心脏停搏液系统压力是顺行心脏停搏药输送过程中冠状动脉灌注的最佳测量指标吗?最佳顺产分娩压力测量的关键回顾。

IF 1.1 4区 医学 Q4 CARDIAC & CARDIOVASCULAR SYSTEMS
Perfusion-Uk Pub Date : 2024-11-01 Epub Date: 2023-10-19 DOI:10.1177/02676591231206016
Allison Hacker, Peter Maggs, Patrick Treanor, Kevin Lilly, Vladimir Birjiniuk
{"title":"心脏停搏液系统压力是顺行心脏停搏药输送过程中冠状动脉灌注的最佳测量指标吗?最佳顺产分娩压力测量的关键回顾。","authors":"Allison Hacker, Peter Maggs, Patrick Treanor, Kevin Lilly, Vladimir Birjiniuk","doi":"10.1177/02676591231206016","DOIUrl":null,"url":null,"abstract":"<p><p>Antegrade cardioplegia is routinely given during cardiac surgery. The delivery of antegrade cardioplegia from the cardiopulmonary bypass machine has many variables. Many perfusionists rely exclusively on cardioplegia system pressure to ensure safe antegrade delivery. Our group reviewed antegrade cardioplegia delivery in 50 patients undergoing coronary artery bypass graft. The data collected included the cardioplegia system pressure and the patient's direct aortic root pressure. The analysis of the data found <i>weak correlation</i> between the two pressures with a <i>large</i> mean difference and a <i>wide</i> standard deviation. The results suggest the direct measurement of aortic root pressure as guidance to antegrade cardioplegia instead of relying solely on cardioplegia system pressure.</p>","PeriodicalId":49707,"journal":{"name":"Perfusion-Uk","volume":" ","pages":"1621-1628"},"PeriodicalIF":1.1000,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Is cardioplegia system pressure the optimal measure of coronary perfusion during antegrade cardioplegia delivery? A critical review of pressure measurements for optimal antegrade delivery.\",\"authors\":\"Allison Hacker, Peter Maggs, Patrick Treanor, Kevin Lilly, Vladimir Birjiniuk\",\"doi\":\"10.1177/02676591231206016\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Antegrade cardioplegia is routinely given during cardiac surgery. The delivery of antegrade cardioplegia from the cardiopulmonary bypass machine has many variables. Many perfusionists rely exclusively on cardioplegia system pressure to ensure safe antegrade delivery. Our group reviewed antegrade cardioplegia delivery in 50 patients undergoing coronary artery bypass graft. The data collected included the cardioplegia system pressure and the patient's direct aortic root pressure. The analysis of the data found <i>weak correlation</i> between the two pressures with a <i>large</i> mean difference and a <i>wide</i> standard deviation. The results suggest the direct measurement of aortic root pressure as guidance to antegrade cardioplegia instead of relying solely on cardioplegia system pressure.</p>\",\"PeriodicalId\":49707,\"journal\":{\"name\":\"Perfusion-Uk\",\"volume\":\" \",\"pages\":\"1621-1628\"},\"PeriodicalIF\":1.1000,\"publicationDate\":\"2024-11-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Perfusion-Uk\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1177/02676591231206016\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2023/10/19 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q4\",\"JCRName\":\"CARDIAC & CARDIOVASCULAR SYSTEMS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Perfusion-Uk","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/02676591231206016","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/10/19 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

摘要

心脏手术期间常规给予顺行停搏液。体外循环机顺行心脏停搏液的输送有很多变量。许多灌注师完全依靠心脏停搏液系统的压力来确保安全顺产。我们的小组回顾了50例接受冠状动脉搭桥术的患者顺行心脏停搏液的输送情况。收集的数据包括心脏停搏液系统压力和患者的直接主动脉根部压力。对数据的分析发现,这两种压力之间的相关性很弱,平均差很大,标准差很大。结果表明,直接测量主动脉根部压力可作为顺行停搏液的指导,而不是仅依赖停搏液系统压力。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Is cardioplegia system pressure the optimal measure of coronary perfusion during antegrade cardioplegia delivery? A critical review of pressure measurements for optimal antegrade delivery.

Antegrade cardioplegia is routinely given during cardiac surgery. The delivery of antegrade cardioplegia from the cardiopulmonary bypass machine has many variables. Many perfusionists rely exclusively on cardioplegia system pressure to ensure safe antegrade delivery. Our group reviewed antegrade cardioplegia delivery in 50 patients undergoing coronary artery bypass graft. The data collected included the cardioplegia system pressure and the patient's direct aortic root pressure. The analysis of the data found weak correlation between the two pressures with a large mean difference and a wide standard deviation. The results suggest the direct measurement of aortic root pressure as guidance to antegrade cardioplegia instead of relying solely on cardioplegia system pressure.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Perfusion-Uk
Perfusion-Uk 医学-外周血管病
CiteScore
3.00
自引率
8.30%
发文量
203
审稿时长
6-12 weeks
期刊介绍: Perfusion is an ISI-ranked, peer-reviewed scholarly journal, which provides current information on all aspects of perfusion, oxygenation and biocompatibility and their use in modern cardiac surgery. The journal is at the forefront of international research and development and presents an appropriately multidisciplinary approach to perfusion science.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信