H. Defino, H. R. D. Costa, Leonardo Ribeiro Nascimento, Izabella Meirelles Guarato
{"title":"无椎体系统在脊柱疾病的外科治疗中的应用","authors":"H. Defino, H. R. D. Costa, Leonardo Ribeiro Nascimento, Izabella Meirelles Guarato","doi":"10.1590/s1808-185120222201262504","DOIUrl":null,"url":null,"abstract":"ABSTRACT Objectives: Evaluate the treatment outcome and the performance of the uCentum spinal fixation system in treating traumatic, degenerative, and tumoral diseases of the spine. Methods: This is a therapeutic study to investigate treatment outcomes and level of evidence III, including twenty-three adult patients of both sexes undergoing surgical treatment of degenerative (13 patients), traumatic (04 patients), or tumor diseases (06 patients). Patients were prospectively evaluated using clinical parameters: pain (visual analog scale), clinical and functional assessment questionnaires (SF-36, Oswestry and Roland-Morris), and radiological criteria (arthrodesis consolidation, loosening, breakage or deformation of the implants). Results: Twenty patients were followed for a period of 01 month to 12 month (mean 6,5±7,77). Three patients died due to complications unrelated to the primary disease (traumatic brain injury, septicemia, and lung tumor). Improvements were observed in clinical parameters and scores of the evaluation questionnaires used. No implant-related complications (breakage, loosening, deformation) were observed. Conclusion: the uCentum fixation system showed great versatility for performing the surgical treatment, allowing the performance of open, percutaneous procedures, the introduction of acrylic cement inside the implants, and conversion of polyaxial screws into monoaxial screws intraoperatively. Level of Evidence III; Therapeutic Studies - Investigating the Results of Treatment.","PeriodicalId":40025,"journal":{"name":"Coluna/ Columna","volume":"1 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2023-03-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"USE OF THE uCentrum SYSTEM IN THE SURGICAL TREATMENT OF DISEASES OF THE VERTEBRAL SPINE\",\"authors\":\"H. Defino, H. R. D. Costa, Leonardo Ribeiro Nascimento, Izabella Meirelles Guarato\",\"doi\":\"10.1590/s1808-185120222201262504\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"ABSTRACT Objectives: Evaluate the treatment outcome and the performance of the uCentum spinal fixation system in treating traumatic, degenerative, and tumoral diseases of the spine. Methods: This is a therapeutic study to investigate treatment outcomes and level of evidence III, including twenty-three adult patients of both sexes undergoing surgical treatment of degenerative (13 patients), traumatic (04 patients), or tumor diseases (06 patients). Patients were prospectively evaluated using clinical parameters: pain (visual analog scale), clinical and functional assessment questionnaires (SF-36, Oswestry and Roland-Morris), and radiological criteria (arthrodesis consolidation, loosening, breakage or deformation of the implants). Results: Twenty patients were followed for a period of 01 month to 12 month (mean 6,5±7,77). Three patients died due to complications unrelated to the primary disease (traumatic brain injury, septicemia, and lung tumor). Improvements were observed in clinical parameters and scores of the evaluation questionnaires used. No implant-related complications (breakage, loosening, deformation) were observed. Conclusion: the uCentum fixation system showed great versatility for performing the surgical treatment, allowing the performance of open, percutaneous procedures, the introduction of acrylic cement inside the implants, and conversion of polyaxial screws into monoaxial screws intraoperatively. Level of Evidence III; Therapeutic Studies - Investigating the Results of Treatment.\",\"PeriodicalId\":40025,\"journal\":{\"name\":\"Coluna/ Columna\",\"volume\":\"1 1\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2023-03-13\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Coluna/ Columna\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1590/s1808-185120222201262504\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Coluna/ Columna","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1590/s1808-185120222201262504","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
USE OF THE uCentrum SYSTEM IN THE SURGICAL TREATMENT OF DISEASES OF THE VERTEBRAL SPINE
ABSTRACT Objectives: Evaluate the treatment outcome and the performance of the uCentum spinal fixation system in treating traumatic, degenerative, and tumoral diseases of the spine. Methods: This is a therapeutic study to investigate treatment outcomes and level of evidence III, including twenty-three adult patients of both sexes undergoing surgical treatment of degenerative (13 patients), traumatic (04 patients), or tumor diseases (06 patients). Patients were prospectively evaluated using clinical parameters: pain (visual analog scale), clinical and functional assessment questionnaires (SF-36, Oswestry and Roland-Morris), and radiological criteria (arthrodesis consolidation, loosening, breakage or deformation of the implants). Results: Twenty patients were followed for a period of 01 month to 12 month (mean 6,5±7,77). Three patients died due to complications unrelated to the primary disease (traumatic brain injury, septicemia, and lung tumor). Improvements were observed in clinical parameters and scores of the evaluation questionnaires used. No implant-related complications (breakage, loosening, deformation) were observed. Conclusion: the uCentum fixation system showed great versatility for performing the surgical treatment, allowing the performance of open, percutaneous procedures, the introduction of acrylic cement inside the implants, and conversion of polyaxial screws into monoaxial screws intraoperatively. Level of Evidence III; Therapeutic Studies - Investigating the Results of Treatment.