Barozen Lipid Plus用于脂质状况护理点测试的性能评估:一项方法比较研究。

IF 1 Q3 MEDICINE, GENERAL & INTERNAL
Journal of Yeungnam medical science Pub Date : 2023-11-01 Epub Date: 2023-10-20 DOI:10.12701/jyms.2023.00528
Soojoung Yu, Hwa Yeon Sun, Byungwook Yoo
{"title":"Barozen Lipid Plus用于脂质状况护理点测试的性能评估:一项方法比较研究。","authors":"Soojoung Yu, Hwa Yeon Sun, Byungwook Yoo","doi":"10.12701/jyms.2023.00528","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The quick and easy nature of point-of-care (POC) testing devices allows regular monitoring of serum lipid levels to increase efficiency. The purpose of this study was to assess a POC lipid analyzer, Barozen Lipid Plus (MICO Biomed Co., Ltd.), which uses capillary blood to measure total cholesterol (TC), triglycerides (TGs), and high-density lipoprotein cholesterol (HDL-C).</p><p><strong>Methods: </strong>Capillary and venous blood samples were collected from 110 participants at a single center in Korea between June 10 and June 26, 2021. TC, TG, and HDL-C measurements using Barozen Lipid Plus were compared with measurements using our reference device, the Roche-Hitachi Cobas 8000 c702 (Hitachi High-Technologies Corporation). This study followed the guidelines of the Clinical and Laboratory Standards Institute and the Clinical Laboratory Improvement Amendments. We surveyed participants regarding the convenience of the POC device using a questionnaire following the completion of blood collection.</p><p><strong>Results: </strong>When compared to the reference equipment, the measurements obtained using Barozen Lipid Plus were more than 95% satisfactory within TC±10%, TG±25%, and HDL-C±30%. The coefficient of variation in the repeatability testing was within 5% for TC, 5% for TGs, and 7% for HDL-C. The survey results indicated high levels of satisfaction. No adverse events were reported.</p><p><strong>Conclusion: </strong>These findings suggest that Barozen Lipid Plus is reliable for measuring lipid profiles and can therefore be used to monitor lipid levels at the time and place of patient care.</p>","PeriodicalId":74020,"journal":{"name":"Journal of Yeungnam medical science","volume":null,"pages":null},"PeriodicalIF":1.0000,"publicationDate":"2023-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10718599/pdf/","citationCount":"0","resultStr":"{\"title\":\"Performance evaluation of Barozen Lipid Plus for point-of-care testing of lipid profiles: a method comparison study.\",\"authors\":\"Soojoung Yu, Hwa Yeon Sun, Byungwook Yoo\",\"doi\":\"10.12701/jyms.2023.00528\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>The quick and easy nature of point-of-care (POC) testing devices allows regular monitoring of serum lipid levels to increase efficiency. The purpose of this study was to assess a POC lipid analyzer, Barozen Lipid Plus (MICO Biomed Co., Ltd.), which uses capillary blood to measure total cholesterol (TC), triglycerides (TGs), and high-density lipoprotein cholesterol (HDL-C).</p><p><strong>Methods: </strong>Capillary and venous blood samples were collected from 110 participants at a single center in Korea between June 10 and June 26, 2021. TC, TG, and HDL-C measurements using Barozen Lipid Plus were compared with measurements using our reference device, the Roche-Hitachi Cobas 8000 c702 (Hitachi High-Technologies Corporation). This study followed the guidelines of the Clinical and Laboratory Standards Institute and the Clinical Laboratory Improvement Amendments. We surveyed participants regarding the convenience of the POC device using a questionnaire following the completion of blood collection.</p><p><strong>Results: </strong>When compared to the reference equipment, the measurements obtained using Barozen Lipid Plus were more than 95% satisfactory within TC±10%, TG±25%, and HDL-C±30%. The coefficient of variation in the repeatability testing was within 5% for TC, 5% for TGs, and 7% for HDL-C. The survey results indicated high levels of satisfaction. No adverse events were reported.</p><p><strong>Conclusion: </strong>These findings suggest that Barozen Lipid Plus is reliable for measuring lipid profiles and can therefore be used to monitor lipid levels at the time and place of patient care.</p>\",\"PeriodicalId\":74020,\"journal\":{\"name\":\"Journal of Yeungnam medical science\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.0000,\"publicationDate\":\"2023-11-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10718599/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Yeungnam medical science\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.12701/jyms.2023.00528\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2023/10/20 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q3\",\"JCRName\":\"MEDICINE, GENERAL & INTERNAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Yeungnam medical science","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.12701/jyms.2023.00528","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/10/20 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"MEDICINE, GENERAL & INTERNAL","Score":null,"Total":0}
引用次数: 0

摘要

背景:护理点(POC)检测设备快速简便,可以定期监测血脂水平,提高效率。本研究的目的是评估POC脂质分析仪Barozen lipid Plus(MICO Biomed Co.,有限公司),该分析仪使用毛细管血液测量总胆固醇(TC)、甘油三酯(TG)和高密度脂蛋白胆固醇(HDL-C)。将使用Barozen Lipid Plus的TC、TG和HDL-C测量值与使用我们的参考设备Roche Hitachi Cobas 8000 c702(日立高科技公司)的测量值进行比较。这项研究遵循了临床和实验室标准研究所和临床实验室改进修正案的指导方针。我们在完成血液采集后使用问卷调查了参与者POC设备的便利性。结果:与参考设备相比,使用Barozen Lipid Plus获得的测量结果在TC±10%、TG±25%和HDL-C±30%范围内的满意率超过95%。TC、TG和HDL-C的重复性测试的变异系数分别在5%、5%和7%以内。调查结果显示满意度很高。未报告不良事件。结论:这些发现表明,Barozen Lipid Plus在测量脂质谱方面是可靠的,因此可以用于在患者护理的时间和地点监测脂质水平。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Performance evaluation of Barozen Lipid Plus for point-of-care testing of lipid profiles: a method comparison study.

Background: The quick and easy nature of point-of-care (POC) testing devices allows regular monitoring of serum lipid levels to increase efficiency. The purpose of this study was to assess a POC lipid analyzer, Barozen Lipid Plus (MICO Biomed Co., Ltd.), which uses capillary blood to measure total cholesterol (TC), triglycerides (TGs), and high-density lipoprotein cholesterol (HDL-C).

Methods: Capillary and venous blood samples were collected from 110 participants at a single center in Korea between June 10 and June 26, 2021. TC, TG, and HDL-C measurements using Barozen Lipid Plus were compared with measurements using our reference device, the Roche-Hitachi Cobas 8000 c702 (Hitachi High-Technologies Corporation). This study followed the guidelines of the Clinical and Laboratory Standards Institute and the Clinical Laboratory Improvement Amendments. We surveyed participants regarding the convenience of the POC device using a questionnaire following the completion of blood collection.

Results: When compared to the reference equipment, the measurements obtained using Barozen Lipid Plus were more than 95% satisfactory within TC±10%, TG±25%, and HDL-C±30%. The coefficient of variation in the repeatability testing was within 5% for TC, 5% for TGs, and 7% for HDL-C. The survey results indicated high levels of satisfaction. No adverse events were reported.

Conclusion: These findings suggest that Barozen Lipid Plus is reliable for measuring lipid profiles and can therefore be used to monitor lipid levels at the time and place of patient care.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
0.80
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信