{"title":"视角倾斜和片编码金属伪影校正方法对膝关节假体磁共振成像金属伪影降低的影响","authors":"Parisa Shahbazi, B. Yasrebi, Cyrus Afshar, S. Haghipour, R. Abedi-Firouzjah","doi":"10.18502/fbt.v9i4.10419","DOIUrl":null,"url":null,"abstract":"Purpose: At Magnetic Resonance Imaging (MRI), artifacts arising from metal implants are an obstacle to obtaining optimal images. This study aimed to evaluate the impact of View-Angle Tilting (VAT) and Slice Encoding for Metal Artifact Correction (SEMAC) techniques for the artifact reduction of patients during knee MRI with metal implants. \nMaterials and Methods: The MR images without any intervention of the knee from 20 patients with knee prostheses were used. The VAT and SEMAC metal artifact reduction techniques were applied to all the MR images. Volume and mass of the metal prosthesis were quantified using the MATLAB program and compared with the real measurements using nonparametric Wilcoxon tests in SPSS software. The qualitative analysis was performed by two blinded observers regarding the score of artifact size, distortions, image quality, and visualization of bone marrow and soft tissues adjacent to metal implants. In addition, Cohen’s kappa values were used for inter-observer agreement. \nResults: The average volume of the platinum based on the conventional, VAT, and SEMAC methods was estimated at 14.22 ± 0.43, 14.05 ± 0.4, and 13.3 ± 0.45 cm3, respectively. The statistical analysis showed no significant difference (P > 0.05) between the mean value of the platinum volume for the SEMAC method and the real measurement (13.6 ± 0.33 cm3). Furthermore, regarding the conventional, VAT, and SEMAC sequences, the mean mass of the platinum was obtained at 305.02 ± 9.22, 301.37 ± 8.58, and 285.28 ± 9.65 g, respectively, with the P-Value of 0.005, 0.009, and 0.268, compared to the real measurements (286.81±8.75 g). Notably, the blinded readers demonstrated that the SEMAC method was remarkably superior quality compared with VAT and conventional acquisitions (P-Value< 0.05). \nConclusion: The knee prosthesis metal artifact was reduced using the VAT and SEMAC techniques, in a way that, the reduction was significant by the SEMAC method. In addition, concerning the qualitative observer analysis, the application of the SEMAC technique provides improved visualization of tissue structures adjacent to metal implants.","PeriodicalId":34203,"journal":{"name":"Frontiers in Biomedical Technologies","volume":"16 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2022-08-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"The Impact of View- Angle Tilting and Slice Encoding for Metal Artifact Correction Methods on the Metal Artifact Reduction in Magnetic Resonance Imaging of the Knee Prosthesis\",\"authors\":\"Parisa Shahbazi, B. Yasrebi, Cyrus Afshar, S. Haghipour, R. Abedi-Firouzjah\",\"doi\":\"10.18502/fbt.v9i4.10419\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Purpose: At Magnetic Resonance Imaging (MRI), artifacts arising from metal implants are an obstacle to obtaining optimal images. This study aimed to evaluate the impact of View-Angle Tilting (VAT) and Slice Encoding for Metal Artifact Correction (SEMAC) techniques for the artifact reduction of patients during knee MRI with metal implants. \\nMaterials and Methods: The MR images without any intervention of the knee from 20 patients with knee prostheses were used. The VAT and SEMAC metal artifact reduction techniques were applied to all the MR images. Volume and mass of the metal prosthesis were quantified using the MATLAB program and compared with the real measurements using nonparametric Wilcoxon tests in SPSS software. The qualitative analysis was performed by two blinded observers regarding the score of artifact size, distortions, image quality, and visualization of bone marrow and soft tissues adjacent to metal implants. In addition, Cohen’s kappa values were used for inter-observer agreement. \\nResults: The average volume of the platinum based on the conventional, VAT, and SEMAC methods was estimated at 14.22 ± 0.43, 14.05 ± 0.4, and 13.3 ± 0.45 cm3, respectively. The statistical analysis showed no significant difference (P > 0.05) between the mean value of the platinum volume for the SEMAC method and the real measurement (13.6 ± 0.33 cm3). Furthermore, regarding the conventional, VAT, and SEMAC sequences, the mean mass of the platinum was obtained at 305.02 ± 9.22, 301.37 ± 8.58, and 285.28 ± 9.65 g, respectively, with the P-Value of 0.005, 0.009, and 0.268, compared to the real measurements (286.81±8.75 g). Notably, the blinded readers demonstrated that the SEMAC method was remarkably superior quality compared with VAT and conventional acquisitions (P-Value< 0.05). \\nConclusion: The knee prosthesis metal artifact was reduced using the VAT and SEMAC techniques, in a way that, the reduction was significant by the SEMAC method. In addition, concerning the qualitative observer analysis, the application of the SEMAC technique provides improved visualization of tissue structures adjacent to metal implants.