{"title":"丹麦心脏植入式电子装置患者心脏手术后的铅并发症:一项全国嵌套病例对照研究","authors":"M. Frausing, J. Nielsen, J. Johansen, OD Jorgensen, J. Kristensen, T. Olsen, C. Gerdes, M. Kronborg","doi":"10.1093/europace/euac053.528","DOIUrl":null,"url":null,"abstract":"\n \n \n Type of funding sources: Private grant(s) and/or Sponsorship. Main funding source(s): Karen Elise Jensen’s Foundation\n \n \n \n Manipulation of the heart during cardiac surgery in patients with cardiac implantable electronic devices (CIEDs) may result in lead damage or -displacement, but whether cardiac surgery truly infers an excess risk of lead failure is not known.\n \n \n \n The objective of this study was to examine risk of lead complications as a result of cardiac surgery in CIED patients.\n \n \n \n We conducted a nationwide nested case-control study. Our source population comprised all patients ≥18 of age who underwent a de novo CIED implantation in Denmark between 1998 and 2017. Only patients who were alive and event-free at six months after implantation were eligible for inclusion in the case-control study. Cases were defined as incident lead complications resulting in a revision procedure, and were matched 1:30 to controls using risk set sampling. We used conditional logistic regression to estimate incidence rate ratios (IRRs) for the association between cardiac surgery and lead complications.\n \n \n \n Our population consisted of 61,869 de novo CIED patient. We identified 1324 incident cases of lead complications and 39,708 time-matched controls. 1036 cases and controls underwent cardiac surgery during follow-up; 61 cases (4.5%) and 975 controls (2.5%). Risk of lead-related reoperation was highest within the first six months of cardiac surgery (IRR 10.1, 95% CI 6.5-15.6, adjusted IRR 11.0, 95% CI 7-17.4). At one year, we observed no association between cardiac surgery and lead complications (IRR: 1.0, 95% CI 0.7-1.5 and adjusted IRR: 1.1, 95% CI 0.7-1.6).\n \n \n \n Cardiac surgery was associated with a substantial risk of lead complications in patients with de novo CIEDs. Manipulation of the heart including cannulation of the right atrium is a likely explanation for the subsequent lead problems.\n","PeriodicalId":11720,"journal":{"name":"EP Europace","volume":"35 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2022-05-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Lead complications after cardiac surgery in danish patients with cardiac implantable electronic devices: a nationwide nested case-control study\",\"authors\":\"M. Frausing, J. Nielsen, J. Johansen, OD Jorgensen, J. Kristensen, T. Olsen, C. Gerdes, M. Kronborg\",\"doi\":\"10.1093/europace/euac053.528\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"\\n \\n \\n Type of funding sources: Private grant(s) and/or Sponsorship. Main funding source(s): Karen Elise Jensen’s Foundation\\n \\n \\n \\n Manipulation of the heart during cardiac surgery in patients with cardiac implantable electronic devices (CIEDs) may result in lead damage or -displacement, but whether cardiac surgery truly infers an excess risk of lead failure is not known.\\n \\n \\n \\n The objective of this study was to examine risk of lead complications as a result of cardiac surgery in CIED patients.\\n \\n \\n \\n We conducted a nationwide nested case-control study. Our source population comprised all patients ≥18 of age who underwent a de novo CIED implantation in Denmark between 1998 and 2017. Only patients who were alive and event-free at six months after implantation were eligible for inclusion in the case-control study. Cases were defined as incident lead complications resulting in a revision procedure, and were matched 1:30 to controls using risk set sampling. We used conditional logistic regression to estimate incidence rate ratios (IRRs) for the association between cardiac surgery and lead complications.\\n \\n \\n \\n Our population consisted of 61,869 de novo CIED patient. We identified 1324 incident cases of lead complications and 39,708 time-matched controls. 