西门子E.Cam®Spect系统在加纳Korle-Bu教学医院的性能评估

E. Sosu, F. Hasford, J. Amuasi
{"title":"西门子E.Cam®Spect系统在加纳Korle-Bu教学医院的性能评估","authors":"E. Sosu, F. Hasford, J. Amuasi","doi":"10.4314/JAST.V16I1-2.64789","DOIUrl":null,"url":null,"abstract":"Performance evaluation tests on Siemens e.camR Signature Series Single Photon Emission Computed Tomogra-phy (SPECT) System at Korle-Bu Teaching Hospital (Ghana) were conducted for compliance with Manufactur-erfs specifications and usersf requirements. The tests were performed according to National Electrical Manufac-turers Association protocols, and different measuring instruments and phantoms containing 99m-Tc were used. Extrinsic uniformity, System energy resolution, System spatial resolution (without scattering), Detector shield leakage and System planar sensitivity were evaluated. The System energy resolution measured was 9.38 % as against manufacturersf specification of . 9.9 %, Planar sensitivity value measured was 232.92 c/min/mCi while manufacturersf specification was 202 c/min/mCi, and Detector shield leakage values measured were 0.001 %, 0.022 % and 0.025 % against an accepted value of < 1 %. The performance evaluation confirmed that the Syst-em met requirements for clinical nuclear medicine imaging, and the values could be used as database for future quality control measurements.","PeriodicalId":9207,"journal":{"name":"British Journal of Applied Science and Technology","volume":"39 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2011-03-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Performance evaluation of Siemens E.Cam® Spect System at Korle-Bu Teaching Hospital (Ghana\",\"authors\":\"E. Sosu, F. Hasford, J. Amuasi\",\"doi\":\"10.4314/JAST.V16I1-2.64789\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Performance evaluation tests on Siemens e.camR Signature Series Single Photon Emission Computed Tomogra-phy (SPECT) System at Korle-Bu Teaching Hospital (Ghana) were conducted for compliance with Manufactur-erfs specifications and usersf requirements. The tests were performed according to National Electrical Manufac-turers Association protocols, and different measuring instruments and phantoms containing 99m-Tc were used. Extrinsic uniformity, System energy resolution, System spatial resolution (without scattering), Detector shield leakage and System planar sensitivity were evaluated. The System energy resolution measured was 9.38 % as against manufacturersf specification of . 9.9 %, Planar sensitivity value measured was 232.92 c/min/mCi while manufacturersf specification was 202 c/min/mCi, and Detector shield leakage values measured were 0.001 %, 0.022 % and 0.025 % against an accepted value of < 1 %. The performance evaluation confirmed that the Syst-em met requirements for clinical nuclear medicine imaging, and the values could be used as database for future quality control measurements.\",\"PeriodicalId\":9207,\"journal\":{\"name\":\"British Journal of Applied Science and Technology\",\"volume\":\"39 1\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2011-03-25\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"British Journal of Applied Science and Technology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.4314/JAST.V16I1-2.64789\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"British Journal of Applied Science and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4314/JAST.V16I1-2.64789","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

在加纳Korle-Bu教学医院对西门子e.camR签名系列单光子发射计算机断层扫描(SPECT)系统进行了性能评估测试,以符合制造商的规范和用户的要求。测试是根据国家电气制造商协会的协议进行的,使用了不同的测量仪器和含有99m-Tc的样品。评估了系统的外部均匀性、系统能量分辨率、系统空间分辨率(无散射)、探测器屏蔽泄漏和系统平面灵敏度。系统的能量分辨率为9.38%,与制造商的规格相符。测量的平面灵敏度值为232.92 c/min/mCi,而制造商的规格为202 c/min/mCi;测量的探测器屏蔽泄漏值分别为0.001 %,0.022%和0.025%,而可接受值< 1%。性能评价证实该系统满足临床核医学成像的要求,其值可作为将来质量控制测量的数据库。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Performance evaluation of Siemens E.Cam® Spect System at Korle-Bu Teaching Hospital (Ghana
Performance evaluation tests on Siemens e.camR Signature Series Single Photon Emission Computed Tomogra-phy (SPECT) System at Korle-Bu Teaching Hospital (Ghana) were conducted for compliance with Manufactur-erfs specifications and usersf requirements. The tests were performed according to National Electrical Manufac-turers Association protocols, and different measuring instruments and phantoms containing 99m-Tc were used. Extrinsic uniformity, System energy resolution, System spatial resolution (without scattering), Detector shield leakage and System planar sensitivity were evaluated. The System energy resolution measured was 9.38 % as against manufacturersf specification of . 9.9 %, Planar sensitivity value measured was 232.92 c/min/mCi while manufacturersf specification was 202 c/min/mCi, and Detector shield leakage values measured were 0.001 %, 0.022 % and 0.025 % against an accepted value of < 1 %. The performance evaluation confirmed that the Syst-em met requirements for clinical nuclear medicine imaging, and the values could be used as database for future quality control measurements.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信