雅培C-16000分析仪溶血指数的分析性能及其在测量无浆细胞血红蛋白中的潜在用途。

IF 0.4 4区 医学 Q4 MEDICAL LABORATORY TECHNOLOGY
Anne-Julie Schapira, Klara Lunte, Carole Hennequin, Stéphanie Vicca, Jean-Louis Beaudeux, Rana Alkoury, Valérie Nivet-Antoine, Alexandre Raynor, Charles-Henry Cottart
{"title":"雅培C-16000分析仪溶血指数的分析性能及其在测量无浆细胞血红蛋白中的潜在用途。","authors":"Anne-Julie Schapira,&nbsp;Klara Lunte,&nbsp;Carole Hennequin,&nbsp;Stéphanie Vicca,&nbsp;Jean-Louis Beaudeux,&nbsp;Rana Alkoury,&nbsp;Valérie Nivet-Antoine,&nbsp;Alexandre Raynor,&nbsp;Charles-Henry Cottart","doi":"10.1684/abc.2023.1785","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Measurement of the haemolysis index (HI) is usually performed in clinical chemistry laboratories in order to inform about whether biological analyses are influenced by in vivo or in vitro haemolysis of the specimen. Our aim was to evaluate the analytical performance of Abbott C-16000 analyser HI measurement in order to determine whether this could be used to reliably measure cell-free haemoglobin (fHB) in plasma samples.</p><p><strong>Methods: </strong>The repeatability, reproducibility, lower limit of detection (LLOD) and lower limit of quantification (LLOQ) of C-16000 HI measurement were determined as well as the potential interference of bilirubin, triglycerides and myoglobin. C-16000 HI values of biological samples with various ranges of fHB were compared to those measured using the established reference method, second-derivate spectroscopy.</p><p><strong>Results: </strong>Results: C-16000 HI determination showed excellent linear correlation with the reference method (y = 1.0043x – 1.248, R² = 0.998), a broad analytical measurement range (400-20,000 mg/L; y = 0.9904x + 72.972, R² = 0.999), clinically relevant LLOD (56 mg/L) and LLOQ (84 mg/L), good repeatability (coefficient of variation (CV) = 1-15%) and good reproducibility (CV = 5-7%). No interference was observed with myoglobin at concentrations as high as 35,447 mg/L, unconjugated and conjugated bilirubin (at concentrations up to 500 mg/L and 375 mg/L, respectively) or triglycerides up to 6.8 mmol/L. However, a significant underestimation of fHB concentrations was observed at higher triglyceride levels.</p><p><strong>Conclusion: </strong>This study demonstrates that Abbott C-16000 analyser HI is reliable and accurately measures plasma fHB concentrations under pathophysiological conditions except when there are high blood concentrations of triglycerides.</p>","PeriodicalId":7892,"journal":{"name":"Annales de biologie clinique","volume":"81 1","pages":"44-51"},"PeriodicalIF":0.4000,"publicationDate":"2023-03-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Analytical performance of Abbott C-16000 analyser haemolysis index and its potential use in measuring plasma cell-free haemoglobin.\",\"authors\":\"Anne-Julie Schapira,&nbsp;Klara Lunte,&nbsp;Carole Hennequin,&nbsp;Stéphanie Vicca,&nbsp;Jean-Louis Beaudeux,&nbsp;Rana Alkoury,&nbsp;Valérie Nivet-Antoine,&nbsp;Alexandre Raynor,&nbsp;Charles-Henry Cottart\",\"doi\":\"10.1684/abc.2023.1785\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>Measurement of the haemolysis index (HI) is usually performed in clinical chemistry laboratories in order to inform about whether biological analyses are influenced by in vivo or in vitro haemolysis of the specimen. Our aim was to evaluate the analytical performance of Abbott C-16000 analyser HI measurement in order to determine whether this could be used to reliably measure cell-free haemoglobin (fHB) in plasma samples.</p><p><strong>Methods: </strong>The repeatability, reproducibility, lower limit of detection (LLOD) and lower limit of quantification (LLOQ) of C-16000 HI measurement were determined as well as the potential interference of bilirubin, triglycerides and myoglobin. C-16000 HI values of biological samples with various ranges of fHB were compared to those measured using the established reference method, second-derivate spectroscopy.</p><p><strong>Results: </strong>Results: C-16000 HI determination showed excellent linear correlation with the reference method (y = 1.0043x – 1.248, R² = 0.998), a broad analytical measurement range (400-20,000 mg/L; y = 0.9904x + 72.972, R² = 0.999), clinically relevant LLOD (56 mg/L) and LLOQ (84 mg/L), good repeatability (coefficient of variation (CV) = 1-15%) and good reproducibility (CV = 5-7%). No interference was observed with myoglobin at concentrations as high as 35,447 mg/L, unconjugated and conjugated bilirubin (at concentrations up to 500 mg/L and 375 mg/L, respectively) or triglycerides up to 6.8 mmol/L. However, a significant underestimation of fHB concentrations was observed at higher triglyceride levels.</p><p><strong>Conclusion: </strong>This study demonstrates that Abbott C-16000 analyser HI is reliable and accurately measures plasma fHB concentrations under pathophysiological conditions except when there are high blood concentrations of triglycerides.</p>\",\"PeriodicalId\":7892,\"journal\":{\"name\":\"Annales de biologie clinique\",\"volume\":\"81 1\",\"pages\":\"44-51\"},\"PeriodicalIF\":0.4000,\"publicationDate\":\"2023-03-15\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Annales de biologie clinique\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1684/abc.2023.1785\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annales de biologie clinique","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1684/abc.2023.1785","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

