Xiangli Cui, K. Jee, Hongzhi Wang, Man Hu, Hsiao-Ming Lu
{"title":"质子治疗中CT霍斯菲尔德单位转换及相对停止力的研究及脂肪组织改善","authors":"Xiangli Cui, K. Jee, Hongzhi Wang, Man Hu, Hsiao-Ming Lu","doi":"10.3760/CMA.J.ISSN.0254-5098.2019.10.011","DOIUrl":null,"url":null,"abstract":"Objective \nTo introduce a method for calibrating the conversion from CT Hounsfield units (HU) to relative stopping power (RSP) for proton therapy, and improve the precision of the conversion in the region for adipose tissues. \n \n \nMethods \nThe HU and RSP values of human tissues were calculated by a stoichiometric calibration method. Animal tissue was used to simulate subcutaneous adipose tissue of patients, and the HU and RSP of the animal tissue were measured. The effect of subcutaneous adipose tissue on conversion between HU and RSP were analyzed by piecewise fitting. \n \n \nResults \nThe precision of conversion curve was improved significantly with the measured HU and RSP of adipose tissue in the fitting. The effect caused by different choice in different ionization energy was less than 0.6%, and the effect of proton energy differential was less than 0.8%. \n \n \nConclusions \nThe precision of conversion curve for the transformation of HU into RSP in adipose tissues could be improved by taking subcutaneous adipose tissue into account, which would reduce the range error of proton beams when such tissues are present in the target volumes or in the beam path. \n \n \nKey words: \nProton therapy; Relative stopping power; Stoichiometric calibration method","PeriodicalId":36403,"journal":{"name":"中华放射医学与防护杂志","volume":"81 1","pages":"772-777"},"PeriodicalIF":0.0000,"publicationDate":"2019-10-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Study on the conversion between CT hounsfield units and relative stopping power in proton therapy and the improvement in adipose tissue\",\"authors\":\"Xiangli Cui, K. Jee, Hongzhi Wang, Man Hu, Hsiao-Ming Lu\",\"doi\":\"10.3760/CMA.J.ISSN.0254-5098.2019.10.011\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Objective \\nTo introduce a method for calibrating the conversion from CT Hounsfield units (HU) to relative stopping power (RSP) for proton therapy, and improve the precision of the conversion in the region for adipose tissues. \\n \\n \\nMethods \\nThe HU and RSP values of human tissues were calculated by a stoichiometric calibration method. Animal tissue was used to simulate subcutaneous adipose tissue of patients, and the HU and RSP of the animal tissue were measured. The effect of subcutaneous adipose tissue on conversion between HU and RSP were analyzed by piecewise fitting. \\n \\n \\nResults \\nThe precision of conversion curve was improved significantly with the measured HU and RSP of adipose tissue in the fitting. The effect caused by different choice in different ionization energy was less than 0.6%, and the effect of proton energy differential was less than 0.8%. \\n \\n \\nConclusions \\nThe precision of conversion curve for the transformation of HU into RSP in adipose tissues could be improved by taking subcutaneous adipose tissue into account, which would reduce the range error of proton beams when such tissues are present in the target volumes or in the beam path. \\n \\n \\nKey words: \\nProton therapy; Relative stopping power; Stoichiometric calibration method\",\"PeriodicalId\":36403,\"journal\":{\"name\":\"中华放射医学与防护杂志\",\"volume\":\"81 1\",\"pages\":\"772-777\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2019-10-25\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"中华放射医学与防护杂志\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.3760/CMA.J.ISSN.0254-5098.2019.10.011\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"中华放射医学与防护杂志","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3760/CMA.J.ISSN.0254-5098.2019.10.011","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
Study on the conversion between CT hounsfield units and relative stopping power in proton therapy and the improvement in adipose tissue
Objective
To introduce a method for calibrating the conversion from CT Hounsfield units (HU) to relative stopping power (RSP) for proton therapy, and improve the precision of the conversion in the region for adipose tissues.
Methods
The HU and RSP values of human tissues were calculated by a stoichiometric calibration method. Animal tissue was used to simulate subcutaneous adipose tissue of patients, and the HU and RSP of the animal tissue were measured. The effect of subcutaneous adipose tissue on conversion between HU and RSP were analyzed by piecewise fitting.
Results
The precision of conversion curve was improved significantly with the measured HU and RSP of adipose tissue in the fitting. The effect caused by different choice in different ionization energy was less than 0.6%, and the effect of proton energy differential was less than 0.8%.
Conclusions
The precision of conversion curve for the transformation of HU into RSP in adipose tissues could be improved by taking subcutaneous adipose tissue into account, which would reduce the range error of proton beams when such tissues are present in the target volumes or in the beam path.
Key words:
Proton therapy; Relative stopping power; Stoichiometric calibration method