{"title":"水密度的比较研究及其在体积校准不确定预算中的应用","authors":"A. Balsa, Sheila Preste, G. Almeida, R. Esteves","doi":"10.26461/21.05","DOIUrl":null,"url":null,"abstract":"In the quantity of volume, gravimetric method is considered the best accuracy method for the quantitative determination of an analyte based on its mass. The distilled water is selected as calibration liquid because it is well known how its density changes with temperature. For volume calculation at reference temperature, the water density value is obtained by known mathematic models, for example, Tanaka equation (Tanaka, et al., 2001). According to ISO 3696 (International Organization for Standardization, 1987), there are requirements for water (Grade 3), the lab needs to perform periodic studies of conductivity, but it is necessary to study the water density deviation and comparing this value with that obtained by Tanaka equation. If this deviation value (calculated by Tanaka Equation vs Oscillation – type density meter is significant, it must be considered in the uncertainty estimation for volume calibration using gravimetric method.","PeriodicalId":30552,"journal":{"name":"Innotec","volume":"1 1","pages":"25-35"},"PeriodicalIF":0.0000,"publicationDate":"2020-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Estudio comparativo de la densidad del agua y su aplicación en el presupuesto de incertidumbre para calibraciones en volumen\",\"authors\":\"A. Balsa, Sheila Preste, G. Almeida, R. Esteves\",\"doi\":\"10.26461/21.05\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"In the quantity of volume, gravimetric method is considered the best accuracy method for the quantitative determination of an analyte based on its mass. The distilled water is selected as calibration liquid because it is well known how its density changes with temperature. For volume calculation at reference temperature, the water density value is obtained by known mathematic models, for example, Tanaka equation (Tanaka, et al., 2001). According to ISO 3696 (International Organization for Standardization, 1987), there are requirements for water (Grade 3), the lab needs to perform periodic studies of conductivity, but it is necessary to study the water density deviation and comparing this value with that obtained by Tanaka equation. If this deviation value (calculated by Tanaka Equation vs Oscillation – type density meter is significant, it must be considered in the uncertainty estimation for volume calibration using gravimetric method.\",\"PeriodicalId\":30552,\"journal\":{\"name\":\"Innotec\",\"volume\":\"1 1\",\"pages\":\"25-35\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2020-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Innotec\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.26461/21.05\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Innotec","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.26461/21.05","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

在体积的数量上,重量法被认为是根据分析物的质量进行定量测定的最准确的方法。选择蒸馏水作为校准液,因为众所周知,蒸馏水的密度随温度的变化是怎样的。对于参考温度下的体积计算,水密度值由已知的数学模型获得,例如Tanaka方程(Tanaka, et al., 2001)。根据ISO 3696(国际标准化组织,1987),对水(3级)有要求,实验室需要进行电导率的定期研究,但有必要研究水密度偏差,并将该值与田中方程得到的值进行比较。如果该偏差值(由田中方程与振荡式密度计计算)显著,则在用重量法进行体积校准的不确定度估计时必须考虑该偏差值。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Estudio comparativo de la densidad del agua y su aplicación en el presupuesto de incertidumbre para calibraciones en volumen
In the quantity of volume, gravimetric method is considered the best accuracy method for the quantitative determination of an analyte based on its mass. The distilled water is selected as calibration liquid because it is well known how its density changes with temperature. For volume calculation at reference temperature, the water density value is obtained by known mathematic models, for example, Tanaka equation (Tanaka, et al., 2001). According to ISO 3696 (International Organization for Standardization, 1987), there are requirements for water (Grade 3), the lab needs to perform periodic studies of conductivity, but it is necessary to study the water density deviation and comparing this value with that obtained by Tanaka equation. If this deviation value (calculated by Tanaka Equation vs Oscillation – type density meter is significant, it must be considered in the uncertainty estimation for volume calibration using gravimetric method.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
10
审稿时长
16 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信