{"title":"印度手术室麻醉工作区域开放式系统填充适配器的污染率:一项体外研究","authors":"A. Jain, N. Sharma, Niranjan Patil","doi":"10.33545/26648849.2021.v3.i1a.17","DOIUrl":null,"url":null,"abstract":"Objective: To evaluate the rate and type of contamination in open-system adapters used for filling the anesthetic agent sevoflurane across hospitals in India. Methods: Nineteen reusable adapters used for filling sevoflurane were collected from nine hospitals in India and evaluated for microbial contamination. Samples were collected by flushing the inner lumen and swabbing the outer surfaces of adapters and dispatched to a central laboratory in sealed sterile tubes/containers for culture and microbial identification. Results: Out of 19 swab samples collected from the outer surfaces, 11 (57.9%) were found to be contaminated/colonized. Out of these, 5 (26.3%) were found to be contaminated with Gram-positive cocci belonging to the Staphylococcus species, and 6 (31.6%), with Gram-negative organisms Pseudomonas aeruginosa (5 [26.3%]) , and Burkholderia cepacian (1 [5%]). Out of 19 rinse samples from the inner lumens, 8 (42.1%) were found to be contaminated, 2 (10.5%) with Gram-positive cocci belonging to Staphylococcus species and 6 (31.6%) with Gram-negative organisms Achromobacter denitrificans and Aeromonas enteropelogene (one (5.3%) sample each), and Pseudomonas aeruginosa and Ralstonia pickettii (2 (10.5%) samples each). Conclusion: Reusable open-system filling adapters are susceptible to contamination with nosocomial pathogens; therefore, single-use adapters to inhibit contamination and prevent hospital ‐ acquired infections should be encouraged.","PeriodicalId":91883,"journal":{"name":"International journal of anesthesiology & research","volume":"1 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2021-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Contamination rate in open-system filling adapters in operating room anesthesia work area in India: An in-vitro study\",\"authors\":\"A. Jain, N. Sharma, Niranjan Patil\",\"doi\":\"10.33545/26648849.2021.v3.i1a.17\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Objective: To evaluate the rate and type of contamination in open-system adapters used for filling the anesthetic agent sevoflurane across hospitals in India. Methods: Nineteen reusable adapters used for filling sevoflurane were collected from nine hospitals in India and evaluated for microbial contamination. Samples were collected by flushing the inner lumen and swabbing the outer surfaces of adapters and dispatched to a central laboratory in sealed sterile tubes/containers for culture and microbial identification. Results: Out of 19 swab samples collected from the outer surfaces, 11 (57.9%) were found to be contaminated/colonized. Out of these, 5 (26.3%) were found to be contaminated with Gram-positive cocci belonging to the Staphylococcus species, and 6 (31.6%), with Gram-negative organisms Pseudomonas aeruginosa (5 [26.3%]) , and Burkholderia cepacian (1 [5%]). Out of 19 rinse samples from the inner lumens, 8 (42.1%) were found to be contaminated, 2 (10.5%) with Gram-positive cocci belonging to Staphylococcus species and 6 (31.6%) with Gram-negative organisms Achromobacter denitrificans and Aeromonas enteropelogene (one (5.3%) sample each), and Pseudomonas aeruginosa and Ralstonia pickettii (2 (10.5%) samples each). Conclusion: Reusable open-system filling adapters are susceptible to contamination with nosocomial pathogens; therefore, single-use adapters to inhibit contamination and prevent hospital ‐ acquired infections should be encouraged.\",\"PeriodicalId\":91883,\"journal\":{\"name\":\"International journal of anesthesiology & research\",\"volume\":\"1 1\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2021-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"International journal of anesthesiology & research\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.33545/26648849.2021.v3.i1a.17\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"International journal of anesthesiology & research","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.33545/26648849.2021.v3.i1a.17","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

目的:评价印度各医院用于填充麻醉剂七氟醚的开放式系统适配器的污染率和类型。方法:从印度9家医院收集19个可重复使用的七氟醚填充适配器,并对其微生物污染进行评估。通过冲洗内腔和擦拭适配器的外表面来收集样本,并将样本放入密封的无菌管/容器中送往中心实验室进行培养和微生物鉴定。结果:在外表面采集19份拭子样本,11份(57.9%)被感染/定植。其中,5例(26.3%)感染了属于葡萄球菌种类的革兰氏阳性球菌,6例(31.6%)感染了革兰氏阴性菌铜绿假单胞菌(5例[26.3%])和伯克霍尔德菌(1例[5%])。19份内腔冲洗标本中,检出污染8份(42.1%),其中葡萄球菌属革兰氏阳性球菌2份(10.5%),革兰氏阴性菌反硝化无色杆菌和肠产气单胞菌各1份(5.3%),铜绿假单胞菌和皮氏Ralstonia各2份(10.5%)。结论:可重复使用的开放式系统灌装适配器容易受到医院病原体的污染;因此,应该鼓励使用一次性适配器来抑制污染和防止医院获得性感染。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Contamination rate in open-system filling adapters in operating room anesthesia work area in India: An in-vitro study
Objective: To evaluate the rate and type of contamination in open-system adapters used for filling the anesthetic agent sevoflurane across hospitals in India. Methods: Nineteen reusable adapters used for filling sevoflurane were collected from nine hospitals in India and evaluated for microbial contamination. Samples were collected by flushing the inner lumen and swabbing the outer surfaces of adapters and dispatched to a central laboratory in sealed sterile tubes/containers for culture and microbial identification. Results: Out of 19 swab samples collected from the outer surfaces, 11 (57.9%) were found to be contaminated/colonized. Out of these, 5 (26.3%) were found to be contaminated with Gram-positive cocci belonging to the Staphylococcus species, and 6 (31.6%), with Gram-negative organisms Pseudomonas aeruginosa (5 [26.3%]) , and Burkholderia cepacian (1 [5%]). Out of 19 rinse samples from the inner lumens, 8 (42.1%) were found to be contaminated, 2 (10.5%) with Gram-positive cocci belonging to Staphylococcus species and 6 (31.6%) with Gram-negative organisms Achromobacter denitrificans and Aeromonas enteropelogene (one (5.3%) sample each), and Pseudomonas aeruginosa and Ralstonia pickettii (2 (10.5%) samples each). Conclusion: Reusable open-system filling adapters are susceptible to contamination with nosocomial pathogens; therefore, single-use adapters to inhibit contamination and prevent hospital ‐ acquired infections should be encouraged.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信