新型冠状病毒肺炎防控中CT模块的剂量监测及放射防护分析

Q4 Medicine
Ansheng Liu, Bolin Huang, Cuiling Li, Kai Yang, H. Du, Guilin Yi
{"title":"新型冠状病毒肺炎防控中CT模块的剂量监测及放射防护分析","authors":"Ansheng Liu, Bolin Huang, Cuiling Li, Kai Yang, H. Du, Guilin Yi","doi":"10.3760/CMA.J.ISSN.0254-5098.0008","DOIUrl":null,"url":null,"abstract":"Objective \nTo investigate the safety and protection level of radiological treatment in the CT module in Wuhan during the prevention and treatment of COVID-19. \n \n \nMethods \nThe layout position of the CT module in makeshift hospitals, radiological protection facilities and personal protective equipment were investigated. Based on the national standards, CT dose index and evaluated radiological protection level of the CT module were measured. \n \n \nResults \nThe layout position of the CT module in makeshift hospitals are reasonable, and radiological protection facilities are well-equipped. 87.0% of them had been accepted. In 23 CT module, 20 of which were up to standards. The other three had 1 detection points which exceed the national standards, and then had been modified immediately, so as to reach the national standards. In addition, CT dose index was measured in 7 CT module of them, the variation of body CTDIW was less than 7.5%. \n \n \nConclusions \nThe CT module in Wuhan meet the requirements of the radiological safety and protection level during the prevention and treatment of COVID-19. \n \n \nKey words: \nCOVID-19; CT module; Radiological protection; CT dose index (CTDI)","PeriodicalId":36403,"journal":{"name":"中华放射医学与防护杂志","volume":"11 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2020-03-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Monitoring of CT dose and analysis of radiological protection in CT module applied for COVID-19 prevention and control\",\"authors\":\"Ansheng Liu, Bolin Huang, Cuiling Li, Kai Yang, H. Du, Guilin Yi\",\"doi\":\"10.3760/CMA.J.ISSN.0254-5098.0008\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Objective \\nTo investigate the safety and protection level of radiological treatment in the CT module in Wuhan during the prevention and treatment of COVID-19. \\n \\n \\nMethods \\nThe layout position of the CT module in makeshift hospitals, radiological protection facilities and personal protective equipment were investigated. Based on the national standards, CT dose index and evaluated radiological protection level of the CT module were measured. \\n \\n \\nResults \\nThe layout position of the CT module in makeshift hospitals are reasonable, and radiological protection facilities are well-equipped. 87.0% of them had been accepted. In 23 CT module, 20 of which were up to standards. The other three had 1 detection points which exceed the national standards, and then had been modified immediately, so as to reach the national standards. In addition, CT dose index was measured in 7 CT module of them, the variation of body CTDIW was less than 7.5%. \\n \\n \\nConclusions \\nThe CT module in Wuhan meet the requirements of the radiological safety and protection level during the prevention and treatment of COVID-19. \\n \\n \\nKey words: \\nCOVID-19; CT module; Radiological protection; CT dose index (CTDI)\",\"PeriodicalId\":36403,\"journal\":{\"name\":\"中华放射医学与防护杂志\",\"volume\":\"11 1\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2020-03-16\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"中华放射医学与防护杂志\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.3760/CMA.J.ISSN.0254-5098.0008\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"中华放射医学与防护杂志","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3760/CMA.J.ISSN.0254-5098.0008","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

摘要

目的了解武汉市新型冠状病毒肺炎防治过程中CT模块放射治疗的安全性和防护水平。方法对方舱医院、放射防护设施、个人防护用品的CT模块布置位置进行调查。根据国家标准,测量了CT模块的剂量指数和评估放射防护水平。结果方舱医院CT模块布置位置合理,放射防护设施齐全。录取率为87.0%。在23个CT模块中,20个模块达到标准。另外3家有1个检测点超过国家标准,随即进行了整改,达到国家标准。另外,对其中7个CT模块的CT剂量指数进行了测量,机体CTDIW的变化均小于7.5%。结论武汉市CT模块满足新型冠状病毒肺炎防治期间的放射安全防护水平要求。关键词:COVID-19;CT模块;放射防护;CT剂量指数(CTDI)
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Monitoring of CT dose and analysis of radiological protection in CT module applied for COVID-19 prevention and control
Objective To investigate the safety and protection level of radiological treatment in the CT module in Wuhan during the prevention and treatment of COVID-19. Methods The layout position of the CT module in makeshift hospitals, radiological protection facilities and personal protective equipment were investigated. Based on the national standards, CT dose index and evaluated radiological protection level of the CT module were measured. Results The layout position of the CT module in makeshift hospitals are reasonable, and radiological protection facilities are well-equipped. 87.0% of them had been accepted. In 23 CT module, 20 of which were up to standards. The other three had 1 detection points which exceed the national standards, and then had been modified immediately, so as to reach the national standards. In addition, CT dose index was measured in 7 CT module of them, the variation of body CTDIW was less than 7.5%. Conclusions The CT module in Wuhan meet the requirements of the radiological safety and protection level during the prevention and treatment of COVID-19. Key words: COVID-19; CT module; Radiological protection; CT dose index (CTDI)
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
中华放射医学与防护杂志
中华放射医学与防护杂志 Medicine-Radiology, Nuclear Medicine and Imaging
CiteScore
0.60
自引率
0.00%
发文量
6377
期刊介绍:
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信