基于光学表面监测技术的患者定位新方法在头部放疗中的应用

Q4 Medicine
Yixiao Du, H. Yue, Meijiao Wang, Shun Zhou, Songmao Yu
{"title":"基于光学表面监测技术的患者定位新方法在头部放疗中的应用","authors":"Yixiao Du, H. Yue, Meijiao Wang, Shun Zhou, Songmao Yu","doi":"10.3760/CMA.J.ISSN.0254-5098.2020.03.009","DOIUrl":null,"url":null,"abstract":"Objective \nTo develop a new method to set up patients using optical surface monitoring system and to compare it with the conventional method in head radiotherapy. \n \n \nMethods \nA total of 358 set-ups (130 with the conventional method and 228 with the new method ), which were from 99 head tumor patients in Beijing Cancer Hospital treated between May 2018 to April 2019, obtained by using Image Guided Radiotherapy were retrospectively analyzed. The distributions of set-up errors, the number of abnormal positions, and the set-up time were compared to evaluate the potential advantages of the new method . \n \n \nResults \nThe mean (± standard deviation) absolute values of setup errors of the new method were (0.07±0.07) , (0.08±0.06) and (0.06±0.06) cm for the vertical, lateral, and longitudinal, (0.53±0.41)°, (0.59±0.44)° and (0.59±0.46)° for the rotation, pitch and roll, respectively. In the new method , the setup accuracy was improved(t=3.24-6.10, P 0.05). \n \n \nConclusions \nThe new high-precision method to set up patients using optical surface monitoring system improves the accuracy of patients′ position, decreases the corrections applied by 6DoF couch, reduces the probability of abnormal positions, and suggests the potential benefit in head radiotherapy. \n \n \nKey words: \nSet-up errors; Image guided radiotherapy; Six degrees of freedom (6DoF) couch; Optical surface monitoring technology","PeriodicalId":36403,"journal":{"name":"中华放射医学与防护杂志","volume":"70 1","pages":"209-215"},"PeriodicalIF":0.0000,"publicationDate":"2020-03-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Application of a new method for patient positioning based on optical surface monitoring technology to head radiotherapy\",\"authors\":\"Yixiao Du, H. Yue, Meijiao Wang, Shun Zhou, Songmao Yu\",\"doi\":\"10.3760/CMA.J.ISSN.0254-5098.2020.03.009\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Objective \\nTo develop a new method to set up patients using optical surface monitoring system and to compare it with the conventional method in head radiotherapy. \\n \\n \\nMethods \\nA total of 358 set-ups (130 with the conventional method and 228 with the new method ), which were from 99 head tumor patients in Beijing Cancer Hospital treated between May 2018 to April 2019, obtained by using Image Guided Radiotherapy were retrospectively analyzed. The distributions of set-up errors, the number of abnormal positions, and the set-up time were compared to evaluate the potential advantages of the new method . \\n \\n \\nResults \\nThe mean (± standard deviation) absolute values of setup errors of the new method were (0.07±0.07) , (0.08±0.06) and (0.06±0.06) cm for the vertical, lateral, and longitudinal, (0.53±0.41)°, (0.59±0.44)° and (0.59±0.46)° for the rotation, pitch and roll, respectively. In the new method , the setup accuracy was improved(t=3.24-6.10, P 0.05). \\n \\n \\nConclusions \\nThe new high-precision method to set up patients using optical surface monitoring system improves the accuracy of patients′ position, decreases the corrections applied by 6DoF couch, reduces the probability of abnormal positions, and suggests the potential benefit in head radiotherapy. \\n \\n \\nKey words: \\nSet-up errors; Image guided radiotherapy; Six degrees of freedom (6DoF) couch; Optical surface monitoring technology\",\"PeriodicalId\":36403,\"journal\":{\"name\":\"中华放射医学与防护杂志\",\"volume\":\"70 1\",\"pages\":\"209-215\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2020-03-25\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"中华放射医学与防护杂志\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.3760/CMA.J.ISSN.0254-5098.2020.03.009\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"中华放射医学与防护杂志","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3760/CMA.J.ISSN.0254-5098.2020.03.009","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

摘要

目的建立一种利用光学表面监测系统设置患者的新方法,并将其与传统的头部放疗方法进行比较。方法回顾性分析2018年5月至2019年4月北京肿瘤医院影像引导放疗治疗的99例头部肿瘤患者的358例(常规方法130例,新方法228例)。比较了设置误差分布、异常位置数和设置时间,评价了新方法的潜在优势。结果新方法的设置误差均值(±标准差)绝对值分别为(0.07±0.07)、(0.08±0.06)、(0.06±0.06)cm、(0.53±0.41)°、(0.59±0.44)°、(0.59±0.46)°,分别为纵向、横向和纵向设置误差均值(±标准差)为(0.07±0.07)、(0.08±0.06)cm。新方法提高了设置精度(t=3.24 ~ 6.10, P 0.05)。结论利用光学体表监测系统对患者进行高精度立位,提高了患者的体位精度,减少了6DoF躺椅的校正,降低了体位异常的概率,在头部放疗中具有潜在的应用价值。关键词:设置误差;影像引导放疗;六自由度(6DoF)沙发;光学表面监测技术
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Application of a new method for patient positioning based on optical surface monitoring technology to head radiotherapy
Objective To develop a new method to set up patients using optical surface monitoring system and to compare it with the conventional method in head radiotherapy. Methods A total of 358 set-ups (130 with the conventional method and 228 with the new method ), which were from 99 head tumor patients in Beijing Cancer Hospital treated between May 2018 to April 2019, obtained by using Image Guided Radiotherapy were retrospectively analyzed. The distributions of set-up errors, the number of abnormal positions, and the set-up time were compared to evaluate the potential advantages of the new method . Results The mean (± standard deviation) absolute values of setup errors of the new method were (0.07±0.07) , (0.08±0.06) and (0.06±0.06) cm for the vertical, lateral, and longitudinal, (0.53±0.41)°, (0.59±0.44)° and (0.59±0.46)° for the rotation, pitch and roll, respectively. In the new method , the setup accuracy was improved(t=3.24-6.10, P 0.05). Conclusions The new high-precision method to set up patients using optical surface monitoring system improves the accuracy of patients′ position, decreases the corrections applied by 6DoF couch, reduces the probability of abnormal positions, and suggests the potential benefit in head radiotherapy. Key words: Set-up errors; Image guided radiotherapy; Six degrees of freedom (6DoF) couch; Optical surface monitoring technology
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
中华放射医学与防护杂志
中华放射医学与防护杂志 Medicine-Radiology, Nuclear Medicine and Imaging
CiteScore
0.60
自引率
0.00%
发文量
6377
期刊介绍:
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信