1036 cases and controls underwent cardiac surgery during follow-up; 61 cases (4.5%) and 975 controls (2.5%). Risk of lead-related reoperation was highest within the first six months of cardiac surgery (IRR 10.1, 95% CI 6.5-15.6, adjusted IRR 11.0, 95% CI 7-17.4). At one year, we observed no association between cardiac surgery and lead complications (IRR: 1.0, 95% CI 0.7-1.5 and adjusted IRR: 1.1, 95% CI 0.7-1.6).\\n \\n \\n \\n Cardiac surgery was associated with a substantial risk of lead complications in patients with de novo CIEDs. Manipulation of the heart including cannulation of the right atrium is a likely explanation for the subsequent lead problems.\\n\",\"PeriodicalId\":11720,\"journal\":{\"name\":\"EP Europace\",\"volume\":\"35 1\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2022-05-18\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"EP Europace\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1093/europace/euac053.528\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"EP Europace","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1093/europace/euac053.528","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

资金来源类型:私人资助和/或赞助。主要资金来源:Karen Elise Jensen基金会在心脏植入电子装置(cied)患者的心脏手术中对心脏的操作可能导致铅损伤或位移,但心脏手术是否真的推断出铅衰竭的过度风险尚不清楚。本研究的目的是检查CIED患者因心脏手术导致的铅并发症的风险。我们在全国范围内进行了巢式病例对照研究。我们的源人群包括1998年至2017年间在丹麦接受从头植入CIED的所有年龄≥18岁的患者。只有在植入后6个月存活且无事件的患者才有资格纳入病例对照研究。病例被定义为导致修订程序的意外铅并发症,并使用风险集抽样与对照进行1:30匹配。我们使用条件逻辑回归来估计心脏手术与导联并发症之间关联的发生率比(IRRs)。我们的人群包括61,869名新发CIED患者。我们确定了1324例铅并发症和39,708例时间匹配的对照。随访期间,1036例患者和对照组接受了心脏手术;61例(4.5%),对照组975例(2.5%)。铅相关再手术的风险在心脏手术后的前6个月内最高(IRR 10.1, 95% CI 6.5-15.6,校正IRR 11.0, 95% CI 7-17.4)。一年后,我们观察到心脏手术与导联并发症无关联(IRR: 1.0, 95% CI 0.7-1.5,校正IRR: 1.1, 95% CI 0.7-1.6)。心脏手术与新发cied患者发生主要并发症的重大风险相关。包括右心房插管在内的心脏操作可能是导致后续铅问题的一个原因。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Lead complications after cardiac surgery in danish patients with cardiac implantable electronic devices: a nationwide nested case-control study
Type of funding sources: Private grant(s) and/or Sponsorship. Main funding source(s): Karen Elise Jensen’s Foundation Manipulation of the heart during cardiac surgery in patients with cardiac implantable electronic devices (CIEDs) may result in lead damage or -displacement, but whether cardiac surgery truly infers an excess risk of lead failure is not known. The objective of this study was to examine risk of lead complications as a result of cardiac surgery in CIED patients. We conducted a nationwide nested case-control study. Our source population comprised all patients ≥18 of age who underwent a de novo CIED implantation in Denmark between 1998 and 2017. Only patients who were alive and event-free at six months after implantation were eligible for inclusion in the case-control study. Cases were defined as incident lead complications resulting in a revision procedure, and were matched 1:30 to controls using risk set sampling. We used conditional logistic regression to estimate incidence rate ratios (IRRs) for the association between cardiac surgery and lead complications. Our population consisted of 61,869 de novo CIED patient. We identified 1324 incident cases of lead complications and 39,708 time-matched controls. 1036 cases and controls underwent cardiac surgery during follow-up; 61 cases (4.5%) and 975 controls (2.5%). Risk of lead-related reoperation was highest within the first six months of cardiac surgery (IRR 10.1, 95% CI 6.5-15.6, adjusted IRR 11.0, 95% CI 7-17.4). At one year, we observed no association between cardiac surgery and lead complications (IRR: 1.0, 95% CI 0.7-1.5 and adjusted IRR: 1.1, 95% CI 0.7-1.6). Cardiac surgery was associated with a substantial risk of lead complications in patients with de novo CIEDs. Manipulation of the heart including cannulation of the right atrium is a likely explanation for the subsequent lead problems.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信