肱骨骨干无菌假关节的外科治疗。研究力学套装

D. Roca Romalde , D. Lacroix , V.L. Caja López , I. Proubasta Renart , J.A. Planell Estany
{"title":"肱骨骨干无菌假关节的外科治疗。研究力学套装","authors":"D. Roca Romalde ,&nbsp;D. Lacroix ,&nbsp;V.L. Caja López ,&nbsp;I. Proubasta Renart ,&nbsp;J.A. Planell Estany","doi":"10.1016/S0482-5985(07)74574-1","DOIUrl":null,"url":null,"abstract":"<div><h3>Objetivos</h3><p>Comparar la rigidez y la distribución de tensiones en los implantes de dos modelos experimentales: un húmero con pseudoartrosis diafisaria estabilizado con placa y un húmero con pseudoartrosis diafisaria estabilizado con clavo encerrojado.</p></div><div><h3>Material y método</h3><p>Con un húmero de cadáver y los dos dispositivos de fijación se crearon las geometrías con el programa de diseño CATIA v4.2<sup>®</sup> (IBM, Armonk, USA). Posteriormente estas geometrías se modelaron con el preprocesador informático MSC-PATRAN<sup>®</sup> (IBM, Armonk, USA). Finalmente se establecieron las propiedades mecánicas de los materiales, las condiciones de contorno y las cargas a las que fueron sometidos los modelos.</p></div><div><h3>Resultados</h3><p>El modelo con clavo fue más rígido que el modelo con placa a compresión, tracción y torsión, sin embargo, el modelo con placa fue más rígido que el modelo con clavo en flexión anteroposterior, lateromedial y cizallamiento lateromedial. La distribución de tensiones ha sido más heterogénea en el clavo que en la placa, siendo esta última el implante que soporta los mayores valores tensionales en todos los estados de carga estudiados.</p></div><div><h3>Conclusiones</h3><p>La rigidez del modelo experimental depende no solo del implante sino del estado de carga aplicado, siendo superior el clavo en unas condiciones de carga y la placa en otras. El clavo sin embargo, absorbe menos tensión que la placa en todos los estados de carga estudiados.</p></div><div><h3>Purpose</h3><p>To compare the firmess and stress distribution patterns in the implants of two experimental models: a humerus with shaft pseudoarthrosis stabilized with a plate, and a humerus with shaft nonunions stabilized with a locking nail.</p></div><div><h3>Materials and methods</h3><p>The two fixation devices are attached to cadaver humerus bones; geometries are created with the CATIA 4.2 design software (IBM, Armonk, USA). Subsequently, these geometries were modelled with the MSC PATRAN<sup>®</sup> computer processor (IBM, Armonk, USA). Finally, the mechanical properties of the materials were established as well as the contour properties and the loads the models were subjected to.</p></div><div><h3>Results</h3><p>The nailed model was firmer that the plated model as regards compression, traction and torsion. Nonetheless, as far as A/P and lateral-medial flexion and lateral-medial shear stresses were concerned, the plated model was firmer than the nailed model. Stress distribution was more heterogeneous in the nail than in the plate, the latter being the implant type supporting the highest stress levels in all the loading phases studied.</p></div><div><h3>Conclusions</h3><p>For some stress levels, the nail proved to be more stable than the plate, although for other stress levels the opposite was the case. In any case, the nail provides a better stress distribution than the plate.</p></div>","PeriodicalId":101102,"journal":{"name":"Revista de Ortopedia y Traumatología","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2007-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1016/S0482-5985(07)74574-1","citationCount":"1","resultStr":"{\"title\":\"Tratamiento quirúrgico de las pseudoartrosis asépticas de diáfisis humeral. Estudio biomecánico\",\"authors\":\"D. Roca Romalde ,&nbsp;D. Lacroix ,&nbsp;V.L. Caja López ,&nbsp;I. Proubasta Renart ,&nbsp;J.A. Planell Estany\",\"doi\":\"10.1016/S0482-5985(07)74574-1\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Objetivos</h3><p>Comparar la rigidez y la distribución de tensiones en los implantes de dos modelos experimentales: un húmero con pseudoartrosis diafisaria estabilizado con placa y un húmero con pseudoartrosis diafisaria estabilizado con clavo encerrojado.