目的:溶血指数(HI)的测量通常在临床化学实验室进行,以了解生物分析是否受到体内或体外溶血标本的影响。我们的目的是评估雅培C-16000分析仪HI测量的分析性能,以确定其是否可用于可靠地测量血浆样品中的无细胞血红蛋白(fHB)。方法:测定C-16000 HI测定的重复性、再现性、检测下限(LLOD)和定量下限(LLOQ),以及胆红素、甘油三酯和肌红蛋白的潜在干扰。将不同fHB范围的生物样品的C-16000 HI值与建立的参考方法-二阶导数光谱法测量的值进行比较。结果:C-16000 HI测定结果与参比法具有良好的线性相关性(y = 1.0043x ~ 1.248, R²= 0.998),分析测量范围宽(400 ~ 20,000 mg/L;y = 0.9904x + 72.972, R²= 0.999),临床相关LLOD (56 mg/L)和LLOQ (84 mg/L),重复性好(变异系数(CV) = 1 ~ 15%),重现性好(CV = 5 ~ 7%)。肌红蛋白浓度高达35,447 mg/L,未偶联和偶联胆红素(浓度分别高达500 mg/L和375 mg/L)或甘油三酯高达6.8 mmol/L时均未观察到干扰。然而,在较高的甘油三酯水平下观察到fHB浓度的显著低估。结论:本研究表明,除了血液中甘油三酯浓度较高的情况外,雅培C-16000分析仪HI在病理生理条件下是可靠的,可以准确测量血浆中fHB浓度。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Analytical performance of Abbott C-16000 analyser haemolysis index and its potential use in measuring plasma cell-free haemoglobin.

Purpose: Measurement of the haemolysis index (HI) is usually performed in clinical chemistry laboratories in order to inform about whether biological analyses are influenced by in vivo or in vitro haemolysis of the specimen. Our aim was to evaluate the analytical performance of Abbott C-16000 analyser HI measurement in order to determine whether this could be used to reliably measure cell-free haemoglobin (fHB) in plasma samples.

Methods: The repeatability, reproducibility, lower limit of detection (LLOD) and lower limit of quantification (LLOQ) of C-16000 HI measurement were determined as well as the potential interference of bilirubin, triglycerides and myoglobin. C-16000 HI values of biological samples with various ranges of fHB were compared to those measured using the established reference method, second-derivate spectroscopy.

Results: Results: C-16000 HI determination showed excellent linear correlation with the reference method (y = 1.0043x – 1.248, R² = 0.998), a broad analytical measurement range (400-20,000 mg/L; y = 0.9904x + 72.972, R² = 0.999), clinically relevant LLOD (56 mg/L) and LLOQ (84 mg/L), good repeatability (coefficient of variation (CV) = 1-15%) and good reproducibility (CV = 5-7%). No interference was observed with myoglobin at concentrations as high as 35,447 mg/L, unconjugated and conjugated bilirubin (at concentrations up to 500 mg/L and 375 mg/L, respectively) or triglycerides up to 6.8 mmol/L. However, a significant underestimation of fHB concentrations was observed at higher triglyceride levels.

Conclusion: This study demonstrates that Abbott C-16000 analyser HI is reliable and accurately measures plasma fHB concentrations under pathophysiological conditions except when there are high blood concentrations of triglycerides.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Annales de biologie clinique
Annales de biologie clinique 医学-医学:研究与实验
CiteScore
0.80
自引率
20.00%
发文量
53
审稿时长
6-12 weeks
期刊介绍: Multidisciplinary information with direct relevance to everyday practice Annales de Biologie Clinique, the official journal of the French Society of Clinical Biology (SFBC), supports biologists in areas including continuing education, laboratory accreditation and technique validation. With original articles, abstracts and accounts of everyday practice, the journal provides details of advances in knowledge, techniques and equipment, as well as a forum for discussion open to the entire community.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信