{"title":"非胶结水杯的后期脱落——恢复分析及失效机理的假设","authors":"Chao-Tse Chiu, Ta-I Wang, Jen-Ting Huang, Kui-Chou Huang","doi":"10.6492/FJMD.201805_9(2).0008","DOIUrl":null,"url":null,"abstract":"Uncemented modular acetabulum components demonstrate a lower rate of radiographic loosening. However, most of late failure was caused by the osteolysis, inducing bone destruction and loss of fixation. Loosening of the well-fixed cup caused by late shedding of the ingrowth surface of uncemented cup is a rare complication. We report of three cases, and through retrieval analysis of the cup we postulate the possible mechanism of late shedding of well-fixed cup. We found that improper position of the cup and eccentric polyethylene (PE) wear may cause impingement of the cup by neck and increase tensile and shear forces which may cause late shedding of good bone-ingrowth cup. Early revision of the eccentric wear PE is indicated for prevention of this complication.","PeriodicalId":100551,"journal":{"name":"Formosan Journal of Musculoskeletal Disorders","volume":"10 1","pages":"82-88"},"PeriodicalIF":0.0000,"publicationDate":"2018-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Late shedding of uncemented cup-Retrieval analysis and postulate the mechanism of failure\",\"authors\":\"Chao-Tse Chiu, Ta-I Wang, Jen-Ting Huang, Kui-Chou Huang\",\"doi\":\"10.6492/FJMD.201805_9(2).0008\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Uncemented modular acetabulum components demonstrate a lower rate of radiographic loosening. However, most of late failure was caused by the osteolysis, inducing bone destruction and loss of fixation. Loosening of the well-fixed cup caused by late shedding of the ingrowth surface of uncemented cup is a rare complication. We report of three cases, and through retrieval analysis of the cup we postulate the possible mechanism of late shedding of well-fixed cup. We found that improper position of the cup and eccentric polyethylene (PE) wear may cause impingement of the cup by neck and increase tensile and shear forces which may cause late shedding of good bone-ingrowth cup. Early revision of the eccentric wear PE is indicated for prevention of this complication.\",\"PeriodicalId\":100551,\"journal\":{\"name\":\"Formosan Journal of Musculoskeletal Disorders\",\"volume\":\"10 1\",\"pages\":\"82-88\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2018-05-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Formosan Journal of Musculoskeletal Disorders\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.6492/FJMD.201805_9(2).0008\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Formosan Journal of Musculoskeletal Disorders","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.6492/FJMD.201805_9(2).0008","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

未骨水泥的模块化髋臼假体显示出较低的x线片松动率。然而,大多数晚期失败是由骨溶解引起的,导致骨破坏和固定丧失。由于未胶结的杯体生长面脱落而导致固定良好的杯体松动是一种罕见的并发症。我们报告了三个病例,并通过对杯子的检索分析,我们假设了固定良好的杯子晚脱落的可能机制。我们发现罩杯位置不当和聚乙烯(PE)偏心磨损可能会导致颈部撞击罩杯,增加拉伸和剪切力,从而导致良好的骨长入罩杯延迟脱落。建议早期对偏心磨损PE进行翻修,以预防这种并发症。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Late shedding of uncemented cup-Retrieval analysis and postulate the mechanism of failure
Uncemented modular acetabulum components demonstrate a lower rate of radiographic loosening. However, most of late failure was caused by the osteolysis, inducing bone destruction and loss of fixation. Loosening of the well-fixed cup caused by late shedding of the ingrowth surface of uncemented cup is a rare complication. We report of three cases, and through retrieval analysis of the cup we postulate the possible mechanism of late shedding of well-fixed cup. We found that improper position of the cup and eccentric polyethylene (PE) wear may cause impingement of the cup by neck and increase tensile and shear forces which may cause late shedding of good bone-ingrowth cup. Early revision of the eccentric wear PE is indicated for prevention of this complication.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信