with attention paid to utility, aesthetics, and cost. To study the cleanability of metal surfaces of varying roughness and finish, soil made of bovine blood and egg yolk was placed on nickel alloy surface of varying average roughness (Ra) values (nominal values of 2-500 μin) and finishes (lapped, ground, milled, profiled, and shape turned). A commercially available wipe consisting of quaternary ammonium compound and less than 20% alcohol was applied for a total of eight wipe cycles to remove the soil. The wipe was applied in both horizontal and vertical directions. Evaluation methods for cleanliness include visual inspection and adenosine triphosphate (ATP) measurement. Rougher surfaces above nominal Ra of 250 were found to have higher ATP readings when wiped in both horizontal and vertical directions. In addition, different surface finishes have different cleanabilities despite similar nominal Ra. To ensure optimal cleaning, surfaces should be cleaned in multiple directions. In the future, similar studies will be coupled with efficacy studies and surfaces made with other materials will be investigated.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Biomedical Instrumentation and Technology
Biomedical Instrumentation and Technology Computer Science-Computer Networks and Communications
CiteScore
1.10
自引率
0.00%
发文量
16
期刊介绍: AAMI publishes Biomedical Instrumentation & Technology (BI&T) a bi-monthly peer-reviewed journal dedicated to the developers, managers, and users of medical instrumentation and technology.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信