{"title":"高危前列腺癌轮换设置误差的剂量学影响","authors":"M. Poncyljusz, W. Bulski","doi":"10.2478/pjmpe-2022-0023","DOIUrl":null,"url":null,"abstract":"Abstract Introduction: Cone-beam computed tomography (CBCT) provides an excellent solution to quantitative assessment and correction of patient set-up errors during radiotherapy. However, most linear accelerators are equipped with conventional therapy tables that can be moved in three translational directions and perform only yaw rotation. Uncorrected roll and pitch result in rotational set-up errors, particularly when the distance from the isocenter to the target border is large. The aim of this study was to investigate the impact of rotational errors on the dose delivered to the clinical target volume (CTV), the planning target volume (PTV) and organs at risk (OAR). Material and methods: 30 patients with prostate cancer treated with VMAT technique had daily CBCT scans (840 CBCTs in total) prior to treatment delivery. The rotational errors remaining after on-line correction were retrospectively analysed. The sum plans simulating the dose distribution during the treatment course were calculated for selected patients with significant rotational errors. Results: The dose delivered to the prostate bed CTV reported in the sum plan was not lower than in the original plan for all selected patients. For four patients from the selected group, the D98% for prostate bed PTV was less than 95%. The V47.88Gy for pelvic lymph nodes PTV was less than 98% for two of the selected patients. Conclusions: The analysis of the dosimetric parameters showed that the impact of uncorrected rotations is not clinically significant in terms of the dose delivered to OAR and the dose coverage of CTV. However, the PTV dose coverage is correlated with distance away from the isocenter and is smaller than planned.","PeriodicalId":53955,"journal":{"name":"Polish Journal of Medical Physics and Engineering","volume":"1 1","pages":"200 - 206"},"PeriodicalIF":0.7000,"publicationDate":"2022-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Dosimetric impact of rotational set-up errors in high-risk prostate cancer\",\"authors\":\"M. Poncyljusz, W. Bulski\",\"doi\":\"10.2478/pjmpe-2022-0023\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Abstract Introduction: Cone-beam computed tomography (CBCT) provides an excellent solution to quantitative assessment and correction of patient set-up errors during radiotherapy. However, most linear accelerators are equipped with conventional therapy tables that can be moved in three translational directions and perform only yaw rotation. Uncorrected roll and pitch result in rotational set-up errors, particularly when the distance from the isocenter to the target border is large. The aim of this study was to investigate the impact of rotational errors on the dose delivered to the clinical target volume (CTV), the planning target volume (PTV) and organs at risk (OAR). Material and methods: 30 patients with prostate cancer treated with VMAT technique had daily CBCT scans (840 CBCTs in total) prior to treatment delivery. The rotational errors remaining after on-line correction were retrospectively analysed. The sum plans simulating the dose distribution during the treatment course were calculated for selected patients with significant rotational errors. Results: The dose delivered to the prostate bed CTV reported in the sum plan was not lower than in the original plan for all selected patients. For four patients from the selected group, the D98% for prostate bed PTV was less than 95%. The V47.88Gy for pelvic lymph nodes PTV was less than 98% for two of the selected patients. Conclusions: The analysis of the dosimetric parameters showed that the impact of uncorrected rotations is not clinically significant in terms of the dose delivered to OAR and the dose coverage of CTV. However, the PTV dose coverage is correlated with distance away from the isocenter and is smaller than planned.\",\"PeriodicalId\":53955,\"journal\":{\"name\":\"Polish Journal of Medical Physics and Engineering\",\"volume\":\"1 1\",\"pages\":\"200 - 206\"},\"PeriodicalIF\":0.7000,\"publicationDate\":\"2022-12-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Polish Journal of Medical Physics and Engineering\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.2478/pjmpe-2022-0023\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Polish Journal of Medical Physics and Engineering","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2478/pjmpe-2022-0023","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

摘要简介:锥形束计算机断层扫描(CBCT)为放射治疗期间患者设置错误的定量评估和纠正提供了极好的解决方案。然而,大多数线性加速器都配备了传统的治疗台,可以在三个平移方向上移动,只执行偏航旋转。未校正的滚转和俯仰会导致旋转设置错误,特别是当从等中心到目标边界的距离很大时。本研究的目的是探讨旋转误差对临床靶体积(CTV)、计划靶体积(PTV)和危险器官(OAR)剂量的影响。材料和方法:30例接受VMAT技术治疗的前列腺癌患者在治疗前每日进行CBCT扫描(共840次)。对在线校正后的旋转误差进行回顾性分析。选取旋转误差较大的患者,计算模拟治疗过程中剂量分布的总和计划。结果:在所有入选患者中,综合方案中报告的前列腺床CTV剂量均不低于原方案。其中4例患者前列腺床PTV的D98%小于95%。其中2例患者盆腔淋巴结PTV V47.88Gy小于98%。结论:剂量学参数分析显示,未校正旋转对OAR的剂量和CTV的剂量覆盖的影响没有临床意义。然而,PTV剂量覆盖与离等中心的距离相关,比计划的要小。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Dosimetric impact of rotational set-up errors in high-risk prostate cancer
Abstract Introduction: Cone-beam computed tomography (CBCT) provides an excellent solution to quantitative assessment and correction of patient set-up errors during radiotherapy. However, most linear accelerators are equipped with conventional therapy tables that can be moved in three translational directions and perform only yaw rotation. Uncorrected roll and pitch result in rotational set-up errors, particularly when the distance from the isocenter to the target border is large. The aim of this study was to investigate the impact of rotational errors on the dose delivered to the clinical target volume (CTV), the planning target volume (PTV) and organs at risk (OAR). Material and methods: 30 patients with prostate cancer treated with VMAT technique had daily CBCT scans (840 CBCTs in total) prior to treatment delivery. The rotational errors remaining after on-line correction were retrospectively analysed. The sum plans simulating the dose distribution during the treatment course were calculated for selected patients with significant rotational errors. Results: The dose delivered to the prostate bed CTV reported in the sum plan was not lower than in the original plan for all selected patients. For four patients from the selected group, the D98% for prostate bed PTV was less than 95%. The V47.88Gy for pelvic lymph nodes PTV was less than 98% for two of the selected patients. Conclusions: The analysis of the dosimetric parameters showed that the impact of uncorrected rotations is not clinically significant in terms of the dose delivered to OAR and the dose coverage of CTV. However, the PTV dose coverage is correlated with distance away from the isocenter and is smaller than planned.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Polish Journal of Medical Physics and Engineering
Polish Journal of Medical Physics and Engineering RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
1.30
自引率
0.00%
发文量
19
期刊介绍: Polish Journal of Medical Physics and Engineering (PJMPE) (Online ISSN: 1898-0309; Print ISSN: 1425-4689) is an official publication of the Polish Society of Medical Physics. It is a peer-reviewed, open access scientific journal with no publication fees. The issues are published quarterly online. The Journal publishes original contribution in medical physics and biomedical engineering.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信