{"title":"努瓦克肖特肿瘤中心直线加速器质量控制的初步结果","authors":"Ould El-Mouna, O. Moussa, Choukri Abdelmajid, Ghassoun Jilali, Hakam Keltoum, Semghouli Slimane","doi":"10.9734/bjmmr/2017/33918","DOIUrl":null,"url":null,"abstract":"The National Center of Oncology in Nouakchott is equipped with a linear accelerator which provides two energies in photon regime 6 MV and 18 MV. The basic dosimetric data required for quality of operation, in the clinical routine of the accelerator for photon beams are percentage depth dose (PDD) curves and beam profile (BP) curves. The aim of this work is to measure the percentage depth dose (PDD) and beam/dose profile (BP) for 6 MV and 18 MV photon beams for different field sizes and for different Skin-Source-Distance (SSD), and to compare the measured results with the results calculated by Treatment Planning Systems (TPS). The measured results of (PDD) and (BF) are consistent with those calculated by TPS, which allows us to conclude that the linear accelerator at oncology Center in Nouakchott is in good condition. Other Parameters: (energy, (SSD) and field sizes) have been also studied in this work, in order to perform a comprehensive quality study for this accelerator a unique oncology at Mauritania.","PeriodicalId":9249,"journal":{"name":"British journal of medicine and medical research","volume":"4 1","pages":"1-8"},"PeriodicalIF":0.0000,"publicationDate":"2017-01-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":"{\"title\":\"Preliminary Results of Quality Control for Linear Accelerator at Oncology Center in Nouakchott\",\"authors\":\"Ould El-Mouna, O. Moussa, Choukri Abdelmajid, Ghassoun Jilali, Hakam Keltoum, Semghouli Slimane\",\"doi\":\"10.9734/bjmmr/2017/33918\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"The National Center of Oncology in Nouakchott is equipped with a linear accelerator which provides two energies in photon regime 6 MV and 18 MV. The basic dosimetric data required for quality of operation, in the clinical routine of the accelerator for photon beams are percentage depth dose (PDD) curves and beam profile (BP) curves. The aim of this work is to measure the percentage depth dose (PDD) and beam/dose profile (BP) for 6 MV and 18 MV photon beams for different field sizes and for different Skin-Source-Distance (SSD), and to compare the measured results with the results calculated by Treatment Planning Systems (TPS). The measured results of (PDD) and (BF) are consistent with those calculated by TPS, which allows us to conclude that the linear accelerator at oncology Center in Nouakchott is in good condition. Other Parameters: (energy, (SSD) and field sizes) have been also studied in this work, in order to perform a comprehensive quality study for this accelerator a unique oncology at Mauritania.\",\"PeriodicalId\":9249,\"journal\":{\"name\":\"British journal of medicine and medical research\",\"volume\":\"4 1\",\"pages\":\"1-8\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2017-01-10\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"British journal of medicine and medical research\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.9734/bjmmr/2017/33918\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"British journal of medicine and medical research","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.9734/bjmmr/2017/33918","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

摘要

位于努瓦克肖特的国家肿瘤中心配备了一个线性加速器,提供6毫伏和18毫伏光子态的两种能量。在光子束流加速器的临床常规操作中,需要的基本剂量学数据是百分比深度剂量(PDD)曲线和束流剖面(BP)曲线。本文的目的是测量6 MV和18 MV光子束在不同场大小和不同皮肤-源距离(SSD)下的百分比深度剂量(PDD)和光束/剂量分布(BP),并将测量结果与治疗计划系统(TPS)计算结果进行比较。(PDD)和(BF)的测量结果与TPS计算的结果一致,这使我们得出结论,努瓦克肖特肿瘤中心的直线加速器处于良好的状态。为了对毛里塔尼亚独特的肿瘤加速器进行全面的质量研究,本工作还研究了其他参数(能量、SSD和场地大小)。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Preliminary Results of Quality Control for Linear Accelerator at Oncology Center in Nouakchott
The National Center of Oncology in Nouakchott is equipped with a linear accelerator which provides two energies in photon regime 6 MV and 18 MV. The basic dosimetric data required for quality of operation, in the clinical routine of the accelerator for photon beams are percentage depth dose (PDD) curves and beam profile (BP) curves. The aim of this work is to measure the percentage depth dose (PDD) and beam/dose profile (BP) for 6 MV and 18 MV photon beams for different field sizes and for different Skin-Source-Distance (SSD), and to compare the measured results with the results calculated by Treatment Planning Systems (TPS). The measured results of (PDD) and (BF) are consistent with those calculated by TPS, which allows us to conclude that the linear accelerator at oncology Center in Nouakchott is in good condition. Other Parameters: (energy, (SSD) and field sizes) have been also studied in this work, in order to perform a comprehensive quality study for this accelerator a unique oncology at Mauritania.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信