\",\"PeriodicalId\":34203,\"journal\":{\"name\":\"Frontiers in Biomedical Technologies\",\"volume\":\"16 1\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2022-08-30\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Frontiers in Biomedical Technologies\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.18502/fbt.v9i4.10419\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"Health Professions\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Frontiers in Biomedical Technologies","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.18502/fbt.v9i4.10419","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Health Professions","Score":null,"Total":0}
引用次数: 0

摘要

目的:在磁共振成像(MRI)中,金属植入物产生的伪影是获得最佳图像的障碍。本研究旨在评估视角倾斜(VAT)和片编码金属伪影校正(SEMAC)技术对金属植入膝关节MRI患者伪影减少的影响。材料与方法:对20例膝关节假体患者进行无干预的MR图像分析。所有MR图像均采用VAT和SEMAC金属伪影还原技术。使用MATLAB程序对金属假体的体积和质量进行量化,并在SPSS软件中使用非参数Wilcoxon检验与实际测量结果进行比较。定性分析由两名盲法观察人员进行,内容涉及伪影大小、畸变、图像质量、骨髓和金属植入物附近软组织的可视化评分。此外,Cohen的kappa值被用于观察者间的一致性。结果:常规、VAT和SEMAC方法的平均铂体积分别为14.22±0.43、14.05±0.4和13.3±0.45 cm3。统计分析显示,SEMAC法测得的铂体积平均值(13.6±0.33 cm3)与实际测量值(13.6±0.33 cm3)差异无统计学意义(P > 0.05)。此外,对于常规、VAT和SEMAC序列,获得的铂的平均质量分别为305.02±9.22、301.37±8.58和285.28±9.65 g,与实际测量值(286.81±8.75 g)相比,p值分别为0.005、0.009和0.268。值得注意的是,盲法读者表明,SEMAC方法的质量明显优于VAT和传统获取方法(p值< 0.05)。结论:采用VAT技术和SEMAC技术对膝关节假体金属假体进行复位,SEMAC方法复位效果显著。此外,在定性观察者分析方面,SEMAC技术的应用改善了金属植入物附近组织结构的可视化。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
The Impact of View- Angle Tilting and Slice Encoding for Metal Artifact Correction Methods on the Metal Artifact Reduction in Magnetic Resonance Imaging of the Knee Prosthesis
Purpose: At Magnetic Resonance Imaging (MRI), artifacts arising from metal implants are an obstacle to obtaining optimal images. This study aimed to evaluate the impact of View-Angle Tilting (VAT) and Slice Encoding for Metal Artifact Correction (SEMAC) techniques for the artifact reduction of patients during knee MRI with metal implants. Materials and Methods: The MR images without any intervention of the knee from 20 patients with knee prostheses were used. The VAT and SEMAC metal artifact reduction techniques were applied to all the MR images. Volume and mass of the metal prosthesis were quantified using the MATLAB program and compared with the real measurements using nonparametric Wilcoxon tests in SPSS software. The qualitative analysis was performed by two blinded observers regarding the score of artifact size, distortions, image quality, and visualization of bone marrow and soft tissues adjacent to metal implants. In addition, Cohen’s kappa values were used for inter-observer agreement. Results: The average volume of the platinum based on the conventional, VAT, and SEMAC methods was estimated at 14.22 ± 0.43, 14.05 ± 0.4, and 13.3 ± 0.45 cm3, respectively. The statistical analysis showed no significant difference (P > 0.05) between the mean value of the platinum volume for the SEMAC method and the real measurement (13.6 ± 0.33 cm3). Furthermore, regarding the conventional, VAT, and SEMAC sequences, the mean mass of the platinum was obtained at 305.02 ± 9.22, 301.37 ± 8.58, and 285.28 ± 9.65 g, respectively, with the P-Value of 0.005, 0.009, and 0.268, compared to the real measurements (286.81±8.75 g). Notably, the blinded readers demonstrated that the SEMAC method was remarkably superior quality compared with VAT and conventional acquisitions (P-Value< 0.05). Conclusion: The knee prosthesis metal artifact was reduced using the VAT and SEMAC techniques, in a way that, the reduction was significant by the SEMAC method. In addition, concerning the qualitative observer analysis, the application of the SEMAC technique provides improved visualization of tissue structures adjacent to metal implants.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Frontiers in Biomedical Technologies
Frontiers in Biomedical Technologies Health Professions-Medical Laboratory Technology
CiteScore
0.80
自引率
0.00%
发文量
34
审稿时长
12 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信