{"title":"心室外引流磁连接装置的设计与分析","authors":"Rongguo Yan,&nbsp;Junjun Ye,&nbsp;Jiahui Wang,&nbsp;Maomao Wang","doi":"10.1049/smt2.12132","DOIUrl":null,"url":null,"abstract":"<p>It is difficult to separate Luer lock adapters during the external ventricular drainage (EVD) since they are often connected by a threaded connection, and the infusion needle put into the patient's body gets pulled out when extubation occurs unexpectedly. The paper described a new magnetic connection assembly made up of ring magnetic permanents and a Hall sensor that could be used to replace Luer lock adapters for quicker disconnection and reconnection of the drainage catheter. For the design and analysis of the device, the finite element method magnetics (FEMM) was used to estimate the size of magnetic permanents and the best detection site of the Hall sensor for detecting whether the magnetic permanents were closely attracted and/or pulled apart. Then an experiment was conducted to confirm our findings. Both simulation and experiment revealed that a pair of hollow cylindrical magnetic permanents, each with height 10 mm, outer diameter 8 mm, and inner diameter 4 mm, could achieve the same force as the Luer lock adapters held, and that the best Hall sensor detection position was in the middle of the permanent magnets, with a radial distance of about 2 mm.</p>","PeriodicalId":54999,"journal":{"name":"Iet Science Measurement & Technology","volume":null,"pages":null},"PeriodicalIF":1.4000,"publicationDate":"2022-11-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1049/smt2.12132","citationCount":"1","resultStr":"{\"title\":\"Design and analysis of a magnetic connection device for external ventricular drain\",\"authors\":\"Rongguo Yan,&nbsp;Junjun Ye,&nbsp;Jiahui Wang,&nbsp;Maomao Wang\",\"doi\":\"10.1049/smt2.12132\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p>It is difficult to separate Luer lock adapters during the external ventricular drainage (EVD) since they are often connected by a threaded connection, and the infusion needle put into the patient's body gets pulled out when extubation occurs unexpectedly. The paper described a new magnetic connection assembly made up of ring magnetic permanents and a Hall sensor that could be used to replace Luer lock adapters for quicker disconnection and reconnection of the drainage catheter. For the design and analysis of the device, the finite element method magnetics (FEMM) was used to estimate the size of magnetic permanents and the best detection site of the Hall sensor for detecting whether the magnetic permanents were closely attracted and/or pulled apart. Then an experiment was conducted to confirm our findings. Both simulation and experiment revealed that a pair of hollow cylindrical magnetic permanents, each with height 10 mm, outer diameter 8 mm, and inner diameter 4 mm, could achieve the same force as the Luer lock adapters held, and that the best Hall sensor detection position was in the middle of the permanent magnets, with a radial distance of about 2 mm.</p>\",\"PeriodicalId\":54999,\"journal\":{\"name\":\"Iet Science Measurement & Technology\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.4000,\"publicationDate\":\"2022-11-22\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://onlinelibrary.wiley.com/doi/epdf/10.1049/smt2.12132\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Iet Science Measurement & Technology\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://onlinelibrary.wiley.com/doi/10.1049/smt2.12132\",\"RegionNum\":4,\"RegionCategory\":\"工程技术\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"ENGINEERING, ELECTRICAL & ELECTRONIC\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Iet Science Measurement & Technology","FirstCategoryId":"5","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1049/smt2.12132","RegionNum":4,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ENGINEERING, ELECTRICAL & ELECTRONIC","Score":null,"Total":0}
引用次数: 1

摘要

在心室外引流(EVD)过程中,很难分离Luer锁适配器,因为它们通常通过螺纹连接,并且当意外拔管时,插入患者体内的输液针会被拔出。本文介绍了一种由环形永磁体和霍尔传感器组成的新型磁性连接组件,该组件可用于取代路厄锁定适配器,以更快地断开和重新连接引流导管。为了设计和分析该装置,使用有限元法磁学(FEMM)来估计永磁体的尺寸和霍尔传感器的最佳检测位置,以检测永磁体是否被紧密吸引和/或拉开。然后进行了一项实验来证实我们的发现。模拟和实验都表明,一对高度为10mm、外径为8mm、内径为4mm的空心圆柱形永磁体可以获得与鲁尔锁适配器相同的力,霍尔传感器的最佳检测位置位于永磁体的中间,径向距离约为2mm。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

Design and analysis of a magnetic connection device for external ventricular drain

Design and analysis of a magnetic connection device for external ventricular drain

It is difficult to separate Luer lock adapters during the external ventricular drainage (EVD) since they are often connected by a threaded connection, and the infusion needle put into the patient's body gets pulled out when extubation occurs unexpectedly. The paper described a new magnetic connection assembly made up of ring magnetic permanents and a Hall sensor that could be used to replace Luer lock adapters for quicker disconnection and reconnection of the drainage catheter. For the design and analysis of the device, the finite element method magnetics (FEMM) was used to estimate the size of magnetic permanents and the best detection site of the Hall sensor for detecting whether the magnetic permanents were closely attracted and/or pulled apart. Then an experiment was conducted to confirm our findings. Both simulation and experiment revealed that a pair of hollow cylindrical magnetic permanents, each with height 10 mm, outer diameter 8 mm, and inner diameter 4 mm, could achieve the same force as the Luer lock adapters held, and that the best Hall sensor detection position was in the middle of the permanent magnets, with a radial distance of about 2 mm.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Iet Science Measurement & Technology
Iet Science Measurement & Technology 工程技术-工程:电子与电气
CiteScore
4.30
自引率
7.10%
发文量
41
审稿时长
7.5 months
期刊介绍: IET Science, Measurement & Technology publishes papers in science, engineering and technology underpinning electronic and electrical engineering, nanotechnology and medical instrumentation.The emphasis of the journal is on theory, simulation methodologies and measurement techniques. The major themes of the journal are: - electromagnetism including electromagnetic theory, computational electromagnetics and EMC - properties and applications of dielectric, magnetic, magneto-optic, piezoelectric materials down to the nanometre scale - measurement and instrumentation including sensors, actuators, medical instrumentation, fundamentals of measurement including measurement standards, uncertainty, dissemination and calibration Applications are welcome for illustrative purposes but the novelty and originality should focus on the proposed new methods.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信