</p></div><div><h3>Material y método</h3><p>Con un húmero de cadáver y los dos dispositivos de fijación se crearon las geometrías con el programa de diseño CATIA v4.2<sup>®</sup> (IBM, Armonk, USA). Posteriormente estas geometrías se modelaron con el preprocesador informático MSC-PATRAN<sup>®</sup> (IBM, Armonk, USA). Finalmente se establecieron las propiedades mecánicas de los materiales, las condiciones de contorno y las cargas a las que fueron sometidos los modelos.</p></div><div><h3>Resultados</h3><p>El modelo con clavo fue más rígido que el modelo con placa a compresión, tracción y torsión, sin embargo, el modelo con placa fue más rígido que el modelo con clavo en flexión anteroposterior, lateromedial y cizallamiento lateromedial. La distribución de tensiones ha sido más heterogénea en el clavo que en la placa, siendo esta última el implante que soporta los mayores valores tensionales en todos los estados de carga estudiados.</p></div><div><h3>Conclusiones</h3><p>La rigidez del modelo experimental depende no solo del implante sino del estado de carga aplicado, siendo superior el clavo en unas condiciones de carga y la placa en otras. El clavo sin embargo, absorbe menos tensión que la placa en todos los estados de carga estudiados.</p></div><div><h3>Purpose</h3><p>To compare the firmess and stress distribution patterns in the implants of two experimental models: a humerus with shaft pseudoarthrosis stabilized with a plate, and a humerus with shaft nonunions stabilized with a locking nail.</p></div><div><h3>Materials and methods</h3><p>The two fixation devices are attached to cadaver humerus bones; geometries are created with the CATIA 4.2 design software (IBM, Armonk, USA). Subsequently, these geometries were modelled with the MSC PATRAN<sup>®</sup> computer processor (IBM, Armonk, USA). Finally, the mechanical properties of the materials were established as well as the contour properties and the loads the models were subjected to.</p></div><div><h3>Results</h3><p>The nailed model was firmer that the plated model as regards compression, traction and torsion. Nonetheless, as far as A/P and lateral-medial flexion and lateral-medial shear stresses were concerned, the plated model was firmer than the nailed model. Stress distribution was more heterogeneous in the nail than in the plate, the latter being the implant type supporting the highest stress levels in all the loading phases studied.</p></div><div><h3>Conclusions</h3><p>For some stress levels, the nail proved to be more stable than the plate, although for other stress levels the opposite was the case. In any case, the nail provides a better stress distribution than the plate.</p></div>\",\"PeriodicalId\":101102,\"journal\":{\"name\":\"Revista de Ortopedia y Traumatología\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2007-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1016/S0482-5985(07)74574-1\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Revista de Ortopedia y Traumatología\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0482598507745741\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Revista de Ortopedia y Traumatología","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0482598507745741","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

