{"title":"根据ISO 81060-2:2018协议,在普通人群中对配有宽量程袖带的Hingmed DBP-01P临床自动血压计进行了评估。","authors":"Claudio Fania, Antonella Giletto, Valentina Zanello, Paolo Palatini","doi":"10.1097/MBP.0000000000000638","DOIUrl":null,"url":null,"abstract":"<p><p>To determine the accuracy of the Hingmed DBP-01P monitor for clinical automatic blood pressure (BP) measurement according to the International Organization for Standardization (ISO) 81060-2 2018 protocol. The DBP-01P was tested in 85 subjects from the general population (mean age, 69.9 years; 43 men) using a wide-range cuff for arm circumferences from 17 to 42 cm. The mean device-observer difference was 1.32 ± 3.18 mmHg for SBP and 0.78 ± 3.15 mmHg for DBP. These data were in agreement with criterion 1 of the protocol standard requirements (≤5 ± 8 mmHg). Also, criterion 2 was satisfied with the SDs of the 85 participants being well below the maximum values required by the protocol (6.80 and 6.89 mmHg for SBP and DBP pressure, respectively). These data show that the Hingmed BP monitor DBP-01P satisfied the ISO 81060-2:2018 standard requirements for a general population across a wide range of arm sizes using a single wide-range cuff.</p>","PeriodicalId":8950,"journal":{"name":"Blood Pressure Monitoring","volume":"28 2","pages":"116-120"},"PeriodicalIF":1.2000,"publicationDate":"2023-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Validation of the Hingmed DBP-01P clinical automatic blood pressure monitor provided with wide-range cuff, evaluated in a general population according to the ISO 81060-2:2018 protocol.\",\"authors\":\"Claudio Fania, Antonella Giletto, Valentina Zanello, Paolo Palatini\",\"doi\":\"10.1097/MBP.0000000000000638\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>To determine the accuracy of the Hingmed DBP-01P monitor for clinical automatic blood pressure (BP) measurement according to the International Organization for Standardization (ISO) 81060-2 2018 protocol. The DBP-01P was tested in 85 subjects from the general population (mean age, 69.9 years; 43 men) using a wide-range cuff for arm circumferences from 17 to 42 cm. The mean device-observer difference was 1.32 ± 3.18 mmHg for SBP and 0.78 ± 3.15 mmHg for DBP. These data were in agreement with criterion 1 of the protocol standard requirements (≤5 ± 8 mmHg). Also, criterion 2 was satisfied with the SDs of the 85 participants being well below the maximum values required by the protocol (6.80 and 6.89 mmHg for SBP and DBP pressure, respectively). These data show that the Hingmed BP monitor DBP-01P satisfied the ISO 81060-2:2018 standard requirements for a general population across a wide range of arm sizes using a single wide-range cuff.</p>\",\"PeriodicalId\":8950,\"journal\":{\"name\":\"Blood Pressure Monitoring\",\"volume\":\"28 2\",\"pages\":\"116-120\"},\"PeriodicalIF\":1.2000,\"publicationDate\":\"2023-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Blood Pressure Monitoring\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1097/MBP.0000000000000638\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2023/2/17 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q4\",\"JCRName\":\"PERIPHERAL VASCULAR DISEASE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBP.0000000000000638","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/2/17 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
Validation of the Hingmed DBP-01P clinical automatic blood pressure monitor provided with wide-range cuff, evaluated in a general population according to the ISO 81060-2:2018 protocol.
To determine the accuracy of the Hingmed DBP-01P monitor for clinical automatic blood pressure (BP) measurement according to the International Organization for Standardization (ISO) 81060-2 2018 protocol. The DBP-01P was tested in 85 subjects from the general population (mean age, 69.9 years; 43 men) using a wide-range cuff for arm circumferences from 17 to 42 cm. The mean device-observer difference was 1.32 ± 3.18 mmHg for SBP and 0.78 ± 3.15 mmHg for DBP. These data were in agreement with criterion 1 of the protocol standard requirements (≤5 ± 8 mmHg). Also, criterion 2 was satisfied with the SDs of the 85 participants being well below the maximum values required by the protocol (6.80 and 6.89 mmHg for SBP and DBP pressure, respectively). These data show that the Hingmed BP monitor DBP-01P satisfied the ISO 81060-2:2018 standard requirements for a general population across a wide range of arm sizes using a single wide-range cuff.
期刊介绍:
Blood Pressure Monitoring is devoted to original research in blood pressure measurement and blood pressure variability. It includes device technology, analytical methodology of blood pressure over time and its variability, clinical trials - including, but not limited to, pharmacology - involving blood pressure monitoring, blood pressure reactivity, patient evaluation, and outcomes and effectiveness research.
This innovative journal contains papers dealing with all aspects of manual, automated, and ambulatory monitoring. Basic and clinical science papers are considered although the emphasis is on clinical medicine.
Submitted articles undergo a preliminary review by the editor. Some articles may be returned to authors without further consideration. Those being considered for publication will undergo further assessment and peer-review by the editors and those invited to do so from a reviewer pool.