{"title":"遮挡效应对助听器预测误差反馈对消系统的影响。","authors":"Renata Coelho Borges, M. Holsbach Costa","doi":"10.1109/EMBC.2015.7318955","DOIUrl":null,"url":null,"abstract":"This work presents a theoretical analysis of the prediction-error method-based adaptive feedback canceller in hearing aid applications. The studied scene takes into account the occlusion effect caused by the partial or complete closing of the ventilation opening. Such a situation may occur in high gain applications to avoid undesired whistling. Deterministic recursive equations and steady-state conditions were derived for the mean weight behaviour of the predictor and the adaptive filter. The expected theoretical predictions were compared to Monte Carlo simulations, showing very accurate agreement. The simulation results suggest the steady-state performance of this feedback canceller is not affected by the occlusion effect, however the occlusion is still perceived, being annoying to the user.","PeriodicalId":72689,"journal":{"name":"Conference proceedings : ... Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual Conference","volume":"68 1","pages":"2725-8"},"PeriodicalIF":0.0000,"publicationDate":"2015-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Influence of the occlusion effect over the prediction-error feedback cancellation system in hearing aids.\",\"authors\":\"Renata Coelho Borges, M. Holsbach Costa\",\"doi\":\"10.1109/EMBC.2015.7318955\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"This work presents a theoretical analysis of the prediction-error method-based adaptive feedback canceller in hearing aid applications. The studied scene takes into account the occlusion effect caused by the partial or complete closing of the ventilation opening. Such a situation may occur in high gain applications to avoid undesired whistling. Deterministic recursive equations and steady-state conditions were derived for the mean weight behaviour of the predictor and the adaptive filter. The expected theoretical predictions were compared to Monte Carlo simulations, showing very accurate agreement. The simulation results suggest the steady-state performance of this feedback canceller is not affected by the occlusion effect, however the occlusion is still perceived, being annoying to the user.\",\"PeriodicalId\":72689,\"journal\":{\"name\":\"Conference proceedings : ... Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual Conference\",\"volume\":\"68 1\",\"pages\":\"2725-8\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2015-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Conference proceedings : ... Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual Conference\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/EMBC.2015.7318955\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Conference proceedings : ... Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual Conference","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/EMBC.2015.7318955","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Influence of the occlusion effect over the prediction-error feedback cancellation system in hearing aids.
This work presents a theoretical analysis of the prediction-error method-based adaptive feedback canceller in hearing aid applications. The studied scene takes into account the occlusion effect caused by the partial or complete closing of the ventilation opening. Such a situation may occur in high gain applications to avoid undesired whistling. Deterministic recursive equations and steady-state conditions were derived for the mean weight behaviour of the predictor and the adaptive filter. The expected theoretical predictions were compared to Monte Carlo simulations, showing very accurate agreement. The simulation results suggest the steady-state performance of this feedback canceller is not affected by the occlusion effect, however the occlusion is still perceived, being annoying to the user.