摘要

目的比较两种实验模型的种植体刚度和应力分布:钢板稳定的肱骨假骨和环钉稳定的肱骨假骨。材料和方法尸体肱骨和两个固定装置的几何形状是使用CATIA v4.2®设计程序(IBM, Armonk,美国)创建的。这些几何图形随后用计算机预处理器MSC-PATRAN®(IBM, Armonk,美国)建模。最后,建立了材料的力学性能、边界条件和模型所受的载荷。结果钉模型在前后弯曲、外侧内侧和外侧内侧剪切方面比钉模型更刚性,而板模型在前后弯曲、外侧内侧剪切方面比钉模型更刚性。在本研究中,我们分析了两种不同的种植体在不同载荷状态下的应力分布。实验模型的刚度不仅取决于种植体的刚度,还取决于所施加的载荷状态,在某些载荷条件下钉子优越,在其他载荷条件下钢板优越。然而,在所有研究的负载状态下,钉子吸收的应力比板少。PurposeTo compare the firmess patterns and应激distribution in the implants of two试点模式:a humerus with shaft pseudoarthrosis枪榴弹with a plate, and a humerus with shaft nonunions枪榴弹with a locking钉。两种固定装置附在肱骨上的材料和方法;= =地理= =根据美国人口普查,这个县的面积为。随后,这些几何图形被建模与MSC PATRAN®计算机处理器(IBM, Armonk,美国)。设立的最后,the mechanical properties of the materials as well as the contour properties and the立即the models是经常会。ResultsThe nailed model was firmer that the plated model期盼compression, traction and torsion。然而,as far as A / P and lateral-medial flexion and lateral-medial shear库尔特是关心的,the plated model was firmer than the nailed model。压力分布was more heterogeneous in the钉than in the plate,后一being the implant的压力水平的佐证type in all the loading阶段studied。ConclusionsFor一些应激水平,the钉证明to be more瘀青than the plate尽管for other压力水平was the opposite the case。在任何情况下,钉子都比板提供更好的应力分布。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Tratamiento quirúrgico de las pseudoartrosis asépticas de diáfisis humeral. Estudio biomecánico

Objetivos

Comparar la rigidez y la distribución de tensiones en los implantes de dos modelos experimentales: un húmero con pseudoartrosis diafisaria estabilizado con placa y un húmero con pseudoartrosis diafisaria estabilizado con clavo encerrojado.

Material y método

Con un húmero de cadáver y los dos dispositivos de fijación se crearon las geometrías con el programa de diseño CATIA v4.2® (IBM, Armonk, USA). Posteriormente estas geometrías se modelaron con el preprocesador informático MSC-PATRAN® (IBM, Armonk, USA). Finalmente se establecieron las propiedades mecánicas de los materiales, las condiciones de contorno y las cargas a las que fueron sometidos los modelos.

Resultados

El modelo con clavo fue más rígido que el modelo con placa a compresión, tracción y torsión, sin embargo, el modelo con placa fue más rígido que el modelo con clavo en flexión anteroposterior, lateromedial y cizallamiento lateromedial. La distribución de tensiones ha sido más heterogénea en el clavo que en la placa, siendo esta última el implante que soporta los mayores valores tensionales en todos los estados de carga estudiados.

Conclusiones

La rigidez del modelo experimental depende no solo del implante sino del estado de carga aplicado, siendo superior el clavo en unas condiciones de carga y la placa en otras. El clavo sin embargo, absorbe menos tensión que la placa en todos los estados de carga estudiados.

Purpose

To compare the firmess and stress distribution patterns in the implants of two experimental models: a humerus with shaft pseudoarthrosis stabilized with a plate, and a humerus with shaft nonunions stabilized with a locking nail.

Materials and methods

The two fixation devices are attached to cadaver humerus bones; geometries are created with the CATIA 4.2 design software (IBM, Armonk, USA). Subsequently, these geometries were modelled with the MSC PATRAN® computer processor (IBM, Armonk, USA). Finally, the mechanical properties of the materials were established as well as the contour properties and the loads the models were subjected to.

Results

The nailed model was firmer that the plated model as regards compression, traction and torsion. Nonetheless, as far as A/P and lateral-medial flexion and lateral-medial shear stresses were concerned, the plated model was firmer than the nailed model. Stress distribution was more heterogeneous in the nail than in the plate, the latter being the implant type supporting the highest stress levels in all the loading phases studied.

Conclusions

For some stress levels, the nail proved to be more stable than the plate, although for other stress levels the opposite was the case. In any case, the nail provides a better stress distribution than the